CVE-2018-8870 (GCVE-0-2018-8870)
Vulnerability from cvelistv5
Published
2018-07-02 18:00
Modified
2025-05-22 18:10
Severity ?
VLAI Severity ?
EPSS score ?
CWE
Summary
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.
References
Impacted products
Vendor | Product | Version | |||||||
---|---|---|---|---|---|---|---|---|---|
► | Medtronic | 24950 MyCareLink Monitor |
Version: All versions |
||||||
|
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-05T07:10:47.291Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "defaultStatus": "unaffected", "product": "24950 MyCareLink Monitor", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "24952 MyCareLink Monitor", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] } ], "credits": [ { "lang": "en", "type": "finder", "value": "Peter Morgan of Clever Security reported this vulnerability" } ], "datePublic": "2018-06-29T06:00:00.000Z", "descriptions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\u003cp\u003eMedtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.\u003c/p\u003e" } ], "value": "Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system." } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "HIGH", "attackVector": "PHYSICAL", "availabilityImpact": "HIGH", "baseScore": 6.4, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H", "version": "3.1" }, "format": "CVSS", "scenarios": [ { "lang": "en", "value": "GENERAL" } ] } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-259", "description": "CWE-259", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2025-05-22T18:10:03.426Z", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-6-28-18.html" }, { "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01" } ], "solutions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\u003cdiv\u003e\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eMedtronic will release several rolling over-the-air product updates that will mitigate the vulnerabilities described within this advisory. These updates will be applied to devices automatically as part of standard, reoccurring update processes. In addition, Medtronic has increased security monitoring of affected devices and related infrastructure. \u003c/span\u003e\n\n\u003c/div\u003e\n\n\u003cp\u003eMedtronic has released additional patient focused information, at the following location:\u003c/p\u003e\u003cp\u003e\u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/security\"\u003ehttps://www.medtronic.com/security\u003c/a\u003e\u003c/p\u003e\n\n\u003cbr\u003e" } ], "value": "Medtronic will release several rolling over-the-air product updates that will mitigate the vulnerabilities described within this advisory. These updates will be applied to devices automatically as part of standard, reoccurring update processes. In addition, Medtronic has increased security monitoring of affected devices and related infrastructure. \n\n\n\n\n\nMedtronic has released additional patient focused information, at the following location:\n\n https://www.medtronic.com/security" } ], "source": { "advisory": "ICSMA-18-179-01", "discovery": "EXTERNAL" }, "title": "Medtronic MyCareLink Patient Monitor Use of Hard-coded Password", "workarounds": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\u003cp\u003eMedtronic recommends users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\u003c/p\u003e\u003cul\u003e\u003cli\u003eMaintain good physical controls over the home monitor as the best mitigation to these vulnerabilities. \u0026nbsp;\u003c/li\u003e\u003cli\u003eOnly use home monitors obtained directly from their healthcare provider or a Medtronic representative to ensure integrity of the system. \u003c/li\u003e\u003cli\u003eReport any concerning behavior regarding their home monitor to their healthcare provider or a Medtronic representative. \u003c/li\u003e\u003c/ul\u003e\u003cp\u003eMedtronic has released additional patient focused information, at the following location:\u003c/p\u003e\u003cp\u003e\u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/security\"\u003ehttps://www.medtronic.com/security\u003c/a\u003e\u003c/p\u003e\n\n\u003cbr\u003e" } ], "value": "Medtronic recommends users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\n * Maintain good physical controls over the home monitor as the best mitigation to these vulnerabilities. \u00a0\n * Only use home monitors obtained directly from their healthcare provider or a Medtronic representative to ensure integrity of the system. \n * Report any concerning behavior regarding their home monitor to their healthcare provider or a Medtronic representative. \n\n\nMedtronic has released additional patient focused information, at the following location:\n\n https://www.medtronic.com/security" } ], "x_generator": { "engine": "Vulnogram 0.2.0" }, "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "DATE_PUBLIC": "2018-06-29T00:00:00", "ID": "CVE-2018-8870", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Medtronic MyCareLink Patient Monitor", "version": { "version_data": [ { "version_value": "24950 MyCareLink Monitor, all versions, 24952 MyCareLink Monitor, all versions." } ] } } ] }, "vendor_name": "ICS-CERT" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "USE OF HARD-CODED PASSWORD CWE-259" } ] } ] }, "references": { "reference_data": [ { "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01", "refsource": "MISC", "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2018-8870", "datePublished": "2018-07-02T18:00:00Z", "dateReserved": "2018-03-20T00:00:00", "dateUpdated": "2025-05-22T18:10:03.426Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2018-10622 (GCVE-0-2018-10622)
Vulnerability from cvelistv5
Published
2018-08-10 18:00
Modified
2025-05-22 16:12
Severity ?
VLAI Severity ?
EPSS score ?
CWE
Summary
Medtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication and encryption of local data at rest.
References
Impacted products
Vendor | Product | Version | |||||||
---|---|---|---|---|---|---|---|---|---|
► | Medtronic | 24950 MyCareLink Monitor |
Version: All versions |
||||||
|
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-05T07:46:47.036Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "name": "105042", "tags": [ "vdb-entry", "x_refsource_BID", "x_transferred" ], "url": "http://www.securityfocus.com/bid/105042" }, { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "defaultStatus": "unaffected", "product": "24950 MyCareLink Monitor", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "24952 MyCareLink Monitor", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] } ], "credits": [ { "lang": "en", "type": "finder", "value": "Billy Rios, Jesse Young, and Jonathan Butts of Whitescope LLC reported these vulnerabilities" } ], "datePublic": "2018-08-07T06:00:00.000Z", "descriptions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\u003cp\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eMedtronic MyCareLink Patient Monitor \u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003euses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication and encryption of local data at rest.\u003c/span\u003e\n\n\u003c/span\u003e\n\n\u003c/p\u003e" } ], "value": "Medtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication and encryption of local data at rest." } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "HIGH", "attackVector": "PHYSICAL", "availabilityImpact": "NONE", "baseScore": 4.9, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "NONE", "privilegesRequired": "NONE", "scope": "CHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:C/C:H/I:N/A:N", "version": "3.1" }, "format": "CVSS", "scenarios": [ { "lang": "en", "value": "GENERAL" } ] } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-257", "description": "CWE-257", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2025-05-22T16:12:44.522Z", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "name": "105042", "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-8-7-18.html" }, { "tags": [ "x_refsource_MISC" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01" }, { "tags": [ "vdb-entry" ], "url": "http://www.securityfocus.com/bid/105042" } ], "source": { "advisory": "ICSMA-18-219-01", "discovery": "EXTERNAL" }, "title": "Medtronic MyCareLink 24950 Patient Monitor Storing Passwords in a Recoverable Format", "workarounds": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\u003cp\u003eMedtronic has made server-side updates to address the insufficient verification vulnerability identified in this advisory. Medtronic is implementing additional server-side mitigations to enhance data integrity and authenticity.\u003c/p\u003e\u003cp\u003eMedtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, users should:\u003c/p\u003e\u003cul\u003e\u003cli\u003eMaintain good physical control over the home monitor.\u003c/li\u003e\u003cli\u003eOnly use home monitors obtained directly from their healthcare provider or a Medtronic representative to ensure integrity of the system.\u003c/li\u003e\u003c/ul\u003e\u003cp\u003eMedtronic has released additional patient focused information, at the following location:\u003c/p\u003e\u003cp\u003e\u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/security\"\u003ehttps://www.medtronic.com/security\u003c/a\u003e\u003c/p\u003e\n\n\u003cbr\u003e" } ], "value": "Medtronic has made server-side updates to address the insufficient verification vulnerability identified in this advisory. Medtronic is implementing additional server-side mitigations to enhance data integrity and authenticity.\n\nMedtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, users should:\n\n * Maintain good physical control over the home monitor.\n * Only use home monitors obtained directly from their healthcare provider or a Medtronic representative to ensure integrity of the system.\n\n\nMedtronic has released additional patient focused information, at the following location:\n\n https://www.medtronic.com/security" } ], "x_generator": { "engine": "Vulnogram 0.2.0" }, "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "DATE_PUBLIC": "2018-08-07T00:00:00", "ID": "CVE-2018-10626", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Medtronic MyCareLink 24950, 24952 Patient Monitor", "version": { "version_data": [ { "version_value": "All versions" } ] } } ] }, "vendor_name": "ICS-CERT" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "A vulnerability was discovered in all versions of Medtronic MyCareLink 24950 and 24952 Patient Monitor. The affected product\u0027s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "INSUFFICIENT VERIFICATION OF DATA AUTHENTICITY CWE-345" } ] } ] }, "references": { "reference_data": [ { "name": "105042", "refsource": "BID", "url": "http://www.securityfocus.com/bid/105042" }, { "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01", "refsource": "MISC", "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2018-10622", "datePublished": "2018-08-10T18:00:00Z", "dateReserved": "2018-05-01T00:00:00", "dateUpdated": "2025-05-22T16:12:44.522Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2018-8868 (GCVE-0-2018-8868)
Vulnerability from cvelistv5
Published
2018-07-02 18:00
Modified
2025-05-22 18:14
Severity ?
VLAI Severity ?
EPSS score ?
CWE
Summary
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.
References
Impacted products
Vendor | Product | Version | |||||||
---|---|---|---|---|---|---|---|---|---|
► | Medtronic | 24950 MyCareLink Monitor |
Version: All versions |
||||||
|
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-05T07:10:46.236Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "defaultStatus": "unaffected", "product": "24950 MyCareLink Monitor", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "24952 MyCareLink Monitor", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] } ], "credits": [ { "lang": "en", "type": "finder", "value": "Peter Morgan of Clever Security reported this vulnerability" } ], "datePublic": "2018-06-29T06:00:00.000Z", "descriptions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\u003cp\u003e\nMedtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor\u0027s communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.\n\n\u003c/p\u003e" } ], "value": "Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor\u0027s communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality." } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "HIGH", "attackVector": "PHYSICAL", "availabilityImpact": "LOW", "baseScore": 6.2, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "LOW", "privilegesRequired": "LOW", "scope": "CHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L", "version": "3.1" }, "format": "CVSS", "scenarios": [ { "lang": "en", "value": "GENERAL" } ] } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-749", "description": "CWE-749", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2025-05-22T18:14:07.710Z", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-6-28-18.html" }, { "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01" } ], "solutions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\u003cdiv\u003e\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eMedtronic will release several rolling over-the-air product updates that will mitigate the vulnerabilities described within this advisory. These updates will be applied to devices automatically as part of standard, reoccurring update processes. In addition, Medtronic has increased security monitoring of affected devices and related infrastructure. \u003c/span\u003e\n\n\u003c/div\u003e\n\n\u003cp\u003eMedtronic has released additional patient focused information, at the following location:\u003c/p\u003e\u003cp\u003e\u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/security\"\u003ehttps://www.medtronic.com/security\u003c/a\u003e\u003c/p\u003e\n\n\u003cbr\u003e" } ], "value": "Medtronic will release several rolling over-the-air product updates that will mitigate the vulnerabilities described within this advisory. These updates will be applied to devices automatically as part of standard, reoccurring update processes. In addition, Medtronic has increased security monitoring of affected devices and related infrastructure. \n\n\n\n\n\nMedtronic has released additional patient focused information, at the following location:\n\n https://www.medtronic.com/security" } ], "source": { "advisory": "ICSMA-18-179-01", "discovery": "EXTERNAL" }, "title": "Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function", "workarounds": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\u003cp\u003eMedtronic recommends users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\u003c/p\u003e\u003cul\u003e\u003cli\u003eMaintain good physical controls over the home monitor as the best mitigation to these vulnerabilities. \u0026nbsp;\u003c/li\u003e\u003cli\u003eOnly use home monitors obtained directly from their healthcare provider or a Medtronic representative to ensure integrity of the system. \u003c/li\u003e\u003cli\u003eReport any concerning behavior regarding their home monitor to their healthcare provider or a Medtronic representative. \u003c/li\u003e\u003c/ul\u003e\u003cp\u003eMedtronic has released additional patient focused information, at the following location:\u003c/p\u003e\u003cp\u003e\u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/security\"\u003ehttps://www.medtronic.com/security\u003c/a\u003e\u003c/p\u003e\n\n\u003cbr\u003e" } ], "value": "Medtronic recommends users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\n * Maintain good physical controls over the home monitor as the best mitigation to these vulnerabilities. \u00a0\n * Only use home monitors obtained directly from their healthcare provider or a Medtronic representative to ensure integrity of the system. \n * Report any concerning behavior regarding their home monitor to their healthcare provider or a Medtronic representative. \n\n\nMedtronic has released additional patient focused information, at the following location:\n\n https://www.medtronic.com/security" } ], "x_generator": { "engine": "Vulnogram 0.2.0" }, "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "DATE_PUBLIC": "2018-06-29T00:00:00", "ID": "CVE-2018-8870", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Medtronic MyCareLink Patient Monitor", "version": { "version_data": [ { "version_value": "24950 MyCareLink Monitor, all versions, 24952 MyCareLink Monitor, all versions." } ] } } ] }, "vendor_name": "ICS-CERT" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "USE OF HARD-CODED PASSWORD CWE-259" } ] } ] }, "references": { "reference_data": [ { "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01", "refsource": "MISC", "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2018-8868", "datePublished": "2018-07-02T18:00:00Z", "dateReserved": "2018-03-20T00:00:00", "dateUpdated": "2025-05-22T18:14:07.710Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2018-10626 (GCVE-0-2018-10626)
Vulnerability from cvelistv5
Published
2018-08-10 18:00
Modified
2025-05-22 16:13
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-345 - Insufficient Verification of Data Authenticity
Summary
Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network.
References
Impacted products
Vendor | Product | Version | |||||||
---|---|---|---|---|---|---|---|---|---|
► | Medtronic | 24950 MyCareLink Monitor |
Version: All versions |
||||||
|
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-05T07:46:45.928Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "name": "105042", "tags": [ "vdb-entry", "x_refsource_BID", "x_transferred" ], "url": "http://www.securityfocus.com/bid/105042" }, { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "defaultStatus": "unaffected", "product": "24950 MyCareLink Monitor", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "24952 MyCareLink Monitor", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] } ], "credits": [ { "lang": "en", "type": "finder", "value": "Billy Rios, Jesse Young, and Jonathan Butts of Whitescope LLC reported these vulnerabilities" } ], "datePublic": "2018-08-07T06:00:00.000Z", "descriptions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\u003cp\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eMedtronic MyCareLink Patient Monitor\u2019s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network.\u003c/span\u003e\n\n\u003c/p\u003e" } ], "value": "Medtronic MyCareLink Patient Monitor\u2019s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network." } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "HIGH", "attackVector": "ADJACENT_NETWORK", "availabilityImpact": "NONE", "baseScore": 4.4, "baseSeverity": "MEDIUM", "confidentialityImpact": "LOW", "integrityImpact": "LOW", "privilegesRequired": "LOW", "scope": "CHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:N", "version": "3.1" }, "format": "CVSS", "scenarios": [ { "lang": "en", "value": "GENERAL" } ] } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-345", "description": "CWE-345 Insufficient Verification of Data Authenticity", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2025-05-22T16:13:52.072Z", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "name": "105042", "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-8-7-18.html" }, { "tags": [ "x_refsource_MISC" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01" }, { "tags": [ "vdb-entry" ], "url": "http://www.securityfocus.com/bid/105042" } ], "source": { "advisory": "ICSMA-18-219-01", "discovery": "EXTERNAL" }, "title": "Medtronic MyCareLink 24950 Patient Monitor Insufficient Verification of Data Authenticity", "workarounds": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\u003cp\u003eMedtronic has made server-side updates to address the insufficient verification vulnerability identified in this advisory. Medtronic is implementing additional server-side mitigations to enhance data integrity and authenticity.\u003c/p\u003e\u003cp\u003eMedtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, users should:\u003c/p\u003e\u003cul\u003e\u003cli\u003eMaintain good physical control over the home monitor.\u003c/li\u003e\u003cli\u003eOnly use home monitors obtained directly from their healthcare provider or a Medtronic representative to ensure integrity of the system.\u003c/li\u003e\u003c/ul\u003e\u003cp\u003eMedtronic has released additional patient focused information, at the following location:\u003c/p\u003e\u003cp\u003e\u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/security\"\u003ehttps://www.medtronic.com/security\u003c/a\u003e\u003c/p\u003e\n\n\u003cbr\u003e" } ], "value": "Medtronic has made server-side updates to address the insufficient verification vulnerability identified in this advisory. Medtronic is implementing additional server-side mitigations to enhance data integrity and authenticity.\n\nMedtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, users should:\n\n * Maintain good physical control over the home monitor.\n * Only use home monitors obtained directly from their healthcare provider or a Medtronic representative to ensure integrity of the system.\n\n\nMedtronic has released additional patient focused information, at the following location:\n\n https://www.medtronic.com/security" } ], "x_generator": { "engine": "Vulnogram 0.2.0" }, "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "DATE_PUBLIC": "2018-08-07T00:00:00", "ID": "CVE-2018-10626", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Medtronic MyCareLink 24950, 24952 Patient Monitor", "version": { "version_data": [ { "version_value": "All versions" } ] } } ] }, "vendor_name": "ICS-CERT" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "A vulnerability was discovered in all versions of Medtronic MyCareLink 24950 and 24952 Patient Monitor. The affected product\u0027s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "INSUFFICIENT VERIFICATION OF DATA AUTHENTICITY CWE-345" } ] } ] }, "references": { "reference_data": [ { "name": "105042", "refsource": "BID", "url": "http://www.securityfocus.com/bid/105042" }, { "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01", "refsource": "MISC", "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2018-10626", "datePublished": "2018-08-10T18:00:00Z", "dateReserved": "2018-05-01T00:00:00", "dateUpdated": "2025-05-22T16:13:52.072Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }