Summary
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
References
► | URL | Tags | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
► | Hospira | Plum A+ Infusion System |
Version: <= 13.4 |
|||||||||||
|
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-06T06:04:02.927Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "product": "Plum A+ Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "\u003c= 13.4" } ] }, { "product": "Plum A+3 Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "\u003c= 13.6" } ] }, { "product": "Symbiq Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "\u003c= 3.13" } ] } ], "datePublic": "2015-06-10T00:00:00", "descriptions": [ { "lang": "en", "value": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue." } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-345", "description": "Insufficient verification of data authenticity CWE-345", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2019-03-25T17:44:44", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" } ], "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2015-3956", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Plum A+ Infusion System", "version": { "version_data": [ { "version_value": "\u003c= 13.4" } ] } }, { "product_name": "Plum A+3 Infusion System", "version": { "version_data": [ { "version_value": "\u003c= 13.6" } ] } }, { "product_name": "Symbiq Infusion System", "version": { "version_data": [ { "version_value": "\u003c= 3.13" } ] } } ] }, "vendor_name": "Hospira" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "Insufficient verification of data authenticity CWE-345" } ] } ] }, "references": { "reference_data": [ { "name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01", "refsource": "MISC", "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2015-3956", "datePublished": "2019-03-25T17:44:44", "dateReserved": "2015-05-12T00:00:00", "dateUpdated": "2024-08-06T06:04:02.927Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1", "vulnerability-lookup:meta": { "nvd": "{\"cve\":{\"id\":\"CVE-2015-3956\",\"sourceIdentifier\":\"ics-cert@hq.dhs.gov\",\"published\":\"2019-03-25T18:29:00.323\",\"lastModified\":\"2024-11-21T02:30:08.517\",\"vulnStatus\":\"Modified\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.\"},{\"lang\":\"es\",\"value\":\"Hospira Plum A+ Infusion System en su versi\u00f3n 13.4 y anteriores, Plum A+3 Infusion System en su versi\u00f3n 13.6 y anteriores y Symbiq Infusion System, en su versi\u00f3n 3.13 y anteriores aceptan bibliotecas de f\u00e1rmacos, actualizaciones de firmware, comandos de bombas y cambios de configuraci\u00f3n no autorizados desde dispositivos no autenticados en la red host. Hospira recomienda que los usuarios cierren el Puerto 20/FTP y el Puerto 23/TELNET en los dispositivos afectados. Hospira ha lanzado tambi\u00e9n el Plum 360 Infusion System que no es vulnerable a este problema.\"}],\"metrics\":{\"cvssMetricV30\":[{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"3.0\",\"vectorString\":\"CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H\",\"baseScore\":9.8,\"baseSeverity\":\"CRITICAL\",\"attackVector\":\"NETWORK\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"HIGH\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":3.9,\"impactScore\":5.9}],\"cvssMetricV2\":[{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"2.0\",\"vectorString\":\"AV:N/AC:L/Au:N/C:C/I:C/A:C\",\"baseScore\":10.0,\"accessVector\":\"NETWORK\",\"accessComplexity\":\"LOW\",\"authentication\":\"NONE\",\"confidentialityImpact\":\"COMPLETE\",\"integrityImpact\":\"COMPLETE\",\"availabilityImpact\":\"COMPLETE\"},\"baseSeverity\":\"HIGH\",\"exploitabilityScore\":10.0,\"impactScore\":10.0,\"acInsufInfo\":false,\"obtainAllPrivilege\":false,\"obtainUserPrivilege\":false,\"obtainOtherPrivilege\":false,\"userInteractionRequired\":false}]},\"weaknesses\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-345\"}]},{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-345\"}]}],\"configurations\":[{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:pifzer:plum_a\\\\+_infusion_system_firmware:*:*:*:*:*:*:*:*\",\"versionEndIncluding\":\"13.4\",\"matchCriteriaId\":\"61C1A9DD-F143-4D0C-871C-B6CD7AF9DAB2\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:pifzer:plum_a\\\\+_infusion_system:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"1DB2490B-0318-4770-BF45-CD7527F15D7F\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:pifzer:plum_a\\\\+3_infusion_system_firmware:*:*:*:*:*:*:*:*\",\"versionEndIncluding\":\"13.6\",\"matchCriteriaId\":\"8536E705-89E8-47CB-9567-6AD65FBA0F1B\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:pifzer:plum_a\\\\+3_infusion_system:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"423AA561-8E38-4378-814B-1008B96F27A6\"}]}]},{\"operator\":\"AND\",\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:*:*:*:*:*:*:*:*\",\"versionEndIncluding\":\"3.13\",\"matchCriteriaId\":\"C6DD5141-72AB-4694-8595-B4BED8EC7773\"}]},{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":false,\"criteria\":\"cpe:2.3:h:pifzer:symbiq_infusion_system:-:*:*:*:*:*:*:*\",\"matchCriteriaId\":\"5C295E1A-BF60-476D-B972-5C5C28D7633B\"}]}]}],\"references\":[{\"url\":\"https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01\",\"source\":\"ics-cert@hq.dhs.gov\",\"tags\":[\"Mitigation\",\"Third Party Advisory\",\"US Government Resource\"]},{\"url\":\"https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01\",\"source\":\"af854a3a-2127-422b-91ae-364da2661108\",\"tags\":[\"Mitigation\",\"Third Party Advisory\",\"US Government Resource\"]}]}}" } }
Loading…
Loading…
Sightings
Author | Source | Type | Date |
---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
- Confirmed: The vulnerability is confirmed from an analyst perspective.
- Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
- Patched: This vulnerability was successfully patched by the user reporting the sighting.
- Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
- Not confirmed: The user expresses doubt about the veracity of the vulnerability.
- Not patched: This vulnerability was not successfully patched by the user reporting the sighting.
Loading…
Loading…