CVE-2021-23236 (GCVE-0-2021-23236)
Vulnerability from cvelistv5
Published
2022-01-21 18:17
Modified
2025-04-16 16:48
CWE
  • CWE-400 - Uncontrolled Resource Consumption
Summary
Requests may be used to interrupt the normal operation of the device. When exploited, Fresenius Kabi Agilia Link+ version 3.0 must be rebooted via a hard reset triggered by pressing a button on the rack system.
References
Impacted products
Vendor Product Version
Fresenius Kabi Agilia Link+ Version: unspecified   < 3.0
Create a notification for this product.
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-03T19:05:55.456Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01"
          }
        ],
        "title": "CVE Program Container"
      },
      {
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2021-23236",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "yes"
                  },
                  {
                    "Technical Impact": "partial"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2025-04-16T16:00:06.380083Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2025-04-16T16:48:05.390Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      }
    ],
    "cna": {
      "affected": [
        {
          "product": "Agilia Link+",
          "vendor": "Fresenius Kabi",
          "versions": [
            {
              "lessThan": "3.0",
              "status": "affected",
              "version": "unspecified",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "value": "Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project)."
        }
      ],
      "descriptions": [
        {
          "lang": "en",
          "value": "Requests may be used to interrupt the normal operation of the device. When exploited, Fresenius Kabi Agilia Link+ version 3.0 must be rebooted via a hard reset triggered by pressing a button on the rack system."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "NETWORK",
            "availabilityImpact": "HIGH",
            "baseScore": 7.5,
            "baseSeverity": "HIGH",
            "confidentialityImpact": "NONE",
            "integrityImpact": "NONE",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
            "version": "3.1"
          }
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-400",
              "description": "CWE-400 Uncontrolled Resource Consumption",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2022-01-21T18:17:35.000Z",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "tags": [
            "x_refsource_MISC"
          ],
          "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "value": "Fresenius Kabi has created new versions to address these vulnerabilities:\n\u2022\tLink+ v3.0 (D16 or later)\n\u2022\tVSS v1.0.3 (or later)\n\u2022\tAgilia Connect Pumps Wifi Module (D29 or later)\n\u2022\tAgilia Connect Partner v3.3.2 (or later) \nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \n"
        }
      ],
      "source": {
        "advisory": "ICSMA-21-355-01",
        "discovery": "EXTERNAL"
      },
      "title": "Fresenius Kabi Agilia Connect Infusion System uncontrolled resource consumption",
      "workarounds": [
        {
          "lang": "en",
          "value": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\u2022\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\n\u2022\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\n\u2022\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\n"
        }
      ],
      "x_generator": {
        "engine": "Vulnogram 0.0.9"
      },
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "ID": "CVE-2021-23236",
          "STATE": "PUBLIC",
          "TITLE": "Fresenius Kabi Agilia Connect Infusion System uncontrolled resource consumption"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "Agilia Link+",
                      "version": {
                        "version_data": [
                          {
                            "version_affected": "\u003c",
                            "version_value": "3.0"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Fresenius Kabi"
              }
            ]
          }
        },
        "credit": [
          {
            "lang": "eng",
            "value": "Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project)."
          }
        ],
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "Requests may be used to interrupt the normal operation of the device. When exploited, Fresenius Kabi Agilia Link+ version 3.0 must be rebooted via a hard reset triggered by pressing a button on the rack system."
            }
          ]
        },
        "generator": {
          "engine": "Vulnogram 0.0.9"
        },
        "impact": {
          "cvss": {
            "attackComplexity": "LOW",
            "attackVector": "NETWORK",
            "availabilityImpact": "HIGH",
            "baseScore": 7.5,
            "baseSeverity": "HIGH",
            "confidentialityImpact": "NONE",
            "integrityImpact": "NONE",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
            "version": "3.1"
          }
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "CWE-400 Uncontrolled Resource Consumption"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01",
              "refsource": "MISC",
              "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01"
            }
          ]
        },
        "solution": [
          {
            "lang": "en",
            "value": "Fresenius Kabi has created new versions to address these vulnerabilities:\n\u2022\tLink+ v3.0 (D16 or later)\n\u2022\tVSS v1.0.3 (or later)\n\u2022\tAgilia Connect Pumps Wifi Module (D29 or later)\n\u2022\tAgilia Connect Partner v3.3.2 (or later) \nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \n"
          }
        ],
        "source": {
          "advisory": "ICSMA-21-355-01",
          "discovery": "EXTERNAL"
        },
        "work_around": [
          {
            "lang": "en",
            "value": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\u2022\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\n\u2022\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\n\u2022\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\n"
          }
        ]
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2021-23236",
    "datePublished": "2022-01-21T18:17:35.000Z",
    "dateReserved": "2021-11-30T00:00:00.000Z",
    "dateUpdated": "2025-04-16T16:48:05.390Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1",
  "vulnerability-lookup:meta": {
    "nvd": "{\"cve\":{\"id\":\"CVE-2021-23236\",\"sourceIdentifier\":\"ics-cert@hq.dhs.gov\",\"published\":\"2022-01-21T19:15:08.177\",\"lastModified\":\"2024-11-21T05:51:25.647\",\"vulnStatus\":\"Modified\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"Requests may be used to interrupt the normal operation of the device. When exploited, Fresenius Kabi Agilia Link+ version 3.0 must be rebooted via a hard reset triggered by pressing a button on the rack system.\"},{\"lang\":\"es\",\"value\":\"Las peticiones pueden ser usadas para interrumpir el funcionamiento normal del dispositivo. Cuando es explotado, Fresenius Kabi Agilia Link+ versi\u00f3n 3.0, debe reiniciarse por medio de un hard reset que es desencadenado al pulsar un bot\u00f3n en el sistema de rack\"}],\"metrics\":{\"cvssMetricV31\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H\",\"baseScore\":7.5,\"baseSeverity\":\"HIGH\",\"attackVector\":\"NETWORK\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"NONE\",\"integrityImpact\":\"NONE\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":3.9,\"impactScore\":3.6},{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H\",\"baseScore\":7.5,\"baseSeverity\":\"HIGH\",\"attackVector\":\"NETWORK\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"NONE\",\"integrityImpact\":\"NONE\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":3.9,\"impactScore\":3.6}],\"cvssMetricV2\":[{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"2.0\",\"vectorString\":\"AV:N/AC:L/Au:N/C:N/I:N/A:C\",\"baseScore\":7.8,\"accessVector\":\"NETWORK\",\"accessComplexity\":\"LOW\",\"authentication\":\"NONE\",\"confidentialityImpact\":\"NONE\",\"integrityImpact\":\"NONE\",\"availabilityImpact\":\"COMPLETE\"},\"baseSeverity\":\"HIGH\",\"exploitabilityScore\":10.0,\"impactScore\":6.9,\"acInsufInfo\":false,\"obtainAllPrivilege\":false,\"obtainUserPrivilege\":false,\"obtainOtherPrivilege\":false,\"userInteractionRequired\":false}]},\"weaknesses\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-400\"}]},{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-400\"}]}],\"configurations\":[{\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:a:fresenius-kabi:agilia_partner_maintenance_software:*:*:*:*:*:*:*:*\",\"versionEndIncluding\":\"3.3.0\",\"matchCriteriaId\":\"3B072164-6AA2-4A14-B7D7-10B4B953004D\"},{\"vulnerable\":true,\"criteria\":\"cpe:2.3:a:fresenius-kabi:vigilant_centerium:1.0:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"C47210A7-4753-4ED7-8E6B-9BE8EBFABC9F\"},{\"vulnerable\":true,\"criteria\":\"cpe:2.3:a:fresenius-kabi:vigilant_insight:1.0:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"9C91B931-F726-4AB2-B3A6-D92F774CF04D\"},{\"vulnerable\":true,\"criteria\":\"cpe:2.3:a:fresenius-kabi:vigilant_mastermed:1.0:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"04AC7167-F5C8-46A2-B937-953E13D76A32\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:fresenius-kabi:agilia_connect_firmware:*:*:*:*:*:*:*:*\",\"versionEndIncluding\":\"d25\",\"matchCriteriaId\":\"8B4979F9-A7D5-4B5C-8FF2-C3C67773EE03\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:fresenius-kabi:agilia_connect:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"D1BBB63E-7E43-4BC1-A08F-4F1F811F839B\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:fresenius-kabi:link\\\\+_agilia_firmware:*:*:*:*:*:*:*:*\",\"versionEndExcluding\":\"3.0\",\"matchCriteriaId\":\"45FA28DE-939F-4146-A6E2-CE8849C9CB16\"},{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:fresenius-kabi:link\\\\+_agilia_firmware:3.0:-:*:*:*:*:*:*\",\"matchCriteriaId\":\"7D5FC3D0-9593-487B-B70A-F8BBCA8A18FF\"},{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:fresenius-kabi:link\\\\+_agilia_firmware:3.0:d15:*:*:*:*:*:*\",\"matchCriteriaId\":\"67E88F2E-C12B-4B50-B087-3247F4748AF3\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:fresenius-kabi:link\\\\+_agilia:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"1548AA3F-659F-43C3-9261-C7FD55465877\"}]}]}],\"references\":[{\"url\":\"https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01\",\"source\":\"ics-cert@hq.dhs.gov\",\"tags\":[\"Third Party Advisory\",\"US Government Resource\"]},{\"url\":\"https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01\",\"source\":\"af854a3a-2127-422b-91ae-364da2661108\",\"tags\":[\"Third Party Advisory\",\"US Government Resource\"]}]}}",
    "vulnrichment": {
      "containers": "{\"adp\": [{\"title\": \"CVE Program Container\", \"references\": [{\"url\": \"https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01\", \"tags\": [\"x_refsource_MISC\", \"x_transferred\"]}], \"providerMetadata\": {\"orgId\": \"af854a3a-2127-422b-91ae-364da2661108\", \"shortName\": \"CVE\", \"dateUpdated\": \"2024-08-03T19:05:55.456Z\"}}, {\"title\": \"CISA ADP Vulnrichment\", \"metrics\": [{\"other\": {\"type\": \"ssvc\", \"content\": {\"id\": \"CVE-2021-23236\", \"role\": \"CISA Coordinator\", \"options\": [{\"Exploitation\": \"none\"}, {\"Automatable\": \"yes\"}, {\"Technical Impact\": \"partial\"}], \"version\": \"2.0.3\", \"timestamp\": \"2025-04-16T16:00:06.380083Z\"}}}], \"providerMetadata\": {\"orgId\": \"134c704f-9b21-4f2e-91b3-4a467353bcc0\", \"shortName\": \"CISA-ADP\", \"dateUpdated\": \"2025-04-16T16:00:08.004Z\"}}], \"cna\": {\"title\": \"Fresenius Kabi Agilia Connect Infusion System uncontrolled resource consumption\", \"source\": {\"advisory\": \"ICSMA-21-355-01\", \"discovery\": \"EXTERNAL\"}, \"credits\": [{\"lang\": \"en\", \"value\": \"Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project).\"}], \"metrics\": [{\"cvssV3_1\": {\"scope\": \"UNCHANGED\", \"version\": \"3.1\", \"baseScore\": 7.5, \"attackVector\": \"NETWORK\", \"baseSeverity\": \"HIGH\", \"vectorString\": \"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H\", \"integrityImpact\": \"NONE\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"LOW\", \"availabilityImpact\": \"HIGH\", \"privilegesRequired\": \"NONE\", \"confidentialityImpact\": \"NONE\"}}], \"affected\": [{\"vendor\": \"Fresenius Kabi\", \"product\": \"Agilia Link+\", \"versions\": [{\"status\": \"affected\", \"version\": \"unspecified\", \"lessThan\": \"3.0\", \"versionType\": \"custom\"}]}], \"solutions\": [{\"lang\": \"en\", \"value\": \"Fresenius Kabi has created new versions to address these vulnerabilities:\\n\\u2022\\tLink+ v3.0 (D16 or later)\\n\\u2022\\tVSS v1.0.3 (or later)\\n\\u2022\\tAgilia Connect Pumps Wifi Module (D29 or later)\\n\\u2022\\tAgilia Connect Partner v3.3.2 (or later) \\nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \\n\"}], \"references\": [{\"url\": \"https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01\", \"tags\": [\"x_refsource_MISC\"]}], \"workarounds\": [{\"lang\": \"en\", \"value\": \"CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\\n\\u2022\\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\\n\\u2022\\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\\n\\u2022\\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\\n\"}], \"x_generator\": {\"engine\": \"Vulnogram 0.0.9\"}, \"descriptions\": [{\"lang\": \"en\", \"value\": \"Requests may be used to interrupt the normal operation of the device. When exploited, Fresenius Kabi Agilia Link+ version 3.0 must be rebooted via a hard reset triggered by pressing a button on the rack system.\"}], \"problemTypes\": [{\"descriptions\": [{\"lang\": \"en\", \"type\": \"CWE\", \"cweId\": \"CWE-400\", \"description\": \"CWE-400 Uncontrolled Resource Consumption\"}]}], \"providerMetadata\": {\"orgId\": \"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6\", \"shortName\": \"icscert\", \"dateUpdated\": \"2022-01-21T18:17:35.000Z\"}, \"x_legacyV4Record\": {\"credit\": [{\"lang\": \"eng\", \"value\": \"Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project).\"}], \"impact\": {\"cvss\": {\"scope\": \"UNCHANGED\", \"version\": \"3.1\", \"baseScore\": 7.5, \"attackVector\": \"NETWORK\", \"baseSeverity\": \"HIGH\", \"vectorString\": \"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H\", \"integrityImpact\": \"NONE\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"LOW\", \"availabilityImpact\": \"HIGH\", \"privilegesRequired\": \"NONE\", \"confidentialityImpact\": \"NONE\"}}, \"source\": {\"advisory\": \"ICSMA-21-355-01\", \"discovery\": \"EXTERNAL\"}, \"affects\": {\"vendor\": {\"vendor_data\": [{\"product\": {\"product_data\": [{\"version\": {\"version_data\": [{\"version_value\": \"3.0\", \"version_affected\": \"\u003c\"}]}, \"product_name\": \"Agilia Link+\"}]}, \"vendor_name\": \"Fresenius Kabi\"}]}}, \"solution\": [{\"lang\": \"en\", \"value\": \"Fresenius Kabi has created new versions to address these vulnerabilities:\\n\\u2022\\tLink+ v3.0 (D16 or later)\\n\\u2022\\tVSS v1.0.3 (or later)\\n\\u2022\\tAgilia Connect Pumps Wifi Module (D29 or later)\\n\\u2022\\tAgilia Connect Partner v3.3.2 (or later) \\nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \\n\"}], \"data_type\": \"CVE\", \"generator\": {\"engine\": \"Vulnogram 0.0.9\"}, \"references\": {\"reference_data\": [{\"url\": \"https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01\", \"name\": \"https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01\", \"refsource\": \"MISC\"}]}, \"data_format\": \"MITRE\", \"description\": {\"description_data\": [{\"lang\": \"eng\", \"value\": \"Requests may be used to interrupt the normal operation of the device. When exploited, Fresenius Kabi Agilia Link+ version 3.0 must be rebooted via a hard reset triggered by pressing a button on the rack system.\"}]}, \"problemtype\": {\"problemtype_data\": [{\"description\": [{\"lang\": \"eng\", \"value\": \"CWE-400 Uncontrolled Resource Consumption\"}]}]}, \"work_around\": [{\"lang\": \"en\", \"value\": \"CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\\n\\u2022\\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\\n\\u2022\\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\\n\\u2022\\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\\n\"}], \"data_version\": \"4.0\", \"CVE_data_meta\": {\"ID\": \"CVE-2021-23236\", \"STATE\": \"PUBLIC\", \"TITLE\": \"Fresenius Kabi Agilia Connect Infusion System uncontrolled resource consumption\", \"ASSIGNER\": \"ics-cert@hq.dhs.gov\"}}}}",
      "cveMetadata": "{\"cveId\": \"CVE-2021-23236\", \"state\": \"PUBLISHED\", \"dateUpdated\": \"2025-04-16T16:48:05.390Z\", \"dateReserved\": \"2021-11-30T00:00:00.000Z\", \"assignerOrgId\": \"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6\", \"datePublished\": \"2022-01-21T18:17:35.000Z\", \"assignerShortName\": \"icscert\"}",
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }
  }
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
  • Confirmed: The vulnerability is confirmed from an analyst perspective.
  • Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
  • Patched: This vulnerability was successfully patched by the user reporting the sighting.
  • Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
  • Not confirmed: The user expresses doubt about the veracity of the vulnerability.
  • Not patched: This vulnerability was not successfully patched by the user reporting the sighting.


Loading…

Loading…