Vulnerabilites related to Becton, Dickinson and Company (BD) - BodyGuard™ Pump
CVE-2022-43557 (GCVE-0-2022-43557)
Vulnerability from cvelistv5
Published
2022-12-05 00:00
Modified
2025-04-23 15:47
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-1299 - Missing Protection Mechanism for Alternate Hardware Interface
Summary
The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specialized equipment and knowledge may be able to configure or disable the pump. No electronic protected health information (ePHI), protected health information (PHI) or personally identifiable information (PII) is stored in the pump.
References
► | URL | Tags |
---|---|---|
Impacted products
Vendor | Product | Version | ||
---|---|---|---|---|
Becton, Dickinson and Company (BD) | BodyGuard™ Pump |
Version: BD BodyGuard™ Version: CME BodyGuard™ 323 (2nd Edition) Version: CME BodyGuard™ 323 Color Vision (2nd Edition) Version: CME BodyGuard™ 323 Color Vision (3rd Edition) Version: CME BodyGuard™ Twins (2nd Edition) |
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-03T13:32:59.566Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "vendor-advisory", "x_transferred" ], "url": "https://www.bd.com/en-us/about-bd/cybersecurity/bulletin/bd-bodyguard-pumps-rs-232-interface-vulnerability" } ], "title": "CVE Program Container" }, { "metrics": [ { "other": { "content": { "id": "CVE-2022-43557", "options": [ { "Exploitation": "none" }, { "Automatable": "no" }, { "Technical Impact": "partial" } ], "role": "CISA Coordinator", "timestamp": "2025-04-23T15:46:47.466198Z", "version": "2.0.3" }, "type": "ssvc" } } ], "providerMetadata": { "dateUpdated": "2025-04-23T15:47:10.734Z", "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0", "shortName": "CISA-ADP" }, "title": "CISA ADP Vulnrichment" } ], "cna": { "affected": [ { "defaultStatus": "unaffected", "product": "BodyGuard\u2122 Pump", "vendor": "Becton, Dickinson and Company (BD)", "versions": [ { "status": "affected", "version": "BD BodyGuard\u2122 " }, { "status": "affected", "version": "CME BodyGuard\u2122 323 (2nd Edition)" }, { "status": "affected", "version": "CME BodyGuard\u2122 323 Color Vision (2nd Edition)" }, { "status": "affected", "version": "CME BodyGuard\u2122 323 Color Vision (3rd Edition)" }, { "status": "affected", "version": "CME BodyGuard\u2122 Twins (2nd Edition)" } ] } ], "datePublic": "2022-10-20T04:00:00.000Z", "descriptions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "The BD BodyGuard\u2122 infusion pumps specified allow for access through the RS-232\u0026nbsp;(serial) port interface. If exploited, threat actors with physical access, specialized equipment and\u0026nbsp;knowledge may be able to configure or disable the pump. No electronic protected health information\u0026nbsp;(ePHI), protected health information (PHI) or personally identifiable information (PII) is stored in the\u0026nbsp;pump." } ], "value": "The BD BodyGuard\u2122 infusion pumps specified allow for access through the RS-232\u00a0(serial) port interface. If exploited, threat actors with physical access, specialized equipment and\u00a0knowledge may be able to configure or disable the pump. No electronic protected health information\u00a0(ePHI), protected health information (PHI) or personally identifiable information (PII) is stored in the\u00a0pump." } ], "impacts": [ { "capecId": "CAPEC-115", "descriptions": [ { "lang": "en", "value": "CAPEC-115 Authentication Bypass" } ] } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "HIGH", "attackVector": "PHYSICAL", "availabilityImpact": "HIGH", "baseScore": 5.3, "baseSeverity": "MEDIUM", "confidentialityImpact": "LOW", "integrityImpact": "LOW", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:H", "version": "3.1" }, "format": "CVSS", "scenarios": [ { "lang": "en", "value": "GENERAL" } ] } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-1299", "description": "CWE-1299 Missing Protection Mechanism for Alternate Hardware Interface", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2023-10-27T20:05:04.787Z", "orgId": "2325d071-eabf-4b7b-a4ea-0819b6629a18", "shortName": "BD" }, "references": [ { "tags": [ "vendor-advisory" ], "url": "https://www.bd.com/en-us/about-bd/cybersecurity/bulletin/bd-bodyguard-pumps-rs-232-interface-vulnerability" } ], "source": { "discovery": "INTERNAL" }, "title": "BD BodyGuard\u2122 Pumps \u2013 RS-232 Interface Vulnerability", "workarounds": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "BD recommends the following mitigations and compensating controls to reduce risk associated with this\u003cbr\u003evulnerability:\u003cbr\u003e\u2022 Ensure physical access controls are in place and only authorized end-users have access to\u003cbr\u003eBD BodyGuard\u2122 pumps.\u003cbr\u003e\u2022 Ensure that only BD-approved equipment is connected to the RS-232 interface of the affected pumps.\u003cbr\u003e\u2022 Ensure that no equipment is connected to the RS-232 interface when the affected pumps are delivering\u003cbr\u003einfusions.\u003cbr\u003e\u2022 Protect connected computer systems with BodyComm\u2122 software with standard security measures.\u003cbr\u003e" } ], "value": "BD recommends the following mitigations and compensating controls to reduce risk associated with this\nvulnerability:\n\u2022 Ensure physical access controls are in place and only authorized end-users have access to\nBD BodyGuard\u2122 pumps.\n\u2022 Ensure that only BD-approved equipment is connected to the RS-232 interface of the affected pumps.\n\u2022 Ensure that no equipment is connected to the RS-232 interface when the affected pumps are delivering\ninfusions.\n\u2022 Protect connected computer systems with BodyComm\u2122 software with standard security measures.\n" } ], "x_generator": { "engine": "Vulnogram 0.1.0-dev" } } }, "cveMetadata": { "assignerOrgId": "2325d071-eabf-4b7b-a4ea-0819b6629a18", "assignerShortName": "BD", "cveId": "CVE-2022-43557", "datePublished": "2022-12-05T00:00:00.000Z", "dateReserved": "2022-10-20T00:00:00.000Z", "dateUpdated": "2025-04-23T15:47:10.734Z", "serial": 1, "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }