fkie_cve-2018-14781
Vulnerability from fkie_nvd
Published
2018-08-13 21:48
Modified
2025-05-22 17:15
Severity ?
Summary
Medtronic MiniMed MMT
devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
References
▶ | URL | Tags | |
---|---|---|---|
ics-cert@hq.dhs.gov | http://www.securityfocus.com/bid/105044 | Third Party Advisory, VDB Entry | |
ics-cert@hq.dhs.gov | https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed.html | ||
ics-cert@hq.dhs.gov | https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02 | Third Party Advisory, US Government Resource | |
af854a3a-2127-422b-91ae-364da2661108 | http://www.securityfocus.com/bid/105044 | Third Party Advisory, VDB Entry | |
af854a3a-2127-422b-91ae-364da2661108 | https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02 | Third Party Advisory, US Government Resource |
Impacted products
{ "configurations": [ { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:medtronicdiabetes:508_minimed_insulin_pump_firmware:-:*:*:*:*:*:*:*", "matchCriteriaId": "EF9C0D70-7924-414E-98EF-245E52ED8838", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:medtronicdiabetes:508_minimed_insulin_pump:-:*:*:*:*:*:*:*", "matchCriteriaId": "FF20B7E1-C7E3-45E6-A981-6A02807A2411", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:medtronicdiabetes:522_paradigm_real-time_firmware:-:*:*:*:*:*:*:*", "matchCriteriaId": "3A8E9BE8-E17F-490D-BCE7-92D43BE14574", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:medtronicdiabetes:522_paradigm_real-time:-:*:*:*:*:*:*:*", "matchCriteriaId": "CCD29F70-6322-4891-8036-6BDE7F98B08A", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:medtronicdiabetes:722_paradigm_real-time_firmware:-:*:*:*:*:*:*:*", "matchCriteriaId": "95CAA730-1862-4D91-B962-28A2D8EB07F2", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:medtronicdiabetes:722_paradigm_real-time:-:*:*:*:*:*:*:*", "matchCriteriaId": "AB82801B-A8AA-47C9-BC9E-A4FCCC4BF10C", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:medtronicdiabetes:523_paradigm_revel_firmware:-:*:*:*:*:*:*:*", "matchCriteriaId": "2C8DF53F-B5DB-4690-87E8-8CECF77783EE", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:medtronicdiabetes:523_paradigm_revel:-:*:*:*:*:*:*:*", "matchCriteriaId": "95A2BE4F-FE1B-496C-8237-73F8BF0639E6", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:medtronicdiabetes:723_paradigm_revel_firmware:-:*:*:*:*:*:*:*", "matchCriteriaId": "A1F8021A-E8CD-4B6A-979D-5E3F18EB6B4A", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:medtronicdiabetes:723_paradigm_revel:-:*:*:*:*:*:*:*", "matchCriteriaId": "AB2EF040-CE07-4430-8707-689666B7AE61", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:medtronicdiabetes:523k_paradigm_revel_firmware:-:*:*:*:*:*:*:*", "matchCriteriaId": "175AE115-73FA-4A1E-B8D6-185522D8AAEC", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:medtronicdiabetes:523k_paradigm_revel:-:*:*:*:*:*:*:*", "matchCriteriaId": "425712F3-B98B-4D0E-97B9-E94D5216A43F", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:medtronicdiabetes:723k_paradigm_revel_firmware:-:*:*:*:*:*:*:*", "matchCriteriaId": "83888F47-99B8-4D7F-9049-6B9168BBFA70", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:medtronicdiabetes:723k_paradigm_revel:-:*:*:*:*:*:*:*", "matchCriteriaId": "04627C77-7F76-4883-8C6D-77579F4AF28E", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:medtronicdiabetes:551_minimed_530g_firmware:-:*:*:*:*:*:*:*", "matchCriteriaId": "86BAFB4D-49CF-4855-8A84-D41BD80A16D4", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:medtronicdiabetes:551_minimed_530g:-:*:*:*:*:*:*:*", "matchCriteriaId": "2AE48E5F-E955-4351-BC2E-E6EF0EB40AAB", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:medtronicdiabetes:751_minimed_530g_firmware:-:*:*:*:*:*:*:*", "matchCriteriaId": "8AE64752-1F20-4F8B-9048-4B004A3B3CB2", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:medtronicdiabetes:751_minimed_530g:-:*:*:*:*:*:*:*", "matchCriteriaId": "B63924E6-F2F3-47A9-BCDC-5EC4B665DC9D", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" } ], "cveTags": [], "descriptions": [ { "lang": "en", "value": "Medtronic MiniMed MMT \n\ndevices when paired with a remote controller and having the \u201ceasy bolus\u201d and \u201cremote bolus\u201d options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery." }, { "lang": "es", "value": "En la v\u00e1lvula de insulina de Medtronic MMT 508 MiniMed, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel y 551 / MMT - 751 MiniMed 530G, estos modelos, cuando se emparejan con un controlador remoto y tienen las opciones \"easy bolus\" y \"remote bolus\" habilitadas (no por defecto), son vulnerables a un ataque de captura y reproducci\u00f3n. Un atacante puede capturar las transmisiones inal\u00e1mbricas entre el controlador remoto y la v\u00e1lvula y reproducirlas para provocar una inyecci\u00f3n de insulina (bolus)." } ], "id": "CVE-2018-14781", "lastModified": "2025-05-22T17:15:22.340", "metrics": { "cvssMetricV2": [ { "acInsufInfo": false, "baseSeverity": "LOW", "cvssData": { "accessComplexity": "MEDIUM", "accessVector": "ADJACENT_NETWORK", "authentication": "NONE", "availabilityImpact": "NONE", "baseScore": 2.9, "confidentialityImpact": "NONE", "integrityImpact": "PARTIAL", "vectorString": "AV:A/AC:M/Au:N/C:N/I:P/A:N", "version": "2.0" }, "exploitabilityScore": 5.5, "impactScore": 2.9, "obtainAllPrivilege": false, "obtainOtherPrivilege": false, "obtainUserPrivilege": false, "source": "nvd@nist.gov", "type": "Primary", "userInteractionRequired": false } ], "cvssMetricV30": [ { "cvssData": { "attackComplexity": "HIGH", "attackVector": "ADJACENT_NETWORK", "availabilityImpact": "NONE", "baseScore": 5.3, "baseSeverity": "MEDIUM", "confidentialityImpact": "NONE", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N", "version": "3.0" }, "exploitabilityScore": 1.6, "impactScore": 3.6, "source": "nvd@nist.gov", "type": "Primary" } ], "cvssMetricV31": [ { "cvssData": { "attackComplexity": "HIGH", "attackVector": "ADJACENT_NETWORK", "availabilityImpact": "NONE", "baseScore": 5.3, "baseSeverity": "MEDIUM", "confidentialityImpact": "NONE", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N", "version": "3.1" }, "exploitabilityScore": 1.6, "impactScore": 3.6, "source": "ics-cert@hq.dhs.gov", "type": "Secondary" } ] }, "published": "2018-08-13T21:48:01.227", "references": [ { "source": "ics-cert@hq.dhs.gov", "tags": [ "Third Party Advisory", "VDB Entry" ], "url": "http://www.securityfocus.com/bid/105044" }, { "source": "ics-cert@hq.dhs.gov", "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed.html" }, { "source": "ics-cert@hq.dhs.gov", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Third Party Advisory", "VDB Entry" ], "url": "http://www.securityfocus.com/bid/105044" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02" } ], "sourceIdentifier": "ics-cert@hq.dhs.gov", "vulnStatus": "Modified", "weaknesses": [ { "description": [ { "lang": "en", "value": "CWE-294" } ], "source": "ics-cert@hq.dhs.gov", "type": "Secondary" }, { "description": [ { "lang": "en", "value": "CWE-287" } ], "source": "nvd@nist.gov", "type": "Primary" } ] }
Loading…
Loading…
Sightings
Author | Source | Type | Date |
---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
- Confirmed: The vulnerability is confirmed from an analyst perspective.
- Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
- Patched: This vulnerability was successfully patched by the user reporting the sighting.
- Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
- Not confirmed: The user expresses doubt about the veracity of the vulnerability.
- Not patched: This vulnerability was not successfully patched by the user reporting the sighting.
Loading…
Loading…