Vulnerabilites related to Boston Scientific - ZOOM LATITUDE
CVE-2021-38400 (GCVE-0-2021-38400)
Vulnerability from cvelistv5
Published
2021-10-04 17:35
Modified
2024-09-16 23:31
CWE
  • CWE-916 - Use of Password Hash With Insufficient Computational Effort
Summary
An attacker with physical access to Boston Scientific Zoom Latitude Model 3120 can remove the hard disk drive or create a specially crafted USB to extract the password hash for brute force reverse engineering of the system password.
References
Impacted products
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-04T01:37:16.609Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "product": "ZOOM LATITUDE",
          "vendor": "Boston Scientific",
          "versions": [
            {
              "status": "affected",
              "version": "Model 3120"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster  University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus  Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices  and reported them to Boston Scientific."
        }
      ],
      "datePublic": "2021-09-30T00:00:00",
      "descriptions": [
        {
          "lang": "en",
          "value": "An attacker with physical access to Boston Scientific Zoom Latitude Model 3120 can remove the hard disk drive or create a specially crafted USB to extract the password hash for brute force reverse engineering of the system password."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "HIGH",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "LOW",
            "baseScore": 6.9,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "HIGH",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "CHANGED",
            "userInteraction": "REQUIRED",
            "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:C/C:H/I:H/A:L",
            "version": "3.1"
          }
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-916",
              "description": "CWE-916 Use of Password Hash With Insufficient Computational Effort",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2021-10-04T17:35:28",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "tags": [
            "x_refsource_MISC"
          ],
          "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
        }
      ],
      "source": {
        "advisory": "ICSMA-21-273-01",
        "defect": [
          "CWE-916"
        ],
        "discovery": "EXTERNAL"
      },
      "title": "Use of Password Hash with Insufficient Computational Effort for Boston Scientific Zoom Latitude",
      "workarounds": [
        {
          "lang": "en",
          "value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security,  the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."
        }
      ],
      "x_generator": {
        "engine": "Vulnogram 0.0.9"
      },
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "DATE_PUBLIC": "2021-09-30T21:02:00.000Z",
          "ID": "CVE-2021-38400",
          "STATE": "PUBLIC",
          "TITLE": "Use of Password Hash with Insufficient Computational Effort for Boston Scientific Zoom Latitude"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "ZOOM LATITUDE",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "Model 3120"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Boston Scientific"
              }
            ]
          }
        },
        "credit": [
          {
            "lang": "eng",
            "value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster  University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus  Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices  and reported them to Boston Scientific."
          }
        ],
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "An attacker with physical access to Boston Scientific Zoom Latitude Model 3120 can remove the hard disk drive or create a specially crafted USB to extract the password hash for brute force reverse engineering of the system password."
            }
          ]
        },
        "generator": {
          "engine": "Vulnogram 0.0.9"
        },
        "impact": {
          "cvss": {
            "attackComplexity": "HIGH",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "LOW",
            "baseScore": 6.9,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "HIGH",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "CHANGED",
            "userInteraction": "REQUIRED",
            "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:C/C:H/I:H/A:L",
            "version": "3.1"
          }
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "CWE-916 Use of Password Hash With Insufficient Computational Effort"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01",
              "refsource": "MISC",
              "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
            }
          ]
        },
        "source": {
          "advisory": "ICSMA-21-273-01",
          "defect": [
            "CWE-916"
          ],
          "discovery": "EXTERNAL"
        },
        "work_around": [
          {
            "lang": "en",
            "value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security,  the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."
          }
        ]
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2021-38400",
    "datePublished": "2021-10-04T17:35:28.144047Z",
    "dateReserved": "2021-08-10T00:00:00",
    "dateUpdated": "2024-09-16T23:31:04.619Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}

CVE-2021-38396 (GCVE-0-2021-38396)
Vulnerability from cvelistv5
Published
2021-10-04 17:35
Modified
2024-09-16 20:46
CWE
  • CWE-353 - Missing Support for Integrity Check
Summary
The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB.
References
Impacted products
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-04T01:37:16.545Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "product": "ZOOM LATITUDE",
          "vendor": "Boston Scientific",
          "versions": [
            {
              "status": "affected",
              "version": "Model 3120"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster  University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus  Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices  and reported them to Boston Scientific."
        }
      ],
      "datePublic": "2021-09-30T00:00:00",
      "descriptions": [
        {
          "lang": "en",
          "value": "The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "LOW",
            "baseScore": 6.5,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "LOW",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "CHANGED",
            "userInteraction": "REQUIRED",
            "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L",
            "version": "3.1"
          }
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-353",
              "description": "CWE-353 Missing Support for Integrity Check",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2021-10-04T17:35:21",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "tags": [
            "x_refsource_MISC"
          ],
          "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
        }
      ],
      "source": {
        "advisory": "ICSMA-21-273-01",
        "defect": [
          "CWE-353"
        ],
        "discovery": "EXTERNAL"
      },
      "title": "Missing Support Integrity Check for Boston Scientific Zoom Latitude",
      "workarounds": [
        {
          "lang": "en",
          "value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security,  the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."
        }
      ],
      "x_generator": {
        "engine": "Vulnogram 0.0.9"
      },
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "DATE_PUBLIC": "2021-09-30T21:02:00.000Z",
          "ID": "CVE-2021-38396",
          "STATE": "PUBLIC",
          "TITLE": "Missing Support Integrity Check for Boston Scientific Zoom Latitude"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "ZOOM LATITUDE",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "Model 3120"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Boston Scientific"
              }
            ]
          }
        },
        "credit": [
          {
            "lang": "eng",
            "value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster  University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus  Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices  and reported them to Boston Scientific."
          }
        ],
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB."
            }
          ]
        },
        "generator": {
          "engine": "Vulnogram 0.0.9"
        },
        "impact": {
          "cvss": {
            "attackComplexity": "LOW",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "LOW",
            "baseScore": 6.5,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "LOW",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "CHANGED",
            "userInteraction": "REQUIRED",
            "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L",
            "version": "3.1"
          }
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "CWE-353 Missing Support for Integrity Check"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01",
              "refsource": "MISC",
              "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
            }
          ]
        },
        "source": {
          "advisory": "ICSMA-21-273-01",
          "defect": [
            "CWE-353"
          ],
          "discovery": "EXTERNAL"
        },
        "work_around": [
          {
            "lang": "en",
            "value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security,  the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."
          }
        ]
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2021-38396",
    "datePublished": "2021-10-04T17:35:21.072612Z",
    "dateReserved": "2021-08-10T00:00:00",
    "dateUpdated": "2024-09-16T20:46:57.503Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}

CVE-2021-38394 (GCVE-0-2021-38394)
Vulnerability from cvelistv5
Published
2021-10-04 17:34
Modified
2024-09-16 18:59
CWE
Summary
An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate of a valid hardware key. The hardware key allows access to special settings when inserted.
References
Impacted products
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-04T01:37:16.553Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "product": "ZOOM LATITUDE",
          "vendor": "Boston Scientific",
          "versions": [
            {
              "status": "affected",
              "version": "Model 3120"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster  University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus  Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices  and reported them to Boston Scientific."
        }
      ],
      "datePublic": "2021-09-30T00:00:00",
      "descriptions": [
        {
          "lang": "en",
          "value": "An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate of a valid hardware key. The hardware key allows access to special settings when inserted."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "HIGH",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "LOW",
            "baseScore": 6.2,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "LOW",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "CHANGED",
            "userInteraction": "REQUIRED",
            "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:C/C:L/I:H/A:L",
            "version": "3.1"
          }
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-1278",
              "description": "CWE-1278",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2021-10-04T17:34:58",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "tags": [
            "x_refsource_MISC"
          ],
          "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
        }
      ],
      "source": {
        "advisory": "ICSMA-21-273-01",
        "defect": [
          "CWE-1278"
        ],
        "discovery": "EXTERNAL"
      },
      "title": "Missing Protection against Hardware Reverse Engineering Using Integrated Circuit Imaging Techniques for Boston Scientific Zoom Latitude",
      "workarounds": [
        {
          "lang": "en",
          "value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security,  the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."
        }
      ],
      "x_generator": {
        "engine": "Vulnogram 0.0.9"
      },
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "DATE_PUBLIC": "2021-09-30T21:02:00.000Z",
          "ID": "CVE-2021-38394",
          "STATE": "PUBLIC",
          "TITLE": "Missing Protection against Hardware Reverse Engineering Using Integrated Circuit Imaging Techniques for Boston Scientific Zoom Latitude"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "ZOOM LATITUDE",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "Model 3120"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Boston Scientific"
              }
            ]
          }
        },
        "credit": [
          {
            "lang": "eng",
            "value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster  University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus  Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices  and reported them to Boston Scientific."
          }
        ],
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate of a valid hardware key. The hardware key allows access to special settings when inserted."
            }
          ]
        },
        "generator": {
          "engine": "Vulnogram 0.0.9"
        },
        "impact": {
          "cvss": {
            "attackComplexity": "HIGH",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "LOW",
            "baseScore": 6.2,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "LOW",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "CHANGED",
            "userInteraction": "REQUIRED",
            "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:C/C:L/I:H/A:L",
            "version": "3.1"
          }
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "CWE-1278"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01",
              "refsource": "MISC",
              "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
            }
          ]
        },
        "source": {
          "advisory": "ICSMA-21-273-01",
          "defect": [
            "CWE-1278"
          ],
          "discovery": "EXTERNAL"
        },
        "work_around": [
          {
            "lang": "en",
            "value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security,  the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."
          }
        ]
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2021-38394",
    "datePublished": "2021-10-04T17:34:58.296378Z",
    "dateReserved": "2021-08-10T00:00:00",
    "dateUpdated": "2024-09-16T18:59:19.736Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}

CVE-2021-38392 (GCVE-0-2021-38392)
Vulnerability from cvelistv5
Published
2021-10-04 17:35
Modified
2024-09-17 03:33
CWE
  • CWE-284 - Improper Access Control
Summary
A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world.
References
Impacted products
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-04T01:37:16.580Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "product": "ZOOM LATITUDE",
          "vendor": "Boston Scientific",
          "versions": [
            {
              "status": "affected",
              "version": "Model 3120"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster  University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus  Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices  and reported them to Boston Scientific."
        }
      ],
      "datePublic": "2021-09-30T00:00:00",
      "descriptions": [
        {
          "lang": "en",
          "value": "A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "LOW",
            "baseScore": 6.5,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "LOW",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "CHANGED",
            "userInteraction": "REQUIRED",
            "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L",
            "version": "3.1"
          }
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-284",
              "description": "CWE-284 Improper Access Control",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2021-10-04T17:35:13",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "tags": [
            "x_refsource_MISC"
          ],
          "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
        }
      ],
      "source": {
        "advisory": "ICSMA-21-273-01",
        "defect": [
          "CWE-284"
        ],
        "discovery": "EXTERNAL"
      },
      "title": "Improper Access Control for Boston Scientific Zoom Latitude",
      "workarounds": [
        {
          "lang": "en",
          "value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security,  the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."
        }
      ],
      "x_generator": {
        "engine": "Vulnogram 0.0.9"
      },
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "DATE_PUBLIC": "2021-09-30T21:02:00.000Z",
          "ID": "CVE-2021-38392",
          "STATE": "PUBLIC",
          "TITLE": "Improper Access Control for Boston Scientific Zoom Latitude"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "ZOOM LATITUDE",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "Model 3120"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Boston Scientific"
              }
            ]
          }
        },
        "credit": [
          {
            "lang": "eng",
            "value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster  University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus  Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices  and reported them to Boston Scientific."
          }
        ],
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world."
            }
          ]
        },
        "generator": {
          "engine": "Vulnogram 0.0.9"
        },
        "impact": {
          "cvss": {
            "attackComplexity": "LOW",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "LOW",
            "baseScore": 6.5,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "LOW",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "CHANGED",
            "userInteraction": "REQUIRED",
            "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L",
            "version": "3.1"
          }
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "CWE-284 Improper Access Control"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01",
              "refsource": "MISC",
              "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
            }
          ]
        },
        "source": {
          "advisory": "ICSMA-21-273-01",
          "defect": [
            "CWE-284"
          ],
          "discovery": "EXTERNAL"
        },
        "work_around": [
          {
            "lang": "en",
            "value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security,  the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."
          }
        ]
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2021-38392",
    "datePublished": "2021-10-04T17:35:13.879926Z",
    "dateReserved": "2021-08-10T00:00:00",
    "dateUpdated": "2024-09-17T03:33:08.540Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}

CVE-2021-38398 (GCVE-0-2021-38398)
Vulnerability from cvelistv5
Published
2021-10-04 17:35
Modified
2024-09-16 17:58
CWE
Summary
The affected device uses off-the-shelf software components that contain unpatched vulnerabilities. A malicious attacker with physical access to the affected device could exploit these vulnerabilities.
References
Impacted products
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-04T01:37:16.493Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "product": "ZOOM LATITUDE",
          "vendor": "Boston Scientific",
          "versions": [
            {
              "status": "affected",
              "version": "Model 3120"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster  University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus  Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices  and reported them to Boston Scientific."
        }
      ],
      "datePublic": "2021-09-30T00:00:00",
      "descriptions": [
        {
          "lang": "en",
          "value": "The affected device uses off-the-shelf software components that contain unpatched vulnerabilities. A malicious attacker with physical access to the affected device could exploit these vulnerabilities."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "LOW",
            "baseScore": 6.5,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "LOW",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "CHANGED",
            "userInteraction": "REQUIRED",
            "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L",
            "version": "3.1"
          }
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-1329",
              "description": "CWE-1329",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2021-10-04T17:35:06",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "tags": [
            "x_refsource_MISC"
          ],
          "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
        }
      ],
      "source": {
        "advisory": "ICSMA-21-273-01",
        "defect": [
          "CWE-1329"
        ],
        "discovery": "EXTERNAL"
      },
      "title": "Reliance on Component that is not Updateable for Boston Scientific Zoom Latitude",
      "workarounds": [
        {
          "lang": "en",
          "value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security,  the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."
        }
      ],
      "x_generator": {
        "engine": "Vulnogram 0.0.9"
      },
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "DATE_PUBLIC": "2021-09-30T21:02:00.000Z",
          "ID": "CVE-2021-38398",
          "STATE": "PUBLIC",
          "TITLE": "Reliance on Component that is not Updateable for Boston Scientific Zoom Latitude"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "ZOOM LATITUDE",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "Model 3120"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Boston Scientific"
              }
            ]
          }
        },
        "credit": [
          {
            "lang": "eng",
            "value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster  University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus  Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices  and reported them to Boston Scientific."
          }
        ],
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "The affected device uses off-the-shelf software components that contain unpatched vulnerabilities. A malicious attacker with physical access to the affected device could exploit these vulnerabilities."
            }
          ]
        },
        "generator": {
          "engine": "Vulnogram 0.0.9"
        },
        "impact": {
          "cvss": {
            "attackComplexity": "LOW",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "LOW",
            "baseScore": 6.5,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "LOW",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "CHANGED",
            "userInteraction": "REQUIRED",
            "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L",
            "version": "3.1"
          }
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "CWE-1329"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01",
              "refsource": "MISC",
              "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
            }
          ]
        },
        "source": {
          "advisory": "ICSMA-21-273-01",
          "defect": [
            "CWE-1329"
          ],
          "discovery": "EXTERNAL"
        },
        "work_around": [
          {
            "lang": "en",
            "value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security,  the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."
          }
        ]
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2021-38398",
    "datePublished": "2021-10-04T17:35:06.578644Z",
    "dateReserved": "2021-08-10T00:00:00",
    "dateUpdated": "2024-09-16T17:58:22.688Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1"
}