fkie_cve-2014-5434
Vulnerability from fkie_nvd
Published
2019-03-26 15:29
Modified
2024-11-21 02:12
Severity ?
Summary
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
References
▶ | URL | Tags | |
---|---|---|---|
ics-cert@hq.dhs.gov | https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01 | Mitigation, Third Party Advisory, US Government Resource | |
af854a3a-2127-422b-91ae-364da2661108 | https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01 | Mitigation, Third Party Advisory, US Government Resource |
Impacted products
Vendor | Product | Version | |
---|---|---|---|
baxter | sigma_spectrum_infusion_system_firmware | 6.05 | |
baxter | sigma_spectrum_infusion_system | - | |
baxter | wireless_battery_module | 16 |
{ "configurations": [ { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:6.05:*:*:*:*:*:*:*", "matchCriteriaId": "E6B02D87-6C75-4D60-8D4C-84628757214F", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:*", "matchCriteriaId": "49E25260-EC14-4E98-A86B-CBBE47E26AE5", "vulnerable": false }, { "criteria": "cpe:2.3:h:baxter:wireless_battery_module:16:*:*:*:*:*:*:*", "matchCriteriaId": "175BDC61-5163-4D78-BF02-6B9EF5D38841", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" } ], "cveTags": [], "descriptions": [ { "lang": "en", "value": "Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes." }, { "lang": "es", "value": "Baxter SIGMA Spectrum Infusion System 6.05 (modelo 35700BAX), con un m\u00f3dulo de bater\u00eda inal\u00e1mbrica (WBM), en su versi\u00f3n 16, tiene una cuenta por defecto con credenciales embebidas empleadas con el protocolo FTP. Baxter afirma que no se pueden transferir archivos desde o hasta el WBM mediante esta cuenta. Baxter ha publicado una nueva versi\u00f3n de SIGMA Spectrum Infusion System, la 8, que incluye cambios en el software y el hardware." } ], "id": "CVE-2014-5434", "lastModified": "2024-11-21T02:12:02.537", "metrics": { "cvssMetricV2": [ { "acInsufInfo": false, "baseSeverity": "MEDIUM", "cvssData": { "accessComplexity": "LOW", "accessVector": "NETWORK", "authentication": "NONE", "availabilityImpact": "NONE", "baseScore": 5.0, "confidentialityImpact": "PARTIAL", "integrityImpact": "NONE", "vectorString": "AV:N/AC:L/Au:N/C:P/I:N/A:N", "version": "2.0" }, "exploitabilityScore": 10.0, "impactScore": 2.9, "obtainAllPrivilege": false, "obtainOtherPrivilege": false, "obtainUserPrivilege": false, "source": "nvd@nist.gov", "type": "Primary", "userInteractionRequired": false } ], "cvssMetricV30": [ { "cvssData": { "attackComplexity": "LOW", "attackVector": "NETWORK", "availabilityImpact": "HIGH", "baseScore": 9.8, "baseSeverity": "CRITICAL", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H", "version": "3.0" }, "exploitabilityScore": 3.9, "impactScore": 5.9, "source": "nvd@nist.gov", "type": "Primary" } ] }, "published": "2019-03-26T15:29:00.287", "references": [ { "source": "ics-cert@hq.dhs.gov", "tags": [ "Mitigation", "Third Party Advisory", "US Government Resource" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Mitigation", "Third Party Advisory", "US Government Resource" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01" } ], "sourceIdentifier": "ics-cert@hq.dhs.gov", "vulnStatus": "Modified", "weaknesses": [ { "description": [ { "lang": "en", "value": "CWE-259" } ], "source": "ics-cert@hq.dhs.gov", "type": "Secondary" }, { "description": [ { "lang": "en", "value": "CWE-798" } ], "source": "nvd@nist.gov", "type": "Primary" } ] }
Loading…
Loading…
Sightings
Author | Source | Type | Date |
---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
- Confirmed: The vulnerability is confirmed from an analyst perspective.
- Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
- Patched: This vulnerability was successfully patched by the user reporting the sighting.
- Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
- Not confirmed: The user expresses doubt about the veracity of the vulnerability.
- Not patched: This vulnerability was not successfully patched by the user reporting the sighting.
Loading…
Loading…