fkie_cve-2018-10596
Vulnerability from fkie_nvd
Published
2018-07-03 01:29
Modified
2025-05-22 18:15
Severity ?
Summary
Medtronic 2090 CareLink Programmer
uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.
References
Impacted products
Vendor | Product | Version | |
---|---|---|---|
medtronic | 2090_carelink_programmer_firmware | * | |
medtronic | 2090_carelink_programmer | - |
{ "configurations": [ { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:medtronic:2090_carelink_programmer_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "884B08DC-B6F9-4B34-9679-61C256437DF9", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:medtronic:2090_carelink_programmer:-:*:*:*:*:*:*:*", "matchCriteriaId": "215ED381-B433-4D05-97AD-4E7287E5820D", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" } ], "cveTags": [], "descriptions": [ { "lang": "en", "value": "Medtronic 2090 CareLink Programmer \n\nuses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications." }, { "lang": "es", "value": "Medtronic 2090 CareLink Programmer en todas sus versiones emplea una conexi\u00f3n de red privada virtual para descargar actualizaciones de forma segura. El producto no verifica si sigue conectado a esta red virtual privada antes de descargar actualizaciones. Un atacante con acceso de red local al programador podr\u00eda influenciar estas comunicaciones." } ], "id": "CVE-2018-10596", "lastModified": "2025-05-22T18:15:21.490", "metrics": { "cvssMetricV2": [ { "acInsufInfo": false, "baseSeverity": "MEDIUM", "cvssData": { "accessComplexity": "LOW", "accessVector": "ADJACENT_NETWORK", "authentication": "SINGLE", "availabilityImpact": "PARTIAL", "baseScore": 5.2, "confidentialityImpact": "PARTIAL", "integrityImpact": "PARTIAL", "vectorString": "AV:A/AC:L/Au:S/C:P/I:P/A:P", "version": "2.0" }, "exploitabilityScore": 5.1, "impactScore": 6.4, "obtainAllPrivilege": false, "obtainOtherPrivilege": false, "obtainUserPrivilege": false, "source": "nvd@nist.gov", "type": "Primary", "userInteractionRequired": false } ], "cvssMetricV30": [ { "cvssData": { "attackComplexity": "LOW", "attackVector": "ADJACENT_NETWORK", "availabilityImpact": "HIGH", "baseScore": 8.0, "baseSeverity": "HIGH", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "privilegesRequired": "LOW", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.0/AV:A/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H", "version": "3.0" }, "exploitabilityScore": 2.1, "impactScore": 5.9, "source": "nvd@nist.gov", "type": "Primary" } ], "cvssMetricV31": [ { "cvssData": { "attackComplexity": "HIGH", "attackVector": "NETWORK", "availabilityImpact": "HIGH", "baseScore": 7.1, "baseSeverity": "HIGH", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "privilegesRequired": "LOW", "scope": "UNCHANGED", "userInteraction": "REQUIRED", "vectorString": "CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H", "version": "3.1" }, "exploitabilityScore": 1.2, "impactScore": 5.9, "source": "ics-cert@hq.dhs.gov", "type": "Secondary" } ] }, "published": "2018-07-03T01:29:00.487", "references": [ { "source": "ics-cert@hq.dhs.gov", "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-2090-29901.html" }, { "source": "ics-cert@hq.dhs.gov", "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-058-01" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01" } ], "sourceIdentifier": "ics-cert@hq.dhs.gov", "vulnStatus": "Modified", "weaknesses": [ { "description": [ { "lang": "en", "value": "CWE-923" } ], "source": "ics-cert@hq.dhs.gov", "type": "Secondary" }, { "description": [ { "lang": "en", "value": "CWE-200" } ], "source": "nvd@nist.gov", "type": "Primary" } ] }
Loading…
Loading…
Sightings
Author | Source | Type | Date |
---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
- Confirmed: The vulnerability is confirmed from an analyst perspective.
- Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
- Patched: This vulnerability was successfully patched by the user reporting the sighting.
- Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
- Not confirmed: The user expresses doubt about the veracity of the vulnerability.
- Not patched: This vulnerability was not successfully patched by the user reporting the sighting.
Loading…
Loading…