Vulnerabilites related to abbott - accent_st
CVE-2017-12714 (GCVE-0-2017-12714)
Vulnerability from cvelistv5
Published
2018-04-25 13:00
Modified
2024-09-17 00:16
Severity ?
CWE
  • CWE-920 - Improper Restriction of power consumption
Summary
Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
References
Impacted products
Vendor Product Version
Abbott Laboratories Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI. Version: All versions of pacemakers manufactured prior to August 28, 2017
Create a notification for this product.
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-05T18:43:56.537Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
          },
          {
            "name": "100523",
            "tags": [
              "vdb-entry",
              "x_refsource_BID",
              "x_transferred"
            ],
            "url": "http://www.securityfocus.com/bid/100523"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "product": "Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI.",
          "vendor": "Abbott Laboratories",
          "versions": [
            {
              "status": "affected",
              "version": "All versions of pacemakers manufactured prior to August 28, 2017"
            }
          ]
        }
      ],
      "datePublic": "2017-08-29T00:00:00",
      "descriptions": [
        {
          "lang": "en",
          "value": "Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted \"RF wake-up\" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-920",
              "description": "Improper Restriction of power consumption CWE-920",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2018-04-26T09:57:01",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "tags": [
            "x_refsource_MISC"
          ],
          "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
        },
        {
          "name": "100523",
          "tags": [
            "vdb-entry",
            "x_refsource_BID"
          ],
          "url": "http://www.securityfocus.com/bid/100523"
        }
      ],
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "DATE_PUBLIC": "2017-08-29T00:00:00",
          "ID": "CVE-2017-12714",
          "STATE": "PUBLIC"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI.",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "All versions of pacemakers manufactured prior to August 28, 2017"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Abbott Laboratories"
              }
            ]
          }
        },
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted \"RF wake-up\" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
            }
          ]
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "Improper Restriction of power consumption CWE-920"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01",
              "refsource": "MISC",
              "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
            },
            {
              "name": "100523",
              "refsource": "BID",
              "url": "http://www.securityfocus.com/bid/100523"
            }
          ]
        }
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2017-12714",
    "datePublished": "2018-04-25T13:00:00Z",
    "dateReserved": "2017-08-09T00:00:00",
    "dateUpdated": "2024-09-17T00:16:50.163Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}

CVE-2017-12712 (GCVE-0-2017-12712)
Vulnerability from cvelistv5
Published
2018-04-25 13:00
Modified
2024-09-17 03:48
Severity ?
CWE
  • CWE-287 - Improper authentication
Summary
The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
References
Impacted products
Vendor Product Version
Abbott Laboratories Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI. Version: All versions of pacemakers manufactured prior to August 28, 2017
Create a notification for this product.
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-05T18:43:56.613Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
          },
          {
            "name": "100523",
            "tags": [
              "vdb-entry",
              "x_refsource_BID",
              "x_transferred"
            ],
            "url": "http://www.securityfocus.com/bid/100523"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "product": "Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI.",
          "vendor": "Abbott Laboratories",
          "versions": [
            {
              "status": "affected",
              "version": "All versions of pacemakers manufactured prior to August 28, 2017"
            }
          ]
        }
      ],
      "datePublic": "2017-08-29T00:00:00",
      "descriptions": [
        {
          "lang": "en",
          "value": "The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-287",
              "description": "Improper authentication CWE-287",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2018-04-26T09:57:01",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "tags": [
            "x_refsource_MISC"
          ],
          "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
        },
        {
          "name": "100523",
          "tags": [
            "vdb-entry",
            "x_refsource_BID"
          ],
          "url": "http://www.securityfocus.com/bid/100523"
        }
      ],
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "DATE_PUBLIC": "2017-08-29T00:00:00",
          "ID": "CVE-2017-12712",
          "STATE": "PUBLIC"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI.",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "All versions of pacemakers manufactured prior to August 28, 2017"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Abbott Laboratories"
              }
            ]
          }
        },
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
            }
          ]
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "Improper authentication CWE-287"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01",
              "refsource": "MISC",
              "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
            },
            {
              "name": "100523",
              "refsource": "BID",
              "url": "http://www.securityfocus.com/bid/100523"
            }
          ]
        }
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2017-12712",
    "datePublished": "2018-04-25T13:00:00Z",
    "dateReserved": "2017-08-09T00:00:00",
    "dateUpdated": "2024-09-17T03:48:37.996Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}

CVE-2017-12716 (GCVE-0-2017-12716)
Vulnerability from cvelistv5
Published
2018-04-25 13:00
Modified
2024-09-17 01:55
Severity ?
CWE
  • CWE-311 - Missing encryption of sensitive data
Summary
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
References
Impacted products
Vendor Product Version
Abbott Laboratories Accent and Anthem Version: All versions of pacemakers manufactured prior to August 28, 2017
Create a notification for this product.
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-05T18:43:56.589Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
          },
          {
            "name": "100523",
            "tags": [
              "vdb-entry",
              "x_refsource_BID",
              "x_transferred"
            ],
            "url": "http://www.securityfocus.com/bid/100523"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "product": "Accent and Anthem",
          "vendor": "Abbott Laboratories",
          "versions": [
            {
              "status": "affected",
              "version": "All versions of pacemakers manufactured prior to August 28, 2017"
            }
          ]
        }
      ],
      "datePublic": "2017-08-29T00:00:00",
      "descriptions": [
        {
          "lang": "en",
          "value": "Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-311",
              "description": "Missing encryption of sensitive data CWE-311",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2018-04-26T09:57:01",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "tags": [
            "x_refsource_MISC"
          ],
          "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
        },
        {
          "name": "100523",
          "tags": [
            "vdb-entry",
            "x_refsource_BID"
          ],
          "url": "http://www.securityfocus.com/bid/100523"
        }
      ],
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "DATE_PUBLIC": "2017-08-29T00:00:00",
          "ID": "CVE-2017-12716",
          "STATE": "PUBLIC"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "Accent and Anthem",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "All versions of pacemakers manufactured prior to August 28, 2017"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Abbott Laboratories"
              }
            ]
          }
        },
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
            }
          ]
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "Missing encryption of sensitive data CWE-311"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01",
              "refsource": "MISC",
              "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
            },
            {
              "name": "100523",
              "refsource": "BID",
              "url": "http://www.securityfocus.com/bid/100523"
            }
          ]
        }
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2017-12716",
    "datePublished": "2018-04-25T13:00:00Z",
    "dateReserved": "2017-08-09T00:00:00",
    "dateUpdated": "2024-09-17T01:55:36.387Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}

Vulnerability from fkie_nvd
Published
2018-04-25 13:29
Modified
2024-11-21 03:10
Summary
Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.



{
  "configurations": [
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:accent_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "BFF25F4E-CF32-41A5-9AEC-5CF2A1D70732",
              "versionEndExcluding": "f0b.0e.7e",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:accent:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "7261AA88-1BD6-4CDF-AFC0-31FD7F52B9E7",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:anthem_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "D6C940D7-5EA3-4D42-8FFE-0C38D2D0065E",
              "versionEndExcluding": "f0b.0e.7e",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:anthem:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "4B28402F-D4DF-448B-8ED3-676E0B438331",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:accent_mri_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "BFC7CD6D-57F8-4479-A25C-9B9937FD3793",
              "versionEndExcluding": "f10.08.6c",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:accent_mri:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "C752B8AA-8990-43DE-AB8B-57329E1E0AE1",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:accent_st_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "3CCA1805-9253-4267-ACE9-B9F3BBB1549A",
              "versionEndExcluding": "f10.08.6c",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:accent_st:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "794620AF-C8D5-4511-B4AF-5E8B4347F558",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:assurity_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "0BF37E32-78D2-48CD-BF19-17533E3CB5DF",
              "versionEndExcluding": "f14.07.80",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:assurity:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "83B2EBFC-FB8A-402E-8C5C-118D4362B143",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:allure_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "858B3E80-0179-4AD7-BC32-3AA87A7341C6",
              "versionEndExcluding": "f14.07.80",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:allure:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "65B30268-EC36-42C3-8028-D345DC22A3DC",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:assurity_mri_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "95E52CB5-DB4A-42FE-B963-CE891D3C1A95",
              "versionEndExcluding": "f17.01.49",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:assurity_mri:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "43025EB2-8644-4BC5-BC3D-D67305C504B7",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    }
  ],
  "cveTags": [],
  "descriptions": [
    {
      "lang": "en",
      "value": "Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted \"RF wake-up\" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
    },
    {
      "lang": "es",
      "value": "Los marcapasos de Abbott Laboratories fabricados antes del 28 de agosto de 2017 no restringen o limitan el n\u00famero de comandos \"RF wake-up\" formateados correctamente que pueden recibir. Esto puede permitir que un atacante cercano env\u00ede comandos repetidamente para reducir la bater\u00eda del marcapasos. Puntuaci\u00f3n base de CVSS v3: 5.3, cadena de vector CVSS: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott ha desarrollado una actualizaci\u00f3n de firmware para ayudar a mitigar las vulnerabilidades identificadas."
    }
  ],
  "id": "CVE-2017-12714",
  "lastModified": "2024-11-21T03:10:04.977",
  "metrics": {
    "cvssMetricV2": [
      {
        "acInsufInfo": false,
        "baseSeverity": "MEDIUM",
        "cvssData": {
          "accessComplexity": "LOW",
          "accessVector": "ADJACENT_NETWORK",
          "authentication": "NONE",
          "availabilityImpact": "COMPLETE",
          "baseScore": 6.1,
          "confidentialityImpact": "NONE",
          "integrityImpact": "NONE",
          "vectorString": "AV:A/AC:L/Au:N/C:N/I:N/A:C",
          "version": "2.0"
        },
        "exploitabilityScore": 6.5,
        "impactScore": 6.9,
        "obtainAllPrivilege": false,
        "obtainOtherPrivilege": false,
        "obtainUserPrivilege": false,
        "source": "nvd@nist.gov",
        "type": "Primary",
        "userInteractionRequired": false
      }
    ],
    "cvssMetricV30": [
      {
        "cvssData": {
          "attackComplexity": "LOW",
          "attackVector": "ADJACENT_NETWORK",
          "availabilityImpact": "HIGH",
          "baseScore": 6.5,
          "baseSeverity": "MEDIUM",
          "confidentialityImpact": "NONE",
          "integrityImpact": "NONE",
          "privilegesRequired": "NONE",
          "scope": "UNCHANGED",
          "userInteraction": "NONE",
          "vectorString": "CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
          "version": "3.0"
        },
        "exploitabilityScore": 2.8,
        "impactScore": 3.6,
        "source": "nvd@nist.gov",
        "type": "Primary"
      }
    ]
  },
  "published": "2018-04-25T13:29:00.287",
  "references": [
    {
      "source": "ics-cert@hq.dhs.gov",
      "tags": [
        "Third Party Advisory",
        "VDB Entry"
      ],
      "url": "http://www.securityfocus.com/bid/100523"
    },
    {
      "source": "ics-cert@hq.dhs.gov",
      "tags": [
        "Third Party Advisory",
        "US Government Resource"
      ],
      "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
    },
    {
      "source": "af854a3a-2127-422b-91ae-364da2661108",
      "tags": [
        "Third Party Advisory",
        "VDB Entry"
      ],
      "url": "http://www.securityfocus.com/bid/100523"
    },
    {
      "source": "af854a3a-2127-422b-91ae-364da2661108",
      "tags": [
        "Third Party Advisory",
        "US Government Resource"
      ],
      "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
    }
  ],
  "sourceIdentifier": "ics-cert@hq.dhs.gov",
  "vulnStatus": "Modified",
  "weaknesses": [
    {
      "description": [
        {
          "lang": "en",
          "value": "CWE-920"
        }
      ],
      "source": "ics-cert@hq.dhs.gov",
      "type": "Secondary"
    },
    {
      "description": [
        {
          "lang": "en",
          "value": "CWE-920"
        }
      ],
      "source": "nvd@nist.gov",
      "type": "Primary"
    }
  ]
}

Vulnerability from fkie_nvd
Published
2018-04-25 13:29
Modified
2024-11-21 03:10
Summary
The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.



{
  "configurations": [
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:accent_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "BFF25F4E-CF32-41A5-9AEC-5CF2A1D70732",
              "versionEndExcluding": "f0b.0e.7e",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:accent:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "7261AA88-1BD6-4CDF-AFC0-31FD7F52B9E7",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:anthem_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "D6C940D7-5EA3-4D42-8FFE-0C38D2D0065E",
              "versionEndExcluding": "f0b.0e.7e",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:anthem:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "4B28402F-D4DF-448B-8ED3-676E0B438331",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:accent_mri_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "BFC7CD6D-57F8-4479-A25C-9B9937FD3793",
              "versionEndExcluding": "f10.08.6c",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:accent_mri:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "C752B8AA-8990-43DE-AB8B-57329E1E0AE1",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:accent_st_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "3CCA1805-9253-4267-ACE9-B9F3BBB1549A",
              "versionEndExcluding": "f10.08.6c",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:accent_st:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "794620AF-C8D5-4511-B4AF-5E8B4347F558",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:assurity_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "0BF37E32-78D2-48CD-BF19-17533E3CB5DF",
              "versionEndExcluding": "f14.07.80",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:assurity:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "83B2EBFC-FB8A-402E-8C5C-118D4362B143",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:allure_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "858B3E80-0179-4AD7-BC32-3AA87A7341C6",
              "versionEndExcluding": "f14.07.80",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:allure:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "65B30268-EC36-42C3-8028-D345DC22A3DC",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:assurity_mri_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "95E52CB5-DB4A-42FE-B963-CE891D3C1A95",
              "versionEndExcluding": "f17.01.49",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:assurity_mri:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "43025EB2-8644-4BC5-BC3D-D67305C504B7",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    }
  ],
  "cveTags": [],
  "descriptions": [
    {
      "lang": "en",
      "value": "The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
    },
    {
      "lang": "es",
      "value": "El algoritmo de autenticaci\u00f3n en los marcapasos de Abbott Laboratories fabricados antes del 28 de agosto de 2017, que est\u00e1 relacionado con una clave de autenticaci\u00f3n y una marca de tiempo, puede comprometerse u omitirse. Esto puede permitir que un atacante cercano env\u00ede comandos no autorizados al marcapasos mediante comunicaciones de radiofrecuencia. Puntuaci\u00f3n base de CVSS v3: 7.5, cadena de vector CVSS: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott ha desarrollado una actualizaci\u00f3n de firmware para ayudar a mitigar las vulnerabilidades identificadas."
    }
  ],
  "id": "CVE-2017-12712",
  "lastModified": "2024-11-21T03:10:04.707",
  "metrics": {
    "cvssMetricV2": [
      {
        "acInsufInfo": false,
        "baseSeverity": "HIGH",
        "cvssData": {
          "accessComplexity": "LOW",
          "accessVector": "ADJACENT_NETWORK",
          "authentication": "NONE",
          "availabilityImpact": "COMPLETE",
          "baseScore": 8.3,
          "confidentialityImpact": "COMPLETE",
          "integrityImpact": "COMPLETE",
          "vectorString": "AV:A/AC:L/Au:N/C:C/I:C/A:C",
          "version": "2.0"
        },
        "exploitabilityScore": 6.5,
        "impactScore": 10.0,
        "obtainAllPrivilege": false,
        "obtainOtherPrivilege": false,
        "obtainUserPrivilege": false,
        "source": "nvd@nist.gov",
        "type": "Primary",
        "userInteractionRequired": false
      }
    ],
    "cvssMetricV30": [
      {
        "cvssData": {
          "attackComplexity": "LOW",
          "attackVector": "ADJACENT_NETWORK",
          "availabilityImpact": "HIGH",
          "baseScore": 8.8,
          "baseSeverity": "HIGH",
          "confidentialityImpact": "HIGH",
          "integrityImpact": "HIGH",
          "privilegesRequired": "NONE",
          "scope": "UNCHANGED",
          "userInteraction": "NONE",
          "vectorString": "CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
          "version": "3.0"
        },
        "exploitabilityScore": 2.8,
        "impactScore": 5.9,
        "source": "nvd@nist.gov",
        "type": "Primary"
      }
    ]
  },
  "published": "2018-04-25T13:29:00.227",
  "references": [
    {
      "source": "ics-cert@hq.dhs.gov",
      "tags": [
        "Third Party Advisory",
        "VDB Entry"
      ],
      "url": "http://www.securityfocus.com/bid/100523"
    },
    {
      "source": "ics-cert@hq.dhs.gov",
      "tags": [
        "Third Party Advisory",
        "US Government Resource"
      ],
      "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
    },
    {
      "source": "af854a3a-2127-422b-91ae-364da2661108",
      "tags": [
        "Third Party Advisory",
        "VDB Entry"
      ],
      "url": "http://www.securityfocus.com/bid/100523"
    },
    {
      "source": "af854a3a-2127-422b-91ae-364da2661108",
      "tags": [
        "Third Party Advisory",
        "US Government Resource"
      ],
      "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
    }
  ],
  "sourceIdentifier": "ics-cert@hq.dhs.gov",
  "vulnStatus": "Modified",
  "weaknesses": [
    {
      "description": [
        {
          "lang": "en",
          "value": "CWE-287"
        }
      ],
      "source": "ics-cert@hq.dhs.gov",
      "type": "Secondary"
    },
    {
      "description": [
        {
          "lang": "en",
          "value": "CWE-287"
        }
      ],
      "source": "nvd@nist.gov",
      "type": "Primary"
    }
  ]
}

Vulnerability from fkie_nvd
Published
2018-04-25 13:29
Modified
2024-11-21 03:10
Summary
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.



{
  "configurations": [
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:accent_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "BFF25F4E-CF32-41A5-9AEC-5CF2A1D70732",
              "versionEndExcluding": "f0b.0e.7e",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:accent:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "7261AA88-1BD6-4CDF-AFC0-31FD7F52B9E7",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:anthem_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "D6C940D7-5EA3-4D42-8FFE-0C38D2D0065E",
              "versionEndExcluding": "f0b.0e.7e",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:anthem:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "4B28402F-D4DF-448B-8ED3-676E0B438331",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:accent_mri_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "BFC7CD6D-57F8-4479-A25C-9B9937FD3793",
              "versionEndExcluding": "f10.08.6c",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:accent_mri:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "C752B8AA-8990-43DE-AB8B-57329E1E0AE1",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:abbott:accent_st_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "3CCA1805-9253-4267-ACE9-B9F3BBB1549A",
              "versionEndExcluding": "f10.08.6c",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:abbott:accent_st:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "794620AF-C8D5-4511-B4AF-5E8B4347F558",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    }
  ],
  "cveTags": [],
  "descriptions": [
    {
      "lang": "en",
      "value": "Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
    },
    {
      "lang": "es",
      "value": "Los marcapasos Accent y Anthem de Abbott Laboratories fabricados antes del 28 de agosto de 2017 transmiten informaci\u00f3n no cifrada del paciente mediante comunicaciones de radiofrecuencia y unidades de monitorizaci\u00f3n dom\u00e9stica. Adem\u00e1s, los marcapasos Accent y Anthem almacenan la informaci\u00f3n opcional del paciente sin cifrar. Puntuaci\u00f3n base de CVSS v3: 3.1, cadena de vector CVSS: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott ha desarrollado una actualizaci\u00f3n de firmware para ayudar a mitigar las vulnerabilidades identificadas."
    }
  ],
  "id": "CVE-2017-12716",
  "lastModified": "2024-11-21T03:10:05.107",
  "metrics": {
    "cvssMetricV2": [
      {
        "acInsufInfo": false,
        "baseSeverity": "LOW",
        "cvssData": {
          "accessComplexity": "LOW",
          "accessVector": "ADJACENT_NETWORK",
          "authentication": "NONE",
          "availabilityImpact": "NONE",
          "baseScore": 3.3,
          "confidentialityImpact": "PARTIAL",
          "integrityImpact": "NONE",
          "vectorString": "AV:A/AC:L/Au:N/C:P/I:N/A:N",
          "version": "2.0"
        },
        "exploitabilityScore": 6.5,
        "impactScore": 2.9,
        "obtainAllPrivilege": false,
        "obtainOtherPrivilege": false,
        "obtainUserPrivilege": false,
        "source": "nvd@nist.gov",
        "type": "Primary",
        "userInteractionRequired": false
      }
    ],
    "cvssMetricV30": [
      {
        "cvssData": {
          "attackComplexity": "LOW",
          "attackVector": "ADJACENT_NETWORK",
          "availabilityImpact": "NONE",
          "baseScore": 6.5,
          "baseSeverity": "MEDIUM",
          "confidentialityImpact": "HIGH",
          "integrityImpact": "NONE",
          "privilegesRequired": "NONE",
          "scope": "UNCHANGED",
          "userInteraction": "NONE",
          "vectorString": "CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N",
          "version": "3.0"
        },
        "exploitabilityScore": 2.8,
        "impactScore": 3.6,
        "source": "nvd@nist.gov",
        "type": "Primary"
      }
    ]
  },
  "published": "2018-04-25T13:29:00.333",
  "references": [
    {
      "source": "ics-cert@hq.dhs.gov",
      "tags": [
        "Third Party Advisory",
        "VDB Entry"
      ],
      "url": "http://www.securityfocus.com/bid/100523"
    },
    {
      "source": "ics-cert@hq.dhs.gov",
      "tags": [
        "Third Party Advisory",
        "US Government Resource"
      ],
      "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
    },
    {
      "source": "af854a3a-2127-422b-91ae-364da2661108",
      "tags": [
        "Third Party Advisory",
        "VDB Entry"
      ],
      "url": "http://www.securityfocus.com/bid/100523"
    },
    {
      "source": "af854a3a-2127-422b-91ae-364da2661108",
      "tags": [
        "Third Party Advisory",
        "US Government Resource"
      ],
      "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
    }
  ],
  "sourceIdentifier": "ics-cert@hq.dhs.gov",
  "vulnStatus": "Modified",
  "weaknesses": [
    {
      "description": [
        {
          "lang": "en",
          "value": "CWE-311"
        }
      ],
      "source": "ics-cert@hq.dhs.gov",
      "type": "Secondary"
    },
    {
      "description": [
        {
          "lang": "en",
          "value": "CWE-319"
        }
      ],
      "source": "nvd@nist.gov",
      "type": "Primary"
    }
  ]
}