Vulnerabilites related to Medtronic - Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States)
CVE-2019-13531 (GCVE-0-2019-13531)
Vulnerability from cvelistv5
Published
2019-11-08 19:46
Modified
2025-05-22 18:37
CWE
  • CWE-287 - Improper Authentication
Summary
In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-04T23:57:39.240Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://www.us-cert.gov/ics/advisories/icsma-19-311-01"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "Valleylab FT10 Energy Platform (VLFT10GEN)",
          "vendor": "Medtronic",
          "versions": [
            {
              "lessThanOrEqual": "2.1.0",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            },
            {
              "lessThanOrEqual": "2.0.3",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States)",
          "vendor": "Medtronic",
          "versions": [
            {
              "lessThanOrEqual": "1.20.2",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Medtronic reported these vulnerabilities to CISA."
        }
      ],
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eIn Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.\u003c/p\u003e"
            }
          ],
          "value": "In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "HIGH",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "LOW",
            "baseScore": 4.8,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "NONE",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:L",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-287",
              "description": "CWE-287 Improper Authentication",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2025-05-22T18:37:04.526Z",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/valleylab-generator-rfid-vulnerabilities.html"
        },
        {
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-311-01"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eA software patch is available now for the affected Valleylab platforms. If you suspect you are in possession of an instrument that is not FDA approved or cleared to be used with Medtronic Valleylab FT10 or LS10, please contact Medtronic or your medical device supplier. If you have concerns about FDA clearance or approval of current or future instruments, please contact your medical device supplier. Please contact \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/covidien/en-us/support/software.html\"\u003eMedtronic\u003c/a\u003e\u0026nbsp;to obtain the software patch.\u003c/p\u003e\u003cp\u003eMedtronic has released additional patient focused information at the following location:\u003c/p\u003e\u003cp\u003e\u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/security\"\u003ehttps://www.medtronic.com/security\u003c/a\u003e\u003c/p\u003e\n\n\u003cbr\u003e"
            }
          ],
          "value": "A software patch is available now for the affected Valleylab platforms. If you suspect you are in possession of an instrument that is not FDA approved or cleared to be used with Medtronic Valleylab FT10 or LS10, please contact Medtronic or your medical device supplier. If you have concerns about FDA clearance or approval of current or future instruments, please contact your medical device supplier. Please contact  https://www.medtronic.com/security"
        }
      ],
      "source": {
        "advisory": "ICSMA-19-311-01",
        "discovery": "INTERNAL"
      },
      "title": "Medtronic Valleylab FT10 and LS10 Improper Authentication",
      "x_generator": {
        "engine": "Vulnogram 0.2.0"
      },
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "ID": "CVE-2019-13531",
          "STATE": "PUBLIC"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "Valleylab FT10 Energy Platform (VLFT10GEN)",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "version 2.1.0 and lower"
                          },
                          {
                            "version_value": "version 2.0.3 and lower"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States)",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "version 1.20.2 and lower"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Medtronic"
              }
            ]
          }
        },
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator."
            }
          ]
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "IMPROPER AUTHENTICATION CWE-287"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://www.us-cert.gov/ics/advisories/icsma-19-311-01",
              "refsource": "MISC",
              "url": "https://www.us-cert.gov/ics/advisories/icsma-19-311-01"
            }
          ]
        }
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2019-13531",
    "datePublished": "2019-11-08T19:46:45",
    "dateReserved": "2019-07-11T00:00:00",
    "dateUpdated": "2025-05-22T18:37:04.526Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}

CVE-2019-13535 (GCVE-0-2019-13535)
Vulnerability from cvelistv5
Published
2019-11-08 19:45
Modified
2025-05-22 18:39
CWE
Summary
In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism does not apply read protection, allowing for full read access of the RFID security mechanism data.
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-04T23:57:39.351Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://www.us-cert.gov/ics/advisories/icsma-19-311-01"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "Valleylab FT10 Energy Platform (VLFT10GEN)",
          "vendor": "Medtronic",
          "versions": [
            {
              "lessThanOrEqual": "2.1.0",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            },
            {
              "lessThanOrEqual": "2.0.3",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States)",
          "vendor": "Medtronic",
          "versions": [
            {
              "lessThanOrEqual": "1.20.2",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Medtronic reported these vulnerabilities to CISA."
        }
      ],
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003e\nIn Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States) version 1.20.2 and lower, the RFID security mechanism does not apply read protection, allowing for full read access of the RFID security mechanism data.\n\n\u003c/p\u003e"
            }
          ],
          "value": "In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States) version 1.20.2 and lower, the RFID security mechanism does not apply read protection, allowing for full read access of the RFID security mechanism data."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "NONE",
            "baseScore": 4.6,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "HIGH",
            "integrityImpact": "NONE",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-693",
              "description": "CWE-693",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2025-05-22T18:39:28.797Z",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/valleylab-generator-rfid-vulnerabilities.html"
        },
        {
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-311-01"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eA software patch is available now for the affected Valleylab platforms. If you suspect you are in possession of an instrument that is not FDA approved or cleared to be used with Medtronic Valleylab FT10 or LS10, please contact Medtronic or your medical device supplier. If you have concerns about FDA clearance or approval of current or future instruments, please contact your medical device supplier. Please contact \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/covidien/en-us/support/software.html\"\u003eMedtronic\u003c/a\u003e\u0026nbsp;to obtain the software patch.\u003c/p\u003e\u003cp\u003eMedtronic has released additional patient focused information at the following location:\u003c/p\u003e\u003cp\u003e\u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/security\"\u003ehttps://www.medtronic.com/security\u003c/a\u003e\u003c/p\u003e\n\n\u003cbr\u003e"
            }
          ],
          "value": "A software patch is available now for the affected Valleylab platforms. If you suspect you are in possession of an instrument that is not FDA approved or cleared to be used with Medtronic Valleylab FT10 or LS10, please contact Medtronic or your medical device supplier. If you have concerns about FDA clearance or approval of current or future instruments, please contact your medical device supplier. Please contact  https://www.medtronic.com/security"
        }
      ],
      "source": {
        "advisory": "ICSMA-19-311-01",
        "discovery": "INTERNAL"
      },
      "title": "Medtronic Valleylab FT10 and LS10 Protection Mechanism Failure",
      "x_generator": {
        "engine": "Vulnogram 0.2.0"
      },
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "ID": "CVE-2019-13531",
          "STATE": "PUBLIC"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "Valleylab FT10 Energy Platform (VLFT10GEN)",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "version 2.1.0 and lower"
                          },
                          {
                            "version_value": "version 2.0.3 and lower"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States)",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "version 1.20.2 and lower"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Medtronic"
              }
            ]
          }
        },
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator."
            }
          ]
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "IMPROPER AUTHENTICATION CWE-287"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://www.us-cert.gov/ics/advisories/icsma-19-311-01",
              "refsource": "MISC",
              "url": "https://www.us-cert.gov/ics/advisories/icsma-19-311-01"
            }
          ]
        }
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2019-13535",
    "datePublished": "2019-11-08T19:45:55",
    "dateReserved": "2019-07-11T00:00:00",
    "dateUpdated": "2025-05-22T18:39:28.797Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",