Vulnerabilites related to Philips - Brilliance CT Scanners
CVE-2018-8853 (GCVE-0-2018-8853)
Vulnerability from cvelistv5
Published
2018-05-04 17:00
Modified
2024-09-16 17:58
Severity ?
CWE
  • CWE-250 - EXECUTION WITH UNNECESSARY PRIVILEGES
Summary
Philips Brilliance CT devices operate user functions from within a contained kiosk in a Microsoft Windows operating system. Windows boots by default with elevated Windows privileges, enabling a kiosk application, user, or an attacker to potentially attain unauthorized elevated privileges in Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior. Also, attackers may gain access to unauthorized resources from the underlying Windows operating system.
Impacted products
Vendor Product Version
Philips Brilliance CT Scanners Version: Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior.
Create a notification for this product.
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-05T07:10:46.272Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_CONFIRM",
              "x_transferred"
            ],
            "url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
          },
          {
            "name": "104088",
            "tags": [
              "vdb-entry",
              "x_refsource_BID",
              "x_transferred"
            ],
            "url": "http://www.securityfocus.com/bid/104088"
          },
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-123-01"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "product": "Brilliance CT Scanners",
          "vendor": "Philips",
          "versions": [
            {
              "status": "affected",
              "version": "Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior."
            }
          ]
        }
      ],
      "datePublic": "2018-05-01T00:00:00",
      "descriptions": [
        {
          "lang": "en",
          "value": "Philips Brilliance CT devices operate user functions from within a contained kiosk in a Microsoft Windows operating system. Windows boots by default with elevated Windows privileges, enabling a kiosk application, user, or an attacker to potentially attain unauthorized elevated privileges in Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior. Also, attackers may gain access to unauthorized resources from the underlying Windows operating system."
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-250",
              "description": "EXECUTION WITH UNNECESSARY PRIVILEGES CWE-250",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2018-05-08T09:57:01",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "tags": [
            "x_refsource_CONFIRM"
          ],
          "url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
        },
        {
          "name": "104088",
          "tags": [
            "vdb-entry",
            "x_refsource_BID"
          ],
          "url": "http://www.securityfocus.com/bid/104088"
        },
        {
          "tags": [
            "x_refsource_MISC"
          ],
          "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-123-01"
        }
      ],
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "DATE_PUBLIC": "2018-05-01T00:00:00",
          "ID": "CVE-2018-8853",
          "STATE": "PUBLIC"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "Brilliance CT Scanners",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior."
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Philips"
              }
            ]
          }
        },
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "Philips Brilliance CT devices operate user functions from within a contained kiosk in a Microsoft Windows operating system. Windows boots by default with elevated Windows privileges, enabling a kiosk application, user, or an attacker to potentially attain unauthorized elevated privileges in Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior. Also, attackers may gain access to unauthorized resources from the underlying Windows operating system."
            }
          ]
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "EXECUTION WITH UNNECESSARY PRIVILEGES CWE-250"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://www.usa.philips.com/healthcare/about/customer-support/product-security",
              "refsource": "CONFIRM",
              "url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
            },
            {
              "name": "104088",
              "refsource": "BID",
              "url": "http://www.securityfocus.com/bid/104088"
            },
            {
              "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-123-01",
              "refsource": "MISC",
              "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-123-01"
            }
          ]
        }
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2018-8853",
    "datePublished": "2018-05-04T17:00:00Z",
    "dateReserved": "2018-03-20T00:00:00",
    "dateUpdated": "2024-09-16T17:58:36.306Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}

CVE-2018-8857 (GCVE-0-2018-8857)
Vulnerability from cvelistv5
Published
2018-05-04 17:00
Modified
2024-09-16 21:02
Severity ?
CWE
  • CWE-798 - USE OF HARD-CODED CREDENTIALS
Summary
Philips Brilliance CT software (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) contains fixed credentials, such as a password or cryptographic key, which it uses for its own inbound authentication, outbound communication to external components, or encryption of internal data. An attacker could compromise these credentials and gain access to the system.
Impacted products
Vendor Product Version
Philips Brilliance CT Scanners Version: Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior.
Create a notification for this product.
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-05T07:10:46.195Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_CONFIRM",
              "x_transferred"
            ],
            "url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
          },
          {
            "name": "104088",
            "tags": [
              "vdb-entry",
              "x_refsource_BID",
              "x_transferred"
            ],
            "url": "http://www.securityfocus.com/bid/104088"
          },
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-123-01"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "product": "Brilliance CT Scanners",
          "vendor": "Philips",
          "versions": [
            {
              "status": "affected",
              "version": "Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior."
            }
          ]
        }
      ],
      "datePublic": "2018-05-01T00:00:00",
      "descriptions": [
        {
          "lang": "en",
          "value": "Philips Brilliance CT software (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) contains fixed credentials, such as a password or cryptographic key, which it uses for its own inbound authentication, outbound communication to external components, or encryption of internal data. An attacker could compromise these credentials and gain access to the system."
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-798",
              "description": "USE OF HARD-CODED CREDENTIALS CWE-798",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2018-05-08T09:57:01",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "tags": [
            "x_refsource_CONFIRM"
          ],
          "url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
        },
        {
          "name": "104088",
          "tags": [
            "vdb-entry",
            "x_refsource_BID"
          ],
          "url": "http://www.securityfocus.com/bid/104088"
        },
        {
          "tags": [
            "x_refsource_MISC"
          ],
          "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-123-01"
        }
      ],
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "DATE_PUBLIC": "2018-05-01T00:00:00",
          "ID": "CVE-2018-8857",
          "STATE": "PUBLIC"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "Brilliance CT Scanners",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior."
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Philips"
              }
            ]
          }
        },
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "Philips Brilliance CT software (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) contains fixed credentials, such as a password or cryptographic key, which it uses for its own inbound authentication, outbound communication to external components, or encryption of internal data. An attacker could compromise these credentials and gain access to the system."
            }
          ]
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "USE OF HARD-CODED CREDENTIALS CWE-798"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://www.usa.philips.com/healthcare/about/customer-support/product-security",
              "refsource": "CONFIRM",
              "url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
            },
            {
              "name": "104088",
              "refsource": "BID",
              "url": "http://www.securityfocus.com/bid/104088"
            },
            {
              "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-123-01",
              "refsource": "MISC",
              "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-123-01"
            }
          ]
        }
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2018-8857",
    "datePublished": "2018-05-04T17:00:00Z",
    "dateReserved": "2018-03-20T00:00:00",
    "dateUpdated": "2024-09-16T21:02:22.963Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}

CVE-2018-8861 (GCVE-0-2018-8861)
Vulnerability from cvelistv5
Published
2018-05-04 17:00
Modified
2024-09-17 03:59
Severity ?
CWE
  • CWE-668 - EXPOSURE OF RESOURCE TO WRONG SPHERE
Summary
Vulnerabilities within the Philips Brilliance CT kiosk environment (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) could enable a limited-access kiosk user or an unauthorized attacker to break-out from the containment of the kiosk environment, attain elevated privileges from the underlying Windows OS, and access unauthorized resources from the operating system.
Impacted products
Vendor Product Version
Philips Brilliance CT Scanners Version: Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior.
Create a notification for this product.
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-05T07:10:46.540Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_CONFIRM",
              "x_transferred"
            ],
            "url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
          },
          {
            "name": "104088",
            "tags": [
              "vdb-entry",
              "x_refsource_BID",
              "x_transferred"
            ],
            "url": "http://www.securityfocus.com/bid/104088"
          },
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-123-01"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "product": "Brilliance CT Scanners",
          "vendor": "Philips",
          "versions": [
            {
              "status": "affected",
              "version": "Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior."
            }
          ]
        }
      ],
      "datePublic": "2018-05-01T00:00:00",
      "descriptions": [
        {
          "lang": "en",
          "value": "Vulnerabilities within the Philips Brilliance CT kiosk environment (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) could enable a limited-access kiosk user or an unauthorized attacker to break-out from the containment of the kiosk environment, attain elevated privileges from the underlying Windows OS, and access unauthorized resources from the operating system."
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-668",
              "description": "EXPOSURE OF RESOURCE TO WRONG SPHERE CWE-668",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2018-05-08T09:57:01",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "tags": [
            "x_refsource_CONFIRM"
          ],
          "url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
        },
        {
          "name": "104088",
          "tags": [
            "vdb-entry",
            "x_refsource_BID"
          ],
          "url": "http://www.securityfocus.com/bid/104088"
        },
        {
          "tags": [
            "x_refsource_MISC"
          ],
          "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-123-01"
        }
      ],
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "DATE_PUBLIC": "2018-05-01T00:00:00",
          "ID": "CVE-2018-8861",
          "STATE": "PUBLIC"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "Brilliance CT Scanners",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior."
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Philips"
              }
            ]
          }
        },
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "Vulnerabilities within the Philips Brilliance CT kiosk environment (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) could enable a limited-access kiosk user or an unauthorized attacker to break-out from the containment of the kiosk environment, attain elevated privileges from the underlying Windows OS, and access unauthorized resources from the operating system."
            }
          ]
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "EXPOSURE OF RESOURCE TO WRONG SPHERE CWE-668"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://www.usa.philips.com/healthcare/about/customer-support/product-security",
              "refsource": "CONFIRM",
              "url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
            },
            {
              "name": "104088",
              "refsource": "BID",
              "url": "http://www.securityfocus.com/bid/104088"
            },
            {
              "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-123-01",
              "refsource": "MISC",
              "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-123-01"
            }
          ]
        }
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2018-8861",
    "datePublished": "2018-05-04T17:00:00Z",
    "dateReserved": "2018-03-20T00:00:00",
    "dateUpdated": "2024-09-17T03:59:05.842Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",