gsd-2018-8870
Vulnerability from gsd
Modified
2023-12-13 01:22
Details
Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.
Aliases
Aliases
{ "GSD": { "alias": "CVE-2018-8870", "description": "Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.", "id": "GSD-2018-8870" }, "gsd": { "metadata": { "exploitCode": "unknown", "remediation": "unknown", "reportConfidence": "confirmed", "type": "vulnerability" }, "osvSchema": { "aliases": [ "CVE-2018-8870" ], "details": "Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.", "id": "GSD-2018-8870", "modified": "2023-12-13T01:22:34.220604Z", "schema_version": "1.4.0" } }, "namespaces": { "cve.org": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "DATE_PUBLIC": "2018-06-29T00:00:00", "ID": "CVE-2018-8870", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Medtronic MyCareLink Patient Monitor", "version": { "version_data": [ { "version_value": "24950 MyCareLink Monitor, all versions, 24952 MyCareLink Monitor, all versions." } ] } } ] }, "vendor_name": "ICS-CERT" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "USE OF HARD-CODED PASSWORD CWE-259" } ] } ] }, "references": { "reference_data": [ { "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01", "refsource": "MISC", "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01" } ] } }, "nvd.nist.gov": { "configurations": { "CVE_data_version": "4.0", "nodes": [ { "children": [ { "children": [], "cpe_match": [ { "cpe23Uri": "cpe:2.3:o:medtronic:24950_mycarelink_monitor_firmware:-:*:*:*:*:*:*:*", "cpe_name": [], "vulnerable": true } ], "operator": "OR" }, { "children": [], "cpe_match": [ { "cpe23Uri": "cpe:2.3:h:medtronic:24950_mycarelink_monitor:-:*:*:*:*:*:*:*", "cpe_name": [], "vulnerable": false } ], "operator": "OR" } ], "cpe_match": [], "operator": "AND" }, { "children": [ { "children": [], "cpe_match": [ { "cpe23Uri": "cpe:2.3:o:medtronic:24952_mycarelink_monitor_firmware:-:*:*:*:*:*:*:*", "cpe_name": [], "vulnerable": true } ], "operator": "OR" }, { "children": [], "cpe_match": [ { "cpe23Uri": "cpe:2.3:h:medtronic:24952_mycarelink_monitor:-:*:*:*:*:*:*:*", "cpe_name": [], "vulnerable": false } ], "operator": "OR" } ], "cpe_match": [], "operator": "AND" } ] }, "cve": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2018-8870" }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "en", "value": "Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "en", "value": "CWE-798" } ] } ] }, "references": { "reference_data": [ { "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01", "refsource": "MISC", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01" } ] } }, "impact": { "baseMetricV2": { "acInsufInfo": true, "cvssV2": { "accessComplexity": "LOW", "accessVector": "LOCAL", "authentication": "NONE", "availabilityImpact": "COMPLETE", "baseScore": 7.2, "confidentialityImpact": "COMPLETE", "integrityImpact": "COMPLETE", "vectorString": "AV:L/AC:L/Au:N/C:C/I:C/A:C", "version": "2.0" }, "exploitabilityScore": 3.9, "impactScore": 10.0, "obtainAllPrivilege": false, "obtainOtherPrivilege": false, "obtainUserPrivilege": false, "severity": "HIGH", "userInteractionRequired": false }, "baseMetricV3": { "cvssV3": { "attackComplexity": "LOW", "attackVector": "PHYSICAL", "availabilityImpact": "HIGH", "baseScore": 6.8, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H", "version": "3.0" }, "exploitabilityScore": 0.9, "impactScore": 5.9 } }, "lastModifiedDate": "2019-10-09T23:42Z", "publishedDate": "2018-07-03T01:29Z" } } }
Loading…
Loading…
Sightings
Author | Source | Type | Date |
---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
- Confirmed: The vulnerability is confirmed from an analyst perspective.
- Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
- Patched: This vulnerability was successfully patched by the user reporting the sighting.
- Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
- Not confirmed: The user expresses doubt about the veracity of the vulnerability.
- Not patched: This vulnerability was not successfully patched by the user reporting the sighting.
Loading…