Summary
In Patient Information Center iX (PICiX) Versions C.02, C.03, the
software parses a formatted message or structure but does not handle or
incorrectly handles a length field that is inconsistent with the actual
length of the associated data, causing the application on the
surveillance station to restart.
References
▶ | URL | Tags | |
---|---|---|---|
ics-cert@hq.dhs.gov | https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01 | Third Party Advisory, US Government Resource | |
ics-cert@hq.dhs.gov | https://www.philips.com/productsecurity | ||
af854a3a-2127-422b-91ae-364da2661108 | https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01 | Third Party Advisory, US Government Resource | |
af854a3a-2127-422b-91ae-364da2661108 | https://www.philips.com/productsecurity |
Impacted products
Vendor | Product | Version | |
---|---|---|---|
philips | patient_information_center_ix | c.02 | |
philips | patient_information_center_ix | c.03 |
{ "configurations": [ { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:a:philips:patient_information_center_ix:c.02:*:*:*:*:*:*:*", "matchCriteriaId": "F3220AB2-AC0D-4AC2-90D8-76C02FC693EB", "vulnerable": true }, { "criteria": "cpe:2.3:a:philips:patient_information_center_ix:c.03:*:*:*:*:*:*:*", "matchCriteriaId": "AE1FB3E9-E269-434A-B1C2-D54C40F437BE", "vulnerable": true } ], "negate": false, "operator": "OR" } ] } ], "cveTags": [], "descriptions": [ { "lang": "en", "value": "In Patient Information Center iX (PICiX) Versions C.02, C.03, the \nsoftware parses a formatted message or structure but does not handle or \nincorrectly handles a length field that is inconsistent with the actual \nlength of the associated data, causing the application on the \nsurveillance station to restart.\n\n\n\n" }, { "lang": "es", "value": "Patient Information Center iX (PICiX) Versiones B.02, C.02, C.03, PerformanceBridge Focal Point Versi\u00f3n A.01, monitores de paciente IntelliVue MX100, MX400-MX850 y MP2-MP90 Versiones N y anteriores, IntelliVue X3 y X2 Versiones N y anteriores.\u0026#xa0;El software analiza una estructura o mensaje formateado, pero no maneja o maneja incorrectamente un campo de longitud que no es consistente con la longitud real de los datos asociados, causando que la aplicaci\u00f3n en la estaci\u00f3n de supervisi\u00f3n se reinicie" } ], "id": "CVE-2020-16224", "lastModified": "2024-11-21T05:06:58.277", "metrics": { "cvssMetricV2": [ { "acInsufInfo": false, "baseSeverity": "LOW", "cvssData": { "accessComplexity": "LOW", "accessVector": "ADJACENT_NETWORK", "authentication": "NONE", "availabilityImpact": "PARTIAL", "baseScore": 3.3, "confidentialityImpact": "NONE", "integrityImpact": "NONE", "vectorString": "AV:A/AC:L/Au:N/C:N/I:N/A:P", "version": "2.0" }, "exploitabilityScore": 6.5, "impactScore": 2.9, "obtainAllPrivilege": false, "obtainOtherPrivilege": false, "obtainUserPrivilege": false, "source": "nvd@nist.gov", "type": "Primary", "userInteractionRequired": false } ], "cvssMetricV31": [ { "cvssData": { "attackComplexity": "LOW", "attackVector": "ADJACENT_NETWORK", "availabilityImpact": "HIGH", "baseScore": 6.5, "baseSeverity": "MEDIUM", "confidentialityImpact": "NONE", "integrityImpact": "NONE", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H", "version": "3.1" }, "exploitabilityScore": 2.8, "impactScore": 3.6, "source": "nvd@nist.gov", "type": "Primary" } ] }, "published": "2020-09-11T14:15:11.567", "references": [ { "source": "ics-cert@hq.dhs.gov", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01" }, { "source": "ics-cert@hq.dhs.gov", "url": "https://www.philips.com/productsecurity" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "url": "https://www.philips.com/productsecurity" } ], "sourceIdentifier": "ics-cert@hq.dhs.gov", "vulnStatus": "Modified", "weaknesses": [ { "description": [ { "lang": "en", "value": "CWE-130" } ], "source": "ics-cert@hq.dhs.gov", "type": "Primary" } ] }
Loading…
Loading…
Sightings
Author | Source | Type | Date |
---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
- Confirmed: The vulnerability is confirmed from an analyst perspective.
- Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
- Patched: This vulnerability was successfully patched by the user reporting the sighting.
- Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
- Not confirmed: The user expresses doubt about the veracity of the vulnerability.
- Not patched: This vulnerability was not successfully patched by the user reporting the sighting.
Loading…
Loading…