CVE-2021-23195 (GCVE-0-2021-23195)
Vulnerability from cvelistv5
Published
2022-01-21 18:17
Modified
2025-04-16 16:47
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-548 - Information Exposure Through Directory Listing
Summary
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 has the option for automated indexing (directory listing) activated. When accessing a directory, a web server delivers its entire content in HTML form. If an index file does not exist and directory listing is enabled, all content of the directory will be displayed, allowing an attacker to identify and access files on the server.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||
---|---|---|---|---|
Fresenius Kabi | Vigilant Software Suite (Mastermed Dashboard) |
Version: unspecified < 2.0.1.3 |
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-03T19:05:54.450Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ], "title": "CVE Program Container" }, { "metrics": [ { "other": { "content": { "id": "CVE-2021-23195", "options": [ { "Exploitation": "none" }, { "Automatable": "yes" }, { "Technical Impact": "partial" } ], "role": "CISA Coordinator", "timestamp": "2025-04-16T15:59:57.055622Z", "version": "2.0.3" }, "type": "ssvc" } } ], "providerMetadata": { "dateUpdated": "2025-04-16T16:47:42.124Z", "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0", "shortName": "CISA-ADP" }, "title": "CISA ADP Vulnrichment" } ], "cna": { "affected": [ { "product": "Vigilant Software Suite (Mastermed Dashboard)", "vendor": "Fresenius Kabi", "versions": [ { "lessThan": "2.0.1.3", "status": "affected", "version": "unspecified", "versionType": "custom" } ] } ], "credits": [ { "lang": "en", "value": "Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project)." } ], "descriptions": [ { "lang": "en", "value": "Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 has the option for automated indexing (directory listing) activated. When accessing a directory, a web server delivers its entire content in HTML form. If an index file does not exist and directory listing is enabled, all content of the directory will be displayed, allowing an attacker to identify and access files on the server." } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "LOW", "attackVector": "NETWORK", "availabilityImpact": "NONE", "baseScore": 5.3, "baseSeverity": "MEDIUM", "confidentialityImpact": "LOW", "integrityImpact": "NONE", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N", "version": "3.1" } } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-548", "description": "CWE-548 Information Exposure Through Directory Listing", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2022-01-21T18:17:38.000Z", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ], "solutions": [ { "lang": "en", "value": "Fresenius Kabi has created new versions to address these vulnerabilities:\n\u2022\tLink+ v3.0 (D16 or later)\n\u2022\tVSS v1.0.3 (or later)\n\u2022\tAgilia Connect Pumps Wifi Module (D29 or later)\n\u2022\tAgilia Connect Partner v3.3.2 (or later) \nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \n" } ], "source": { "advisory": "ICSMA-21-355-01", "discovery": "EXTERNAL" }, "title": "Fresenius Kabi Agilia Connect Infusion System exposure of information through directory listing", "workarounds": [ { "lang": "en", "value": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\u2022\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\n\u2022\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\n\u2022\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\n" } ], "x_generator": { "engine": "Vulnogram 0.0.9" }, "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2021-23195", "STATE": "PUBLIC", "TITLE": "Fresenius Kabi Agilia Connect Infusion System exposure of information through directory listing" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Vigilant Software Suite (Mastermed Dashboard)", "version": { "version_data": [ { "version_affected": "\u003c", "version_value": "2.0.1.3" } ] } } ] }, "vendor_name": "Fresenius Kabi" } ] } }, "credit": [ { "lang": "eng", "value": "Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project)." } ], "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 has the option for automated indexing (directory listing) activated. When accessing a directory, a web server delivers its entire content in HTML form. If an index file does not exist and directory listing is enabled, all content of the directory will be displayed, allowing an attacker to identify and access files on the server." } ] }, "generator": { "engine": "Vulnogram 0.0.9" }, "impact": { "cvss": { "attackComplexity": "LOW", "attackVector": "NETWORK", "availabilityImpact": "NONE", "baseScore": 5.3, "baseSeverity": "MEDIUM", "confidentialityImpact": "LOW", "integrityImpact": "NONE", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N", "version": "3.1" } }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "CWE-548 Information Exposure Through Directory Listing" } ] } ] }, "references": { "reference_data": [ { "name": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01", "refsource": "MISC", "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ] }, "solution": [ { "lang": "en", "value": "Fresenius Kabi has created new versions to address these vulnerabilities:\n\u2022\tLink+ v3.0 (D16 or later)\n\u2022\tVSS v1.0.3 (or later)\n\u2022\tAgilia Connect Pumps Wifi Module (D29 or later)\n\u2022\tAgilia Connect Partner v3.3.2 (or later) \nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \n" } ], "source": { "advisory": "ICSMA-21-355-01", "discovery": "EXTERNAL" }, "work_around": [ { "lang": "en", "value": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\u2022\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\n\u2022\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\n\u2022\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\n" } ] } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2021-23195", "datePublished": "2022-01-21T18:17:38.000Z", "dateReserved": "2021-11-30T00:00:00.000Z", "dateUpdated": "2025-04-16T16:47:42.124Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2021-23207 (GCVE-0-2021-23207)
Vulnerability from cvelistv5
Published
2022-01-21 18:17
Modified
2025-04-16 16:47
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-256 - Plaintext Storage of a Password
Summary
An attacker with physical access to the host can extract the secrets from the registry and create valid JWT tokens for the Fresenius Kabi Vigilant MasterMed version 2.0.1.3 application and impersonate arbitrary users. An attacker could manipulate RabbitMQ queues and messages by impersonating users.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | |||||||
---|---|---|---|---|---|---|---|---|---|
► | Fresenius Kabi | Vigilant Software Suite (Mastermed Dashboard) |
Version: unspecified < 2.0.1.3 |
||||||
|
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-03T19:05:54.432Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ], "title": "CVE Program Container" }, { "metrics": [ { "other": { "content": { "id": "CVE-2021-23207", "options": [ { "Exploitation": "none" }, { "Automatable": "no" }, { "Technical Impact": "partial" } ], "role": "CISA Coordinator", "timestamp": "2025-04-16T15:56:45.510028Z", "version": "2.0.3" }, "type": "ssvc" } } ], "providerMetadata": { "dateUpdated": "2025-04-16T16:47:32.782Z", "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0", "shortName": "CISA-ADP" }, "title": "CISA ADP Vulnrichment" } ], "cna": { "affected": [ { "product": "Vigilant Software Suite (Mastermed Dashboard)", "vendor": "Fresenius Kabi", "versions": [ { "lessThan": "2.0.1.3", "status": "affected", "version": "unspecified", "versionType": "custom" } ] }, { "product": "Agilia Partner", "vendor": "Fresenius Kabi", "versions": [ { "lessThan": "3.0", "status": "affected", "version": "unspecified", "versionType": "custom" } ] } ], "credits": [ { "lang": "en", "value": "Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project)." } ], "descriptions": [ { "lang": "en", "value": "An attacker with physical access to the host can extract the secrets from the registry and create valid JWT tokens for the Fresenius Kabi Vigilant MasterMed version 2.0.1.3 application and impersonate arbitrary users. An attacker could manipulate RabbitMQ queues and messages by impersonating users." } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "LOW", "attackVector": "LOCAL", "availabilityImpact": "NONE", "baseScore": 6.5, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "NONE", "privilegesRequired": "LOW", "scope": "CHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:C/C:H/I:N/A:N", "version": "3.1" } } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-256", "description": "CWE-256 Plaintext Storage of a Password", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2022-01-21T18:17:39.000Z", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ], "solutions": [ { "lang": "en", "value": "Fresenius Kabi has created new versions to address these vulnerabilities:\n\u2022\tLink+ v3.0 (D16 or later)\n\u2022\tVSS v1.0.3 (or later)\n\u2022\tAgilia Connect Pumps Wifi Module (D29 or later)\n\u2022\tAgilia Connect Partner v3.3.2 (or later) \nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \n" } ], "source": { "advisory": "ICSMA-21-355-01", "discovery": "EXTERNAL" }, "title": "Fresenius Kabi Agilia Connect Infusion System plaintext storage of a password", "workarounds": [ { "lang": "en", "value": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\u2022\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\n\u2022\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\n\u2022\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\n" } ], "x_generator": { "engine": "Vulnogram 0.0.9" }, "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2021-23207", "STATE": "PUBLIC", "TITLE": "Fresenius Kabi Agilia Connect Infusion System plaintext storage of a password" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Vigilant Software Suite (Mastermed Dashboard)", "version": { "version_data": [ { "version_affected": "\u003c", "version_value": "2.0.1.3" } ] } }, { "product_name": "Agilia Partner", "version": { "version_data": [ { "version_affected": "\u003c", "version_value": "3.0" } ] } } ] }, "vendor_name": "Fresenius Kabi" } ] } }, "credit": [ { "lang": "eng", "value": "Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project)." } ], "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "An attacker with physical access to the host can extract the secrets from the registry and create valid JWT tokens for the Fresenius Kabi Vigilant MasterMed version 2.0.1.3 application and impersonate arbitrary users. An attacker could manipulate RabbitMQ queues and messages by impersonating users." } ] }, "generator": { "engine": "Vulnogram 0.0.9" }, "impact": { "cvss": { "attackComplexity": "LOW", "attackVector": "LOCAL", "availabilityImpact": "NONE", "baseScore": 6.5, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "NONE", "privilegesRequired": "LOW", "scope": "CHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:C/C:H/I:N/A:N", "version": "3.1" } }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "CWE-256 Plaintext Storage of a Password" } ] } ] }, "references": { "reference_data": [ { "name": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01", "refsource": "MISC", "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ] }, "solution": [ { "lang": "en", "value": "Fresenius Kabi has created new versions to address these vulnerabilities:\n\u2022\tLink+ v3.0 (D16 or later)\n\u2022\tVSS v1.0.3 (or later)\n\u2022\tAgilia Connect Pumps Wifi Module (D29 or later)\n\u2022\tAgilia Connect Partner v3.3.2 (or later) \nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \n" } ], "source": { "advisory": "ICSMA-21-355-01", "discovery": "EXTERNAL" }, "work_around": [ { "lang": "en", "value": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\u2022\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\n\u2022\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\n\u2022\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\n" } ] } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2021-23207", "datePublished": "2022-01-21T18:17:39.000Z", "dateReserved": "2021-11-30T00:00:00.000Z", "dateUpdated": "2025-04-16T16:47:32.782Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2021-33848 (GCVE-0-2021-33848)
Vulnerability from cvelistv5
Published
2022-01-21 18:17
Modified
2025-04-16 16:46
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-79 - Cross-site Scripting (XSS)
Summary
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 is vulnerable to reflected cross-site scripting attacks. An attacker could inject JavaScript in a GET parameter of HTTP requests and perform unauthorized actions such as stealing internal information and performing actions in context of an authenticated user.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||
---|---|---|---|---|
Fresenius Kabi | Vigilant Software Suite (Mastermed Dashboard) |
Version: unspecified < 2.0.1.3 |
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-04T00:05:51.018Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ], "title": "CVE Program Container" }, { "metrics": [ { "other": { "content": { "id": "CVE-2021-33848", "options": [ { "Exploitation": "none" }, { "Automatable": "no" }, { "Technical Impact": "partial" } ], "role": "CISA Coordinator", "timestamp": "2025-04-16T15:56:36.303073Z", "version": "2.0.3" }, "type": "ssvc" } } ], "providerMetadata": { "dateUpdated": "2025-04-16T16:46:52.196Z", "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0", "shortName": "CISA-ADP" }, "title": "CISA ADP Vulnrichment" } ], "cna": { "affected": [ { "product": "Vigilant Software Suite (Mastermed Dashboard)", "vendor": "Fresenius Kabi", "versions": [ { "lessThan": "2.0.1.3", "status": "affected", "version": "unspecified", "versionType": "custom" } ] } ], "credits": [ { "lang": "en", "value": "Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project)." } ], "descriptions": [ { "lang": "en", "value": "Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 is vulnerable to reflected cross-site scripting attacks. An attacker could inject JavaScript in a GET parameter of HTTP requests and perform unauthorized actions such as stealing internal information and performing actions in context of an authenticated user." } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "LOW", "attackVector": "NETWORK", "availabilityImpact": "NONE", "baseScore": 5.4, "baseSeverity": "MEDIUM", "confidentialityImpact": "LOW", "integrityImpact": "LOW", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "REQUIRED", "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:L/A:N", "version": "3.1" } } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-79", "description": "CWE-79 Cross-site Scripting (XSS)", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2022-01-21T18:17:43.000Z", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ], "solutions": [ { "lang": "en", "value": "Fresenius Kabi has created new versions to address these vulnerabilities:\n\u2022\tLink+ v3.0 (D16 or later)\n\u2022\tVSS v1.0.3 (or later)\n\u2022\tAgilia Connect Pumps Wifi Module (D29 or later)\n\u2022\tAgilia Connect Partner v3.3.2 (or later) \nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \n" } ], "source": { "advisory": "ICSMA-21-355-01", "discovery": "EXTERNAL" }, "title": "Fresenius Kabi Agilia Connect Infusion System cross site scripting", "workarounds": [ { "lang": "en", "value": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\u2022\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\n\u2022\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\n\u2022\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\n" } ], "x_generator": { "engine": "Vulnogram 0.0.9" }, "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2021-33848", "STATE": "PUBLIC", "TITLE": "Fresenius Kabi Agilia Connect Infusion System cross site scripting" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Vigilant Software Suite (Mastermed Dashboard)", "version": { "version_data": [ { "version_affected": "\u003c", "version_value": "2.0.1.3" } ] } } ] }, "vendor_name": "Fresenius Kabi" } ] } }, "credit": [ { "lang": "eng", "value": "Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project)." } ], "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 is vulnerable to reflected cross-site scripting attacks. An attacker could inject JavaScript in a GET parameter of HTTP requests and perform unauthorized actions such as stealing internal information and performing actions in context of an authenticated user." } ] }, "generator": { "engine": "Vulnogram 0.0.9" }, "impact": { "cvss": { "attackComplexity": "LOW", "attackVector": "NETWORK", "availabilityImpact": "NONE", "baseScore": 5.4, "baseSeverity": "MEDIUM", "confidentialityImpact": "LOW", "integrityImpact": "LOW", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "REQUIRED", "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:L/A:N", "version": "3.1" } }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "CWE-79 Cross-site Scripting (XSS)" } ] } ] }, "references": { "reference_data": [ { "name": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01", "refsource": "MISC", "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ] }, "solution": [ { "lang": "en", "value": "Fresenius Kabi has created new versions to address these vulnerabilities:\n\u2022\tLink+ v3.0 (D16 or later)\n\u2022\tVSS v1.0.3 (or later)\n\u2022\tAgilia Connect Pumps Wifi Module (D29 or later)\n\u2022\tAgilia Connect Partner v3.3.2 (or later) \nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \n" } ], "source": { "advisory": "ICSMA-21-355-01", "discovery": "EXTERNAL" }, "work_around": [ { "lang": "en", "value": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\u2022\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\n\u2022\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\n\u2022\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\n" } ] } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2021-33848", "datePublished": "2022-01-21T18:17:43.000Z", "dateReserved": "2021-11-30T00:00:00.000Z", "dateUpdated": "2025-04-16T16:46:52.196Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2021-33846 (GCVE-0-2021-33846)
Vulnerability from cvelistv5
Published
2022-01-21 18:17
Modified
2025-04-16 18:01
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-327 - Use of a Broken or Risky Cryptographic Algorithm
Summary
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 issues authentication tokens to authenticated users that are signed with a symmetric encryption key. An attacker in possession of the key can issue valid JWTs and impersonate arbitrary users.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||
---|---|---|---|---|
Fresenius Kabi | Vigilant Software Suite (Mastermed Dashboard) |
Version: unspecified < 2.0.1.3 |
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-04T00:05:51.069Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ], "title": "CVE Program Container" }, { "metrics": [ { "other": { "content": { "id": "CVE-2021-33846", "options": [ { "Exploitation": "none" }, { "Automatable": "no" }, { "Technical Impact": "total" } ], "role": "CISA Coordinator", "timestamp": "2025-04-16T17:31:33.356227Z", "version": "2.0.3" }, "type": "ssvc" } } ], "providerMetadata": { "dateUpdated": "2025-04-16T18:01:40.236Z", "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0", "shortName": "CISA-ADP" }, "title": "CISA ADP Vulnrichment" } ], "cna": { "affected": [ { "product": "Vigilant Software Suite (Mastermed Dashboard)", "vendor": "Fresenius Kabi", "versions": [ { "lessThan": "2.0.1.3", "status": "affected", "version": "unspecified", "versionType": "custom" } ] } ], "credits": [ { "lang": "en", "value": "Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project)." } ], "descriptions": [ { "lang": "en", "value": "Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 issues authentication tokens to authenticated users that are signed with a symmetric encryption key. An attacker in possession of the key can issue valid JWTs and impersonate arbitrary users." } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "HIGH", "attackVector": "NETWORK", "availabilityImpact": "NONE", "baseScore": 5.9, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "privilegesRequired": "HIGH", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:N/AC:H/PR:H/UI:N/S:U/C:H/I:H/A:N", "version": "3.1" } } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-327", "description": "CWE-327 Use of a Broken or Risky Cryptographic Algorithm", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2022-01-21T18:17:37.000Z", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ], "solutions": [ { "lang": "en", "value": "Fresenius Kabi has created new versions to address these vulnerabilities:\n\u2022\tLink+ v3.0 (D16 or later)\n\u2022\tVSS v1.0.3 (or later)\n\u2022\tAgilia Connect Pumps Wifi Module (D29 or later)\n\u2022\tAgilia Connect Partner v3.3.2 (or later) \nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \n" } ], "source": { "advisory": "ICSMA-21-355-01", "discovery": "EXTERNAL" }, "title": "Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm", "workarounds": [ { "lang": "en", "value": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\u2022\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\n\u2022\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\n\u2022\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\n" } ], "x_generator": { "engine": "Vulnogram 0.0.9" }, "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2021-33846", "STATE": "PUBLIC", "TITLE": "Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Vigilant Software Suite (Mastermed Dashboard)", "version": { "version_data": [ { "version_affected": "\u003c", "version_value": "2.0.1.3" } ] } } ] }, "vendor_name": "Fresenius Kabi" } ] } }, "credit": [ { "lang": "eng", "value": "Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project)." } ], "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 issues authentication tokens to authenticated users that are signed with a symmetric encryption key. An attacker in possession of the key can issue valid JWTs and impersonate arbitrary users." } ] }, "generator": { "engine": "Vulnogram 0.0.9" }, "impact": { "cvss": { "attackComplexity": "HIGH", "attackVector": "NETWORK", "availabilityImpact": "NONE", "baseScore": 5.9, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "privilegesRequired": "HIGH", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:N/AC:H/PR:H/UI:N/S:U/C:H/I:H/A:N", "version": "3.1" } }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "CWE-327 Use of a Broken or Risky Cryptographic Algorithm" } ] } ] }, "references": { "reference_data": [ { "name": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01", "refsource": "MISC", "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ] }, "solution": [ { "lang": "en", "value": "Fresenius Kabi has created new versions to address these vulnerabilities:\n\u2022\tLink+ v3.0 (D16 or later)\n\u2022\tVSS v1.0.3 (or later)\n\u2022\tAgilia Connect Pumps Wifi Module (D29 or later)\n\u2022\tAgilia Connect Partner v3.3.2 (or later) \nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \n" } ], "source": { "advisory": "ICSMA-21-355-01", "discovery": "EXTERNAL" }, "work_around": [ { "lang": "en", "value": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\u2022\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\n\u2022\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\n\u2022\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\n" } ] } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2021-33846", "datePublished": "2022-01-21T18:17:37.000Z", "dateReserved": "2021-11-30T00:00:00.000Z", "dateUpdated": "2025-04-16T18:01:40.236Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2021-43355 (GCVE-0-2021-43355)
Vulnerability from cvelistv5
Published
2022-01-21 18:17
Modified
2025-04-16 16:46
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-603 - Use of client side authentication
Summary
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 allows user input to be validated on the client side without authentication by the server. The server should not rely on the correctness of the data because users might not support or block JavaScript or intentionally bypass the client-side checks. An attacker with knowledge of the service user could circumvent the client-side control and login with service privileges.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||
---|---|---|---|---|
Fresenius Kabi | Vigilant Software Suite (Mastermed Dashboard) |
Version: unspecified < 2.0.1.3 |
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-04T03:55:28.496Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ], "title": "CVE Program Container" }, { "metrics": [ { "other": { "content": { "id": "CVE-2021-43355", "options": [ { "Exploitation": "none" }, { "Automatable": "yes" }, { "Technical Impact": "partial" } ], "role": "CISA Coordinator", "timestamp": "2025-04-16T15:59:45.525185Z", "version": "2.0.3" }, "type": "ssvc" } } ], "providerMetadata": { "dateUpdated": "2025-04-16T16:46:31.988Z", "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0", "shortName": "CISA-ADP" }, "title": "CISA ADP Vulnrichment" } ], "cna": { "affected": [ { "product": "Vigilant Software Suite (Mastermed Dashboard)", "vendor": "Fresenius Kabi", "versions": [ { "lessThan": "2.0.1.3", "status": "affected", "version": "unspecified", "versionType": "custom" } ] } ], "credits": [ { "lang": "en", "value": "Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project)." } ], "descriptions": [ { "lang": "en", "value": "Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 allows user input to be validated on the client side without authentication by the server. The server should not rely on the correctness of the data because users might not support or block JavaScript or intentionally bypass the client-side checks. An attacker with knowledge of the service user could circumvent the client-side control and login with service privileges." } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "LOW", "attackVector": "NETWORK", "availabilityImpact": "LOW", "baseScore": 7.3, "baseSeverity": "HIGH", "confidentialityImpact": "LOW", "integrityImpact": "LOW", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L", "version": "3.1" } } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-603", "description": "CWE-603 Use of client side authentication", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2022-01-21T18:17:44.000Z", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ], "solutions": [ { "lang": "en", "value": "Fresenius Kabi has created new versions to address these vulnerabilities:\n\u2022\tLink+ v3.0 (D16 or later)\n\u2022\tVSS v1.0.3 (or later)\n\u2022\tAgilia Connect Pumps Wifi Module (D29 or later)\n\u2022\tAgilia Connect Partner v3.3.2 (or later) \nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \n" } ], "source": { "advisory": "ICSMA-21-355-01", "discovery": "EXTERNAL" }, "title": "Fresenius Kabi Agilia Connect Infusion System use of client side authentication", "workarounds": [ { "lang": "en", "value": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\u2022\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\n\u2022\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\n\u2022\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\n" } ], "x_generator": { "engine": "Vulnogram 0.0.9" }, "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2021-43355", "STATE": "PUBLIC", "TITLE": "Fresenius Kabi Agilia Connect Infusion System use of client side authentication" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Vigilant Software Suite (Mastermed Dashboard)", "version": { "version_data": [ { "version_affected": "\u003c", "version_value": "2.0.1.3" } ] } } ] }, "vendor_name": "Fresenius Kabi" } ] } }, "credit": [ { "lang": "eng", "value": "Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project)." } ], "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 allows user input to be validated on the client side without authentication by the server. The server should not rely on the correctness of the data because users might not support or block JavaScript or intentionally bypass the client-side checks. An attacker with knowledge of the service user could circumvent the client-side control and login with service privileges." } ] }, "generator": { "engine": "Vulnogram 0.0.9" }, "impact": { "cvss": { "attackComplexity": "LOW", "attackVector": "NETWORK", "availabilityImpact": "LOW", "baseScore": 7.3, "baseSeverity": "HIGH", "confidentialityImpact": "LOW", "integrityImpact": "LOW", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L", "version": "3.1" } }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "CWE-603 Use of client side authentication" } ] } ] }, "references": { "reference_data": [ { "name": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01", "refsource": "MISC", "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ] }, "solution": [ { "lang": "en", "value": "Fresenius Kabi has created new versions to address these vulnerabilities:\n\u2022\tLink+ v3.0 (D16 or later)\n\u2022\tVSS v1.0.3 (or later)\n\u2022\tAgilia Connect Pumps Wifi Module (D29 or later)\n\u2022\tAgilia Connect Partner v3.3.2 (or later) \nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \n" } ], "source": { "advisory": "ICSMA-21-355-01", "discovery": "EXTERNAL" }, "work_around": [ { "lang": "en", "value": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\u2022\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\n\u2022\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\n\u2022\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\n" } ] } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2021-43355", "datePublished": "2022-01-21T18:17:44.000Z", "dateReserved": "2021-11-30T00:00:00.000Z", "dateUpdated": "2025-04-16T16:46:31.988Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2021-44464 (GCVE-0-2021-44464)
Vulnerability from cvelistv5
Published
2022-01-21 18:17
Modified
2025-04-16 16:47
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-798 - Use of Hard-coded Credentials
Summary
Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 contains service credentials likely to be common across all instances. An attacker in possession of the password may gain privileges on all installations of this software.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||
---|---|---|---|---|
Fresenius Kabi | Vigilant Software Suite (Mastermed Dashboard) |
Version: unspecified < 2.0.1.3 |
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-04T04:25:16.824Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ], "title": "CVE Program Container" }, { "metrics": [ { "other": { "content": { "id": "CVE-2021-44464", "options": [ { "Exploitation": "none" }, { "Automatable": "no" }, { "Technical Impact": "partial" } ], "role": "CISA Coordinator", "timestamp": "2025-04-16T15:56:40.775925Z", "version": "2.0.3" }, "type": "ssvc" } } ], "providerMetadata": { "dateUpdated": "2025-04-16T16:47:01.630Z", "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0", "shortName": "CISA-ADP" }, "title": "CISA ADP Vulnrichment" } ], "cna": { "affected": [ { "product": "Vigilant Software Suite (Mastermed Dashboard)", "vendor": "Fresenius Kabi", "versions": [ { "lessThan": "2.0.1.3", "status": "affected", "version": "unspecified", "versionType": "custom" } ] } ], "credits": [ { "lang": "en", "value": "Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project)." } ], "descriptions": [ { "lang": "en", "value": "Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 contains service credentials likely to be common across all instances. An attacker in possession of the password may gain privileges on all installations of this software." } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "LOW", "attackVector": "NETWORK", "availabilityImpact": "LOW", "baseScore": 6.3, "baseSeverity": "MEDIUM", "confidentialityImpact": "LOW", "integrityImpact": "LOW", "privilegesRequired": "LOW", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:L/A:L", "version": "3.1" } } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-798", "description": "CWE-798 Use of Hard-coded Credentials", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2022-01-21T18:17:42.000Z", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ], "solutions": [ { "lang": "en", "value": "Fresenius Kabi has created new versions to address these vulnerabilities:\n\u2022\tLink+ v3.0 (D16 or later)\n\u2022\tVSS v1.0.3 (or later)\n\u2022\tAgilia Connect Pumps Wifi Module (D29 or later)\n\u2022\tAgilia Connect Partner v3.3.2 (or later) \nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \n" } ], "source": { "advisory": "ICSMA-21-355-01", "discovery": "EXTERNAL" }, "title": "Fresenius Kabi Agilia Connect Infusion System hard coded credentials", "workarounds": [ { "lang": "en", "value": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\u2022\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\n\u2022\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\n\u2022\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\n" } ], "x_generator": { "engine": "Vulnogram 0.0.9" }, "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2021-44464", "STATE": "PUBLIC", "TITLE": "Fresenius Kabi Agilia Connect Infusion System hard coded credentials" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Vigilant Software Suite (Mastermed Dashboard)", "version": { "version_data": [ { "version_affected": "\u003c", "version_value": "2.0.1.3" } ] } } ] }, "vendor_name": "Fresenius Kabi" } ] } }, "credit": [ { "lang": "eng", "value": "Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project)." } ], "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 contains service credentials likely to be common across all instances. An attacker in possession of the password may gain privileges on all installations of this software." } ] }, "generator": { "engine": "Vulnogram 0.0.9" }, "impact": { "cvss": { "attackComplexity": "LOW", "attackVector": "NETWORK", "availabilityImpact": "LOW", "baseScore": 6.3, "baseSeverity": "MEDIUM", "confidentialityImpact": "LOW", "integrityImpact": "LOW", "privilegesRequired": "LOW", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:L/A:L", "version": "3.1" } }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "CWE-798 Use of Hard-coded Credentials" } ] } ] }, "references": { "reference_data": [ { "name": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01", "refsource": "MISC", "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01" } ] }, "solution": [ { "lang": "en", "value": "Fresenius Kabi has created new versions to address these vulnerabilities:\n\u2022\tLink+ v3.0 (D16 or later)\n\u2022\tVSS v1.0.3 (or later)\n\u2022\tAgilia Connect Pumps Wifi Module (D29 or later)\n\u2022\tAgilia Connect Partner v3.3.2 (or later) \nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \n" } ], "source": { "advisory": "ICSMA-21-355-01", "discovery": "EXTERNAL" }, "work_around": [ { "lang": "en", "value": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\u2022\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\n\u2022\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\n\u2022\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\n" } ] } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2021-44464", "datePublished": "2022-01-21T18:17:42.000Z", "dateReserved": "2021-11-30T00:00:00.000Z", "dateUpdated": "2025-04-16T16:47:01.630Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }