Vulnerabilites related to Hospira - Symbiq Infusion System
CVE-2015-3965 (GCVE-0-2015-3965)
Vulnerability from cvelistv5
Published
2019-03-23 19:23
Modified
2024-08-06 06:04
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- Other
Summary
Hospira Symbiq Infusion System 3.13 and earlier allows remote authenticated users to trigger "unanticipated operations" by leveraging "elevated privileges" for an unspecified call to an incorrectly exposed function.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||
---|---|---|---|---|
Hospira | Symbiq Infusion System |
Version: 3.13 and earlier |
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-06T06:04:02.807Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-174-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "product": "Symbiq Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "3.13 and earlier" } ] } ], "datePublic": "2015-07-21T00:00:00", "descriptions": [ { "lang": "en", "value": "Hospira Symbiq Infusion System 3.13 and earlier allows remote authenticated users to trigger \"unanticipated operations\" by leveraging \"elevated privileges\" for an unspecified call to an incorrectly exposed function." } ], "problemTypes": [ { "descriptions": [ { "description": "Other", "lang": "en", "type": "text" } ] } ], "providerMetadata": { "dateUpdated": "2019-03-23T19:23:49", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-174-01" } ], "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2015-3965", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Symbiq Infusion System", "version": { "version_data": [ { "version_value": "3.13 and earlier" } ] } } ] }, "vendor_name": "Hospira" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Hospira Symbiq Infusion System 3.13 and earlier allows remote authenticated users to trigger \"unanticipated operations\" by leveraging \"elevated privileges\" for an unspecified call to an incorrectly exposed function." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "Other" } ] } ] }, "references": { "reference_data": [ { "name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-174-01", "refsource": "MISC", "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-174-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2015-3965", "datePublished": "2019-03-23T19:23:49", "dateReserved": "2015-05-12T00:00:00", "dateUpdated": "2024-08-06T06:04:02.807Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2015-3956 (GCVE-0-2015-3956)
Vulnerability from cvelistv5
Published
2019-03-25 17:44
Modified
2024-08-06 06:04
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-345 - Insufficient verification of data authenticity
Summary
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
► | Hospira | Plum A+ Infusion System |
Version: <= 13.4 |
|||||||||||
|
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-06T06:04:02.927Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "product": "Plum A+ Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "\u003c= 13.4" } ] }, { "product": "Plum A+3 Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "\u003c= 13.6" } ] }, { "product": "Symbiq Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "\u003c= 3.13" } ] } ], "datePublic": "2015-06-10T00:00:00", "descriptions": [ { "lang": "en", "value": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue." } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-345", "description": "Insufficient verification of data authenticity CWE-345", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2019-03-25T17:44:44", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" } ], "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2015-3956", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Plum A+ Infusion System", "version": { "version_data": [ { "version_value": "\u003c= 13.4" } ] } }, { "product_name": "Plum A+3 Infusion System", "version": { "version_data": [ { "version_value": "\u003c= 13.6" } ] } }, { "product_name": "Symbiq Infusion System", "version": { "version_data": [ { "version_value": "\u003c= 3.13" } ] } } ] }, "vendor_name": "Hospira" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "Insufficient verification of data authenticity CWE-345" } ] } ] }, "references": { "reference_data": [ { "name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01", "refsource": "MISC", "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2015-3956", "datePublished": "2019-03-25T17:44:44", "dateReserved": "2015-05-12T00:00:00", "dateUpdated": "2024-08-06T06:04:02.927Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2015-3953 (GCVE-0-2015-3953)
Vulnerability from cvelistv5
Published
2019-03-25 16:02
Modified
2024-08-06 06:04
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-259 - Use of hard-coded password
Summary
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
► | Hospira | Plum A+ Infusion System |
Version: <= 13.4 |
|||||||||||
|
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-06T06:04:00.957Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "product": "Plum A+ Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "\u003c= 13.4" } ] }, { "product": "Plum A+3 Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "\u003c= 13.6" } ] }, { "product": "Symbiq Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "\u003c= 3.13" } ] } ], "datePublic": "2015-06-10T00:00:00", "descriptions": [ { "lang": "en", "value": "Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue." } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-259", "description": "Use of hard-coded password CWE-259", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2019-03-25T16:02:25", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" } ], "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2015-3953", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Plum A+ Infusion System", "version": { "version_data": [ { "version_value": "\u003c= 13.4" } ] } }, { "product_name": "Plum A+3 Infusion System", "version": { "version_data": [ { "version_value": "\u003c= 13.6" } ] } }, { "product_name": "Symbiq Infusion System", "version": { "version_data": [ { "version_value": "\u003c= 3.13" } ] } } ] }, "vendor_name": "Hospira" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "Use of hard-coded password CWE-259" } ] } ] }, "references": { "reference_data": [ { "name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01", "refsource": "MISC", "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2015-3953", "datePublished": "2019-03-25T16:02:25", "dateReserved": "2015-05-12T00:00:00", "dateUpdated": "2024-08-06T06:04:00.957Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2015-3954 (GCVE-0-2015-3954)
Vulnerability from cvelistv5
Published
2019-03-25 16:12
Modified
2024-08-06 06:04
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-285 - Improper authorization
Summary
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
► | Hospira | Plum A+ Infusion System |
Version: <= 13.4 |
|||||||||||
|
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-06T06:04:01.128Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "product": "Plum A+ Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "\u003c= 13.4" } ] }, { "product": "Plum A+3 Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "\u003c= 13.6" } ] }, { "product": "Symbiq Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "\u003c= 3.13" } ] } ], "datePublic": "2015-06-10T00:00:00", "descriptions": [ { "lang": "en", "value": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue." } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-285", "description": "Improper authorization CWE-285", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2019-03-25T16:12:01", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" } ], "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2015-3954", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Plum A+ Infusion System", "version": { "version_data": [ { "version_value": "\u003c= 13.4" } ] } }, { "product_name": "Plum A+3 Infusion System", "version": { "version_data": [ { "version_value": "\u003c= 13.6" } ] } }, { "product_name": "Symbiq Infusion System", "version": { "version_data": [ { "version_value": "\u003c= 3.13" } ] } } ] }, "vendor_name": "Hospira" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "Improper authorization CWE-285" } ] } ] }, "references": { "reference_data": [ { "name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01", "refsource": "MISC", "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2015-3954", "datePublished": "2019-03-25T16:12:01", "dateReserved": "2015-05-12T00:00:00", "dateUpdated": "2024-08-06T06:04:01.128Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2015-3952 (GCVE-0-2015-3952)
Vulnerability from cvelistv5
Published
2019-03-25 15:42
Modified
2024-08-06 06:04
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-312 - Cleartext storage of sensitive information
Summary
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
► | Hospira | Plum A+ Infusion System |
Version: <= 13.4 |
|||||||||||
|
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-06T06:04:01.121Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "product": "Plum A+ Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "\u003c= 13.4" } ] }, { "product": "Plum A+3 Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "\u003c= 13.6" } ] }, { "product": "Symbiq Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "\u003c= 3.13" } ] } ], "datePublic": "2015-06-10T00:00:00", "descriptions": [ { "lang": "en", "value": "Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue." } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-312", "description": "Cleartext storage of sensitive information CWE-312", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2019-03-25T15:42:39", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" } ], "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2015-3952", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Plum A+ Infusion System", "version": { "version_data": [ { "version_value": "\u003c= 13.4" } ] } }, { "product_name": "Plum A+3 Infusion System", "version": { "version_data": [ { "version_value": "\u003c= 13.6" } ] } }, { "product_name": "Symbiq Infusion System", "version": { "version_data": [ { "version_value": "\u003c= 3.13" } ] } } ] }, "vendor_name": "Hospira" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "Cleartext storage of sensitive information CWE-312" } ] } ] }, "references": { "reference_data": [ { "name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01", "refsource": "MISC", "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2015-3952", "datePublished": "2019-03-25T15:42:39", "dateReserved": "2015-05-12T00:00:00", "dateUpdated": "2024-08-06T06:04:01.121Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }