Summary
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
References
Impacted products
Vendor Product Version
Medtronic Conexus Radio Frequency Telemetry Protocol Version: All versions
Create a notification for this product.
   Medtronic MyCareLink Monitor Version: 24950
Version: 24952
Create a notification for this product.
   Medtronic CareLink Monitor Version: 2490C
Create a notification for this product.
   Medtronic CareLink 2090 Programmer Version: All versions
Create a notification for this product.
   Medtronic Amplia CRT-D Version: All versions
Create a notification for this product.
   Medtronic Claria CRT-D Version: All versions
Create a notification for this product.
   Medtronic Compia CRT-D Version: All versions
Create a notification for this product.
   Medtronic Concerto CRT-D Version: All versions
Create a notification for this product.
   Medtronic Concerto II CRT-D Version: All versions
Create a notification for this product.
   Medtronic Consulta CRT-D Version: All versions
Create a notification for this product.
   Medtronic Evera ICD Version: All versions
Create a notification for this product.
   Medtronic Maximo II CRT-D Version: All versions
Create a notification for this product.
   Medtronic Maximo II ICD Version: All versions
Create a notification for this product.
   Medtronic Mirro ICD Version: All versions
Create a notification for this product.
   Medtronic Nayamed ND ICD Version: All versions
Create a notification for this product.
   Medtronic Primo ICD Version: All versions
Create a notification for this product.
   Medtronic Protecta ICD, Protecta CRT-D Version: All versions
Create a notification for this product.
   Medtronic Secura ICD Version: All versions
Create a notification for this product.
   Medtronic Virtuoso ICD Version: All versions
Create a notification for this product.
   Medtronic Virtuoso II ICD Version: All versions
Create a notification for this product.
   Medtronic Visia AF ICD Version: All versions
Create a notification for this product.
   Medtronic Viva CRT-D Version: All versions
Create a notification for this product.
   Medtronic Brava CRT-D Version: All versions
Create a notification for this product.
   Medtronic Mirro MRI ICD Version: All versions
Create a notification for this product.
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-04T20:23:21.992Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "name": "107544",
            "tags": [
              "vdb-entry",
              "x_refsource_BID",
              "x_transferred"
            ],
            "url": "http://www.securityfocus.com/bid/107544"
          },
          {
            "tags": [
              "x_refsource_CONFIRM",
              "x_transferred"
            ],
            "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "Conexus Radio Frequency Telemetry Protocol",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "MyCareLink Monitor",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "24950"
            },
            {
              "status": "affected",
              "version": "24952"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "CareLink Monitor",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "2490C"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "CareLink 2090 Programmer",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Amplia CRT-D",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Claria CRT-D",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Compia CRT-D",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Concerto CRT-D",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Concerto II CRT-D",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Consulta CRT-D",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Evera ICD",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Maximo II CRT-D",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Maximo II ICD",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Mirro ICD",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Nayamed ND ICD",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Primo ICD",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Protecta ICD, Protecta CRT-D",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Secura ICD",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Virtuoso ICD",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Virtuoso II ICD",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Visia AF ICD",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Viva CRT-D",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Brava CRT-D",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Mirro MRI ICD",
          "vendor": "Medtronic",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Peter Morgan of Clever Security; Dave Singel\u00e9e and Bart Preneel of KU Leuven; Eduard Marin formerly of KU Leuven, currently with University of Birmingham; Flavio D. Garcia; Tom Chothia of the University of Birmingham; and Rik Willems of University Hospital Gasthuisberg Leuven reported these vulnerabilities to CISA."
        }
      ],
      "datePublic": "2019-03-21T06:00:00.000Z",
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003e\nThe Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product\u2019s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.\n\n\u003c/p\u003e"
            }
          ],
          "value": "The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product\u2019s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "ADJACENT_NETWORK",
            "availabilityImpact": "HIGH",
            "baseScore": 9.3,
            "baseSeverity": "CRITICAL",
            "confidentialityImpact": "NONE",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "CHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-284",
              "description": "CWE-284",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2025-05-22T19:26:15.126Z",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "tags": [
            "x_refsource_MISC"
          ],
          "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01"
        },
        {
          "name": "107544",
          "tags": [
            "vdb-entry",
            "x_refsource_BID"
          ],
          "url": "http://www.securityfocus.com/bid/107544"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eMedtronic has developed mitigating patches for a subset of the \naffected implanted cardiac device models. These patches are installed \nduring regular office visits. Medtronic has stated that patches for \nadditional impacted models are being developed by Medtronic and will be \ndeployed through future updates. Patches are currently available for the\n following affected models:\u003cbr\u003e\u003c/p\u003e\u003cul\u003e\u003cli\u003eProtecta CRT-D and implantable cardioverter defibrillators (ICDs), all models\u003c/li\u003e\u003c/ul\u003e\u003cul\u003e\u003cli\u003eAmplia MRI CRT-D, all models (patch available in U.S. only)\u003c/li\u003e\u003cli\u003eClaria MRI CRT-D, all models (patch available in U.S. only)\u003c/li\u003e\u003cli\u003eCompia MRI CRT-D, all models (patch available in U.S. only)\u003c/li\u003e\u003cli\u003eVisia AF MRI ICD, all models\u003c/li\u003e\u003cli\u003eVisia AF ICD, all models \u003c/li\u003e\u003cli\u003eBrava CRT-D, all models \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp; \u0026nbsp;\u003c/li\u003e\u003cli\u003eEvera MRI ICD, all models\u003c/li\u003e\u003cli\u003eEvera ICD, all models\u003c/li\u003e\u003cli\u003eMirro MRI ICD, all models\u003c/li\u003e\u003cli\u003ePrimo MRI ICD, all models\u003c/li\u003e\u003cli\u003eViva CRT-D, all models\u003c/li\u003e\u003c/ul\u003e\n\n\u003cbr\u003e"
            }
          ],
          "value": "Medtronic has developed mitigating patches for a subset of the \naffected implanted cardiac device models. These patches are installed \nduring regular office visits. Medtronic has stated that patches for \nadditional impacted models are being developed by Medtronic and will be \ndeployed through future updates. Patches are currently available for the\n following affected models:\n\n\n  *  Protecta CRT-D and implantable cardioverter defibrillators (ICDs), all models\n\n\n  *  Amplia MRI CRT-D, all models (patch available in U.S. only)\n  *  Claria MRI CRT-D, all models (patch available in U.S. only)\n  *  Compia MRI CRT-D, all models (patch available in U.S. only)\n  *  Visia AF MRI ICD, all models\n  *  Visia AF ICD, all models \n  *  Brava CRT-D, all models \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0 \u00a0\n  *  Evera MRI ICD, all models\n  *  Evera ICD, all models\n  *  Mirro MRI ICD, all models\n  *  Primo MRI ICD, all models\n  *  Viva CRT-D, all models"
        }
      ],
      "source": {
        "advisory": "ICSMA-19-080-01",
        "discovery": "EXTERNAL"
      },
      "title": "Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control",
      "workarounds": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eMedtronic has applied additional controls for monitoring and responding to improper use of the Conexus telemetry protocol by the affected implanted cardiac devices. Additional mitigations are being developed and will be deployed through future updates, assuming regulatory approval.\u003c/p\u003e\u003cp\u003eMedtronic recommends that users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\u003c/p\u003e\u003cul\u003e\u003cli\u003eMaintain good physical control over home monitors and programmers.\u003c/li\u003e\u003cli\u003eUse only home monitors, programmers, and implantable devices obtained directly from your healthcare provider or a Medtronic representative to ensure integrity of the system.\u003c/li\u003e\u003cli\u003eDo not connect unapproved devices to home monitors and programmers through USB ports or other physical connections.\u003c/li\u003e\u003cli\u003eOnly use programmers to connect and interact with implanted devices in physically controlled hospital and clinical environments.\u003c/li\u003e\u003cli\u003eOnly use home monitors in private environments such as a home, apartment, or otherwise physically controlled environment.\u003c/li\u003e\u003cli\u003eReport any concerning behavior regarding these products to your healthcare provider or a Medtronic representative.\u003c/li\u003e\u003c/ul\u003e\u003cp\u003eMedtronic has released additional \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/security\"\u003epatient focused information\u003c/a\u003e.\u003c/p\u003e\n\n\u003cbr\u003e"
            }
          ],
          "value": "Medtronic has applied additional controls for monitoring and responding to improper use of the Conexus telemetry protocol by the affected implanted cardiac devices. Additional mitigations are being developed and will be deployed through future updates, assuming regulatory approval.\n\nMedtronic recommends that users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\n  *  Maintain good physical control over home monitors and programmers.\n  *  Use only home monitors, programmers, and implantable devices obtained directly from your healthcare provider or a Medtronic representative to ensure integrity of the system.\n  *  Do not connect unapproved devices to home monitors and programmers through USB ports or other physical connections.\n  *  Only use programmers to connect and interact with implanted devices in physically controlled hospital and clinical environments.\n  *  Only use home monitors in private environments such as a home, apartment, or otherwise physically controlled environment.\n  *  Report any concerning behavior regarding these products to your healthcare provider or a Medtronic representative.\n\n\nMedtronic has released additional  patient focused information https://www.medtronic.com/security ."
        }
      ],
      "x_generator": {
        "engine": "Vulnogram 0.2.0"
      },
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "ID": "CVE-2019-6540",
          "STATE": "PUBLIC"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "Conexus Radio Frequency Telemetry Protocol",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "MyCareLink Monitor",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "24950"
                          },
                          {
                            "version_value": "24952"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "CareLink Monitor",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "2490C"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "CareLink 2090 Programmer",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Amplia CRT-D",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Claria CRT-D",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Compia CRT-D",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Concerto CRT-D",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Concerto II CRT-D",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Consulta CRT-D",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Evera ICD",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Maximo II CRT-D",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Maximo II ICD",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Mirro ICD",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Nayamed ND ICD",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Primo ICD",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Protecta ICD, Protecta CRT-D",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Secura ICD",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Virtuoso ICD",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Virtuoso II ICD",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Visia AF ICD",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Viva CRT-D",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "n/a"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Medtronic"
              }
            ]
          }
        },
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data."
            }
          ]
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "Cleartext transmission of sensitive information CWE-319"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01",
              "refsource": "MISC",
              "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01"
            },
            {
              "name": "107544",
              "refsource": "BID",
              "url": "http://www.securityfocus.com/bid/107544"
            }
          ]
        }
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2019-6538",
    "datePublished": "2019-03-25T21:26:03",
    "dateReserved": "2019-01-22T00:00:00",
    "dateUpdated": "2025-05-22T19:26:15.126Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1",
  "vulnerability-lookup:meta": {
    "nvd": "{\"cve\":{\"id\":\"CVE-2019-6538\",\"sourceIdentifier\":\"ics-cert@hq.dhs.gov\",\"published\":\"2019-03-25T22:29:00.763\",\"lastModified\":\"2025-05-22T20:15:20.660\",\"vulnStatus\":\"Modified\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product\u2019s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.\"},{\"lang\":\"es\",\"value\":\"El protocolo de telemetr\u00eda Conexus utilizado en las versiones 24950 y 24952 de Medtronic MyCareLink Monitor, CareLink Monitor versi\u00f3n 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D e ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD y CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD y Viva CRT- D no implementa autenticaci\u00f3n o autorizaci\u00f3n. Un atacante con acceso adyacente de corto alcance a un producto afectado, en situaciones donde la radio del producto est\u00e1 encendida, puede inyectar, reproducir, modificar o interceptar datos dentro de la comunicaci\u00f3n de telemetr\u00eda. Este protocolo de comunicaci\u00f3n proporciona la capacidad de leer y escribir valores de memoria en dispositivos card\u00edacos implantados afectados; por lo tanto, un atacante podr\u00eda explotar este protocolo de comunicaci\u00f3n para cambiar la memoria en el dispositivo card\u00edaco implantado.\"}],\"metrics\":{\"cvssMetricV31\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H\",\"baseScore\":9.3,\"baseSeverity\":\"CRITICAL\",\"attackVector\":\"ADJACENT_NETWORK\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"CHANGED\",\"confidentialityImpact\":\"NONE\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":2.8,\"impactScore\":5.8},{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N\",\"baseScore\":6.5,\"baseSeverity\":\"MEDIUM\",\"attackVector\":\"ADJACENT_NETWORK\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"NONE\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"NONE\"},\"exploitabilityScore\":2.8,\"impactScore\":3.6}],\"cvssMetricV2\":[{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"2.0\",\"vectorString\":\"AV:A/AC:L/Au:N/C:N/I:P/A:N\",\"baseScore\":3.3,\"accessVector\":\"ADJACENT_NETWORK\",\"accessComplexity\":\"LOW\",\"authentication\":\"NONE\",\"confidentialityImpact\":\"NONE\",\"integrityImpact\":\"PARTIAL\",\"availabilityImpact\":\"NONE\"},\"baseSeverity\":\"LOW\",\"exploitabilityScore\":6.5,\"impactScore\":2.9,\"acInsufInfo\":false,\"obtainAllPrivilege\":false,\"obtainUserPrivilege\":false,\"obtainOtherPrivilege\":false,\"userInteractionRequired\":false}]},\"weaknesses\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-284\"}]},{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-306\"},{\"lang\":\"en\",\"value\":\"CWE-862\"}]}],\"configurations\":[{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:mycarelink_monitor_firmware:24950:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"1683D2B0-A4AA-4035-8E30-4641F399BF77\"},{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:mycarelink_monitor_firmware:24952:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"E9F02305-2B0B-4F12-99FD-986ADB46B845\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:mycarelink_monitor:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"F8588B5D-10B2-4AD3-B550-2AD7C252BAB4\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:carelink_monitor_firmware:2490c:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"9822CAEC-D72E-4B49-98A0-734E35D25C3B\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:carelink_monitor:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"80CF9694-488A-4829-98B2-1C40136C416E\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:carelink_2090_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"8FE6438F-F84A-4876-B13E-4C06050EF7D4\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:carelink_2090:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"CA96FB28-A2B2-4144-87F2-19D0EEB3CBB2\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:amplia_crt-d_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"53F40F63-1FFD-4F1D-BB91-A0109CCE62A3\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:amplia_crt-d:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"E64609AE-EA44-423A-B0E1-5FBF219165C9\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:claria_crt-d_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"A9F3C231-C8D2-40A1-BA0D-8381208AD3B5\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:claria_crt-d:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"756E8B59-2D00-4647-8DCE-A293626FFFDD\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:compia_crt-d_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"832EC0E7-9AD0-42E9-9663-F9033763ADF1\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:compia_crt-d:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"41A7F4D2-233C-4BF7-9054-C814CE5110DF\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:concerto_crt-d_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"008FD034-1206-41DA-9537-1F4244236286\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:concerto_crt-d:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"94A17D5C-5BC8-4066-8C2C-7AB6CA25B4F0\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:concerto_ii_crt-d_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"7FA48969-0F33-434D-B210-B18D0459316C\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:concerto_ii_crt-d:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"77FD6FAE-32FD-48DD-8A95-881C9373F9EE\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:consulta_crt-d_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"D02CD1C6-9229-408A-8E6E-CAB394DC489A\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:consulta_crt-d:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"F6347AAA-ACB2-4CDF-B3F5-49CA8BB17577\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:evera_icd_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"64C40F66-1FA6-4DC6-BA87-4D7811650F81\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:evera_icd:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"DEBB88DD-94C3-478D-8AFE-BC63BA73E94C\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:maximo_ii_crt-d_and_lcd_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"283D42FD-D107-4340-85F3-B85F66CA9AA4\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:maximo_ii_crt-d_and_lcd:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"468E068D-8598-4D42-A483-7D1040665C4C\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:mirro_icd_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"FBAA5C99-D5AA-4377-86A0-48862582F769\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:mirro_icd:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"1525A59B-E8F7-43FD-8ECD-5546C4E268F6\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:nayamed_nd_icd_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"F92F9AA6-56B1-4969-AADD-0D181F1F26F5\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:nayamed_nd_icd:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"F6BA1499-CB4E-4067-984E-048CADDB5473\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:primo_icd_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"374D02EF-E15A-4C1A-BB1B-770F3FB75EC7\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:primo_icd:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"1941334A-A540-4ED3-8DCF-9298DC18D67E\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:protecta_icd_and_crt-d_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"CE78A2F7-1A2F-488C-A129-41F6D0A7C648\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:protecta_icd_and_crt-d:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"A12FFDDC-10A2-4C96-9256-D168C1E3EB5E\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:secura_icd_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"348706DF-2F3D-4875-9450-218A2A4E9210\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:secura_icd:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"A82FFC9C-960C-4709-AB17-0D6C7BC0E39E\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:virtuoso_icd_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"85BD02A6-B421-4E90-A91E-D966F3CE70A8\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:virtuoso_icd:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"618D26D4-6690-4F17-B9F5-2CD05486EB65\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:virtuoso_ii_icd_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"F9B06688-4AF7-4354-A6AF-B335424F1D51\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:virtuoso_ii_icd:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"C2F9A58D-D3E8-4E57-93D9-A1FB1BEB65D6\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:visia_af_icd_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"5D21AD86-5D91-4279-8A54-9A6E6A1B6960\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:visia_af_icd:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"17418686-EC2F-4FDB-88F8-2B4C60A8B48D\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:medtronic:viva_crt-d_firmware:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"468772A4-BCD2-4884-A7C0-4CD49DFC6C1E\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:medtronic:viva_crt-d:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"7222E296-5DA5-4611-BA1C-85059D7918A0\"}]}]}],\"references\":[{\"url\":\"http://www.securityfocus.com/bid/107544\",\"source\":\"ics-cert@hq.dhs.gov\",\"tags\":[\"Third Party Advisory\",\"VDB Entry\"]},{\"url\":\"https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01\",\"source\":\"ics-cert@hq.dhs.gov\",\"tags\":[\"Mitigation\",\"Third Party Advisory\",\"US Government Resource\"]},{\"url\":\"http://www.securityfocus.com/bid/107544\",\"source\":\"af854a3a-2127-422b-91ae-364da2661108\",\"tags\":[\"Third Party Advisory\",\"VDB Entry\"]},{\"url\":\"https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01\",\"source\":\"af854a3a-2127-422b-91ae-364da2661108\",\"tags\":[\"Mitigation\",\"Third Party Advisory\",\"US Government Resource\"]}]}}"
  }
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
  • Confirmed: The vulnerability is confirmed from an analyst perspective.
  • Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
  • Patched: This vulnerability was successfully patched by the user reporting the sighting.
  • Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
  • Not confirmed: The user expresses doubt about the veracity of the vulnerability.
  • Not patched: This vulnerability was not successfully patched by the user reporting the sighting.


Loading…

Loading…