Vulnerabilites related to sooil - diabecare_rs_firmware
Vulnerability from fkie_nvd
Published
2021-01-19 17:15
Modified
2024-11-21 05:20
Severity ?
Summary
SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i & AnyDana-A mobile apps doesn't use adequate measures to authenticate the communicating entities before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the authentication sequence via Bluetooth Low Energy.
References
▶ | URL | Tags | |
---|---|---|---|
ics-cert@hq.dhs.gov | https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01 | Third Party Advisory, US Government Resource | |
af854a3a-2127-422b-91ae-364da2661108 | https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01 | Third Party Advisory, US Government Resource |
Impacted products
Vendor | Product | Version | |
---|---|---|---|
sooil | anydana-a_firmware | * | |
sooil | anydana-a | - | |
sooil | anydana-i_firmware | * | |
sooil | anydana-i | - | |
sooil | diabecare_rs_firmware | * | |
sooil | diabecare_rs | - |
{ "configurations": [ { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:anydana-a_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "808DBA89-C2C2-4C03-814D-DB0B4DFF8B22", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:anydana-a:-:*:*:*:*:*:*:*", "matchCriteriaId": "2624582E-762F-4CDB-B974-1BC971F1544D", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:anydana-i_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "98E25B7D-C1ED-4144-8411-171C7B6426E4", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:anydana-i:-:*:*:*:*:*:*:*", "matchCriteriaId": "E25CB863-07D9-4E90-8022-6C2B4C52EDFC", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:diabecare_rs_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "C5C154F1-7F29-4EC0-9971-76A02C6DA8BC", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:diabecare_rs:-:*:*:*:*:*:*:*", "matchCriteriaId": "8628A530-AF34-453E-968A-40DD5B8EE456", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" } ], "cveTags": [], "descriptions": [ { "lang": "en", "value": "SOOIL Developments Co Ltd DiabecareRS,AnyDana-i \u0026 AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i \u0026 AnyDana-A mobile apps doesn\u0027t use adequate measures to authenticate the communicating entities before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the authentication sequence via Bluetooth Low Energy." }, { "lang": "es", "value": "SOOIL Developments Co Ltd DiabecareRS, AnyDana-i \u0026amp; AnyDana-A, el protocolo de comunicaci\u00f3n de la bomba de insulina y sus aplicaciones m\u00f3viles AnyDana-i, AnyDana-A no utiliza las medidas adecuadas para autenticar a las entidades comunicantes antes de intercambiar claves, lo que permite no autenticar , los atacantes f\u00edsicamente pr\u00f3ximos para rastrear la secuencia de autenticaci\u00f3n por medio de Bluetooth Low Energy" } ], "id": "CVE-2020-27276", "lastModified": "2024-11-21T05:20:58.943", "metrics": { "cvssMetricV2": [ { "acInsufInfo": false, "baseSeverity": "LOW", "cvssData": { "accessComplexity": "MEDIUM", "accessVector": "ADJACENT_NETWORK", "authentication": "NONE", "availabilityImpact": "NONE", "baseScore": 2.9, "confidentialityImpact": "PARTIAL", "integrityImpact": "NONE", "vectorString": "AV:A/AC:M/Au:N/C:P/I:N/A:N", "version": "2.0" }, "exploitabilityScore": 5.5, "impactScore": 2.9, "obtainAllPrivilege": false, "obtainOtherPrivilege": false, "obtainUserPrivilege": false, "source": "nvd@nist.gov", "type": "Primary", "userInteractionRequired": true } ], "cvssMetricV31": [ { "cvssData": { "attackComplexity": "LOW", "attackVector": "ADJACENT_NETWORK", "availabilityImpact": "NONE", "baseScore": 5.7, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "NONE", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "REQUIRED", "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:H/I:N/A:N", "version": "3.1" }, "exploitabilityScore": 2.1, "impactScore": 3.6, "source": "nvd@nist.gov", "type": "Primary" } ] }, "published": "2021-01-19T17:15:12.677", "references": [ { "source": "ics-cert@hq.dhs.gov", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "sourceIdentifier": "ics-cert@hq.dhs.gov", "vulnStatus": "Modified", "weaknesses": [ { "description": [ { "lang": "en", "value": "CWE-290" } ], "source": "ics-cert@hq.dhs.gov", "type": "Secondary" }, { "description": [ { "lang": "en", "value": "CWE-290" } ], "source": "nvd@nist.gov", "type": "Primary" } ] }
Vulnerability from fkie_nvd
Published
2021-01-19 22:15
Modified
2024-11-21 05:20
Severity ?
Summary
In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a client-side control vulnerability in the insulin pump and its AnyDana-i and AnyDana-A mobile applications allows physically proximate attackers to bypass checks for default PINs via Bluetooth Low Energy.
References
▶ | URL | Tags | |
---|---|---|---|
ics-cert@hq.dhs.gov | https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01 | Third Party Advisory, US Government Resource | |
af854a3a-2127-422b-91ae-364da2661108 | https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01 | Third Party Advisory, US Government Resource |
Impacted products
Vendor | Product | Version | |
---|---|---|---|
sooil | anydana-a_firmware | * | |
sooil | anydana-a | - | |
sooil | anydana-i_firmware | * | |
sooil | anydana-i | - | |
sooil | diabecare_rs_firmware | * | |
sooil | diabecare_rs | - |
{ "configurations": [ { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:anydana-a_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "808DBA89-C2C2-4C03-814D-DB0B4DFF8B22", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:anydana-a:-:*:*:*:*:*:*:*", "matchCriteriaId": "2624582E-762F-4CDB-B974-1BC971F1544D", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:anydana-i_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "98E25B7D-C1ED-4144-8411-171C7B6426E4", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:anydana-i:-:*:*:*:*:*:*:*", "matchCriteriaId": "E25CB863-07D9-4E90-8022-6C2B4C52EDFC", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:diabecare_rs_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "C5C154F1-7F29-4EC0-9971-76A02C6DA8BC", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:diabecare_rs:-:*:*:*:*:*:*:*", "matchCriteriaId": "8628A530-AF34-453E-968A-40DD5B8EE456", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" } ], "cveTags": [], "descriptions": [ { "lang": "en", "value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a client-side control vulnerability in the insulin pump and its AnyDana-i and AnyDana-A mobile applications allows physically proximate attackers to bypass checks for default PINs via Bluetooth Low Energy." }, { "lang": "es", "value": "En SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i y AnyDana-A, una vulnerabilidad de control del lado del cliente en la bomba de insulina y sus aplicaciones m\u00f3viles AnyDana-i y AnyDana-A, permite a atacantes pr\u00f3ximos f\u00edsicamente omitir las comprobaciones para los PIN predeterminados por medio de Bluetooth Low Energy" } ], "id": "CVE-2020-27268", "lastModified": "2024-11-21T05:20:58.377", "metrics": { "cvssMetricV2": [ { "acInsufInfo": false, "baseSeverity": "LOW", "cvssData": { "accessComplexity": "LOW", "accessVector": "ADJACENT_NETWORK", "authentication": "NONE", "availabilityImpact": "NONE", "baseScore": 3.3, "confidentialityImpact": "PARTIAL", "integrityImpact": "NONE", "vectorString": "AV:A/AC:L/Au:N/C:P/I:N/A:N", "version": "2.0" }, "exploitabilityScore": 6.5, "impactScore": 2.9, "obtainAllPrivilege": false, "obtainOtherPrivilege": false, "obtainUserPrivilege": false, "source": "nvd@nist.gov", "type": "Primary", "userInteractionRequired": false } ], "cvssMetricV31": [ { "cvssData": { "attackComplexity": "LOW", "attackVector": "ADJACENT_NETWORK", "availabilityImpact": "NONE", "baseScore": 6.5, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "NONE", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N", "version": "3.1" }, "exploitabilityScore": 2.8, "impactScore": 3.6, "source": "nvd@nist.gov", "type": "Primary" } ] }, "published": "2021-01-19T22:15:12.427", "references": [ { "source": "ics-cert@hq.dhs.gov", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "sourceIdentifier": "ics-cert@hq.dhs.gov", "vulnStatus": "Modified", "weaknesses": [ { "description": [ { "lang": "en", "value": "CWE-602" } ], "source": "ics-cert@hq.dhs.gov", "type": "Secondary" }, { "description": [ { "lang": "en", "value": "CWE-669" } ], "source": "nvd@nist.gov", "type": "Primary" } ] }
Vulnerability from fkie_nvd
Published
2021-01-19 21:15
Modified
2024-11-21 05:20
Severity ?
Summary
In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a hard-coded physician PIN in the physician menu of the insulin pump allows attackers with physical access to change insulin therapy settings.
References
▶ | URL | Tags | |
---|---|---|---|
ics-cert@hq.dhs.gov | https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01 | Third Party Advisory, US Government Resource | |
af854a3a-2127-422b-91ae-364da2661108 | https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01 | Third Party Advisory, US Government Resource |
Impacted products
Vendor | Product | Version | |
---|---|---|---|
sooil | anydana-a_firmware | * | |
sooil | anydana-a | - | |
sooil | anydana-i_firmware | * | |
sooil | anydana-i | - | |
sooil | diabecare_rs_firmware | * | |
sooil | diabecare_rs | - |
{ "configurations": [ { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:anydana-a_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "808DBA89-C2C2-4C03-814D-DB0B4DFF8B22", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:anydana-a:-:*:*:*:*:*:*:*", "matchCriteriaId": "2624582E-762F-4CDB-B974-1BC971F1544D", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:anydana-i_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "98E25B7D-C1ED-4144-8411-171C7B6426E4", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:anydana-i:-:*:*:*:*:*:*:*", "matchCriteriaId": "E25CB863-07D9-4E90-8022-6C2B4C52EDFC", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:diabecare_rs_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "C5C154F1-7F29-4EC0-9971-76A02C6DA8BC", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:diabecare_rs:-:*:*:*:*:*:*:*", "matchCriteriaId": "8628A530-AF34-453E-968A-40DD5B8EE456", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" } ], "cveTags": [], "descriptions": [ { "lang": "en", "value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a hard-coded physician PIN in the physician menu of the insulin pump allows attackers with physical access to change insulin therapy settings." }, { "lang": "es", "value": "En SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i y AnyDana-A, un PIN m\u00e9dico embebido en el men\u00fa m\u00e9dico de la bomba de insulina, permite a atacantes con acceso f\u00edsico cambiar la configuraci\u00f3n de la terapia de insulina" } ], "id": "CVE-2020-27256", "lastModified": "2024-11-21T05:20:57.207", "metrics": { "cvssMetricV2": [ { "acInsufInfo": false, "baseSeverity": "MEDIUM", "cvssData": { "accessComplexity": "LOW", "accessVector": "LOCAL", "authentication": "NONE", "availabilityImpact": "PARTIAL", "baseScore": 4.6, "confidentialityImpact": "PARTIAL", "integrityImpact": "PARTIAL", "vectorString": "AV:L/AC:L/Au:N/C:P/I:P/A:P", "version": "2.0" }, "exploitabilityScore": 3.9, "impactScore": 6.4, "obtainAllPrivilege": false, "obtainOtherPrivilege": false, "obtainUserPrivilege": false, "source": "nvd@nist.gov", "type": "Primary", "userInteractionRequired": false } ], "cvssMetricV31": [ { "cvssData": { "attackComplexity": "LOW", "attackVector": "PHYSICAL", "availabilityImpact": "HIGH", "baseScore": 6.8, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H", "version": "3.1" }, "exploitabilityScore": 0.9, "impactScore": 5.9, "source": "nvd@nist.gov", "type": "Primary" } ] }, "published": "2021-01-19T21:15:13.967", "references": [ { "source": "ics-cert@hq.dhs.gov", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "sourceIdentifier": "ics-cert@hq.dhs.gov", "vulnStatus": "Modified", "weaknesses": [ { "description": [ { "lang": "en", "value": "CWE-798" } ], "source": "ics-cert@hq.dhs.gov", "type": "Secondary" }, { "description": [ { "lang": "en", "value": "CWE-798" } ], "source": "nvd@nist.gov", "type": "Primary" } ] }
Vulnerability from fkie_nvd
Published
2021-01-19 22:15
Modified
2024-11-21 05:20
Severity ?
Summary
In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a client-side control vulnerability in the insulin pump and its AnyDana-i and AnyDana-A mobile applications allows physically proximate attackers to bypass user authentication checks via Bluetooth Low Energy.
References
▶ | URL | Tags | |
---|---|---|---|
ics-cert@hq.dhs.gov | https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01 | Third Party Advisory, US Government Resource | |
af854a3a-2127-422b-91ae-364da2661108 | https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01 | Third Party Advisory, US Government Resource |
Impacted products
Vendor | Product | Version | |
---|---|---|---|
sooil | anydana-a_firmware | * | |
sooil | anydana-a | - | |
sooil | anydana-i_firmware | * | |
sooil | anydana-i | - | |
sooil | diabecare_rs_firmware | * | |
sooil | diabecare_rs | - |
{ "configurations": [ { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:anydana-a_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "808DBA89-C2C2-4C03-814D-DB0B4DFF8B22", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:anydana-a:-:*:*:*:*:*:*:*", "matchCriteriaId": "2624582E-762F-4CDB-B974-1BC971F1544D", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:anydana-i_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "98E25B7D-C1ED-4144-8411-171C7B6426E4", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:anydana-i:-:*:*:*:*:*:*:*", "matchCriteriaId": "E25CB863-07D9-4E90-8022-6C2B4C52EDFC", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:diabecare_rs_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "C5C154F1-7F29-4EC0-9971-76A02C6DA8BC", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:diabecare_rs:-:*:*:*:*:*:*:*", "matchCriteriaId": "8628A530-AF34-453E-968A-40DD5B8EE456", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" } ], "cveTags": [], "descriptions": [ { "lang": "en", "value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a client-side control vulnerability in the insulin pump and its AnyDana-i and AnyDana-A mobile applications allows physically proximate attackers to bypass user authentication checks via Bluetooth Low Energy." }, { "lang": "es", "value": "En SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i y AnyDana-A, una vulnerabilidad de control del lado del cliente en la bomba de insulina y sus aplicaciones m\u00f3viles AnyDana-i y AnyDana-A, permite a atacantes f\u00edsicamente pr\u00f3ximos omitir las comprobaciones de autenticaci\u00f3n del usuario por medio de Bluetooth Low Energy" } ], "id": "CVE-2020-27266", "lastModified": "2024-11-21T05:20:58.180", "metrics": { "cvssMetricV2": [ { "acInsufInfo": false, "baseSeverity": "LOW", "cvssData": { "accessComplexity": "LOW", "accessVector": "ADJACENT_NETWORK", "authentication": "NONE", "availabilityImpact": "NONE", "baseScore": 3.3, "confidentialityImpact": "PARTIAL", "integrityImpact": "NONE", "vectorString": "AV:A/AC:L/Au:N/C:P/I:N/A:N", "version": "2.0" }, "exploitabilityScore": 6.5, "impactScore": 2.9, "obtainAllPrivilege": false, "obtainOtherPrivilege": false, "obtainUserPrivilege": false, "source": "nvd@nist.gov", "type": "Primary", "userInteractionRequired": false } ], "cvssMetricV31": [ { "cvssData": { "attackComplexity": "LOW", "attackVector": "ADJACENT_NETWORK", "availabilityImpact": "NONE", "baseScore": 6.5, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "NONE", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N", "version": "3.1" }, "exploitabilityScore": 2.8, "impactScore": 3.6, "source": "nvd@nist.gov", "type": "Primary" } ] }, "published": "2021-01-19T22:15:12.380", "references": [ { "source": "ics-cert@hq.dhs.gov", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "sourceIdentifier": "ics-cert@hq.dhs.gov", "vulnStatus": "Modified", "weaknesses": [ { "description": [ { "lang": "en", "value": "CWE-603" } ], "source": "ics-cert@hq.dhs.gov", "type": "Secondary" }, { "description": [ { "lang": "en", "value": "CWE-287" } ], "source": "nvd@nist.gov", "type": "Primary" } ] }
Vulnerability from fkie_nvd
Published
2021-01-19 22:15
Modified
2024-11-21 05:20
Severity ?
Summary
In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications lacks replay protection measures, which allows unauthenticated, physically proximate attackers to replay communication sequences via Bluetooth Low Energy.
References
▶ | URL | Tags | |
---|---|---|---|
ics-cert@hq.dhs.gov | https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01 | Third Party Advisory, US Government Resource | |
af854a3a-2127-422b-91ae-364da2661108 | https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01 | Third Party Advisory, US Government Resource |
Impacted products
Vendor | Product | Version | |
---|---|---|---|
sooil | anydana-a_firmware | * | |
sooil | anydana-a | - | |
sooil | anydana-i_firmware | * | |
sooil | anydana-i | - | |
sooil | diabecare_rs_firmware | * | |
sooil | diabecare_rs | - |
{ "configurations": [ { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:anydana-a_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "808DBA89-C2C2-4C03-814D-DB0B4DFF8B22", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:anydana-a:-:*:*:*:*:*:*:*", "matchCriteriaId": "2624582E-762F-4CDB-B974-1BC971F1544D", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:anydana-i_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "98E25B7D-C1ED-4144-8411-171C7B6426E4", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:anydana-i:-:*:*:*:*:*:*:*", "matchCriteriaId": "E25CB863-07D9-4E90-8022-6C2B4C52EDFC", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:diabecare_rs_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "C5C154F1-7F29-4EC0-9971-76A02C6DA8BC", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:diabecare_rs:-:*:*:*:*:*:*:*", "matchCriteriaId": "8628A530-AF34-453E-968A-40DD5B8EE456", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" } ], "cveTags": [], "descriptions": [ { "lang": "en", "value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications lacks replay protection measures, which allows unauthenticated, physically proximate attackers to replay communication sequences via Bluetooth Low Energy." }, { "lang": "es", "value": "En SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i y AnyDana-A, el protocolo de comunicaci\u00f3n de la bomba de insulina y sus aplicaciones m\u00f3viles AnyDana-i y AnyDana-A, carece de medidas de protecci\u00f3n de reproducci\u00f3n, lo que permite a los atacantes no autenticados f\u00edsicamente pr\u00f3ximos poder reproducir secuencias de comunicaci\u00f3n por medio de Bluetooth Low Energy" } ], "id": "CVE-2020-27269", "lastModified": "2024-11-21T05:20:58.480", "metrics": { "cvssMetricV2": [ { "acInsufInfo": false, "baseSeverity": "LOW", "cvssData": { "accessComplexity": "MEDIUM", "accessVector": "ADJACENT_NETWORK", "authentication": "NONE", "availabilityImpact": "NONE", "baseScore": 2.9, "confidentialityImpact": "PARTIAL", "integrityImpact": "NONE", "vectorString": "AV:A/AC:M/Au:N/C:P/I:N/A:N", "version": "2.0" }, "exploitabilityScore": 5.5, "impactScore": 2.9, "obtainAllPrivilege": false, "obtainOtherPrivilege": false, "obtainUserPrivilege": false, "source": "nvd@nist.gov", "type": "Primary", "userInteractionRequired": true } ], "cvssMetricV31": [ { "cvssData": { "attackComplexity": "LOW", "attackVector": "ADJACENT_NETWORK", "availabilityImpact": "NONE", "baseScore": 5.7, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "NONE", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "REQUIRED", "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:H/I:N/A:N", "version": "3.1" }, "exploitabilityScore": 2.1, "impactScore": 3.6, "source": "nvd@nist.gov", "type": "Primary" } ] }, "published": "2021-01-19T22:15:12.487", "references": [ { "source": "ics-cert@hq.dhs.gov", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "sourceIdentifier": "ics-cert@hq.dhs.gov", "vulnStatus": "Modified", "weaknesses": [ { "description": [ { "lang": "en", "value": "CWE-294" } ], "source": "ics-cert@hq.dhs.gov", "type": "Secondary" }, { "description": [ { "lang": "en", "value": "CWE-294" } ], "source": "nvd@nist.gov", "type": "Primary" } ] }
Vulnerability from fkie_nvd
Published
2021-01-19 21:15
Modified
2024-11-21 05:20
Severity ?
Summary
In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications use deterministic keys, which allows unauthenticated, physically proximate attackers to brute-force the keys via Bluetooth Low Energy.
References
▶ | URL | Tags | |
---|---|---|---|
ics-cert@hq.dhs.gov | https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01 | Third Party Advisory, US Government Resource | |
af854a3a-2127-422b-91ae-364da2661108 | https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01 | Third Party Advisory, US Government Resource |
Impacted products
Vendor | Product | Version | |
---|---|---|---|
sooil | anydana-a_firmware | * | |
sooil | anydana-a | - | |
sooil | anydana-i_firmware | * | |
sooil | anydana-i | - | |
sooil | diabecare_rs_firmware | * | |
sooil | diabecare_rs | - |
{ "configurations": [ { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:anydana-a_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "808DBA89-C2C2-4C03-814D-DB0B4DFF8B22", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:anydana-a:-:*:*:*:*:*:*:*", "matchCriteriaId": "2624582E-762F-4CDB-B974-1BC971F1544D", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:anydana-i_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "98E25B7D-C1ED-4144-8411-171C7B6426E4", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:anydana-i:-:*:*:*:*:*:*:*", "matchCriteriaId": "E25CB863-07D9-4E90-8022-6C2B4C52EDFC", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:diabecare_rs_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "C5C154F1-7F29-4EC0-9971-76A02C6DA8BC", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:diabecare_rs:-:*:*:*:*:*:*:*", "matchCriteriaId": "8628A530-AF34-453E-968A-40DD5B8EE456", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" } ], "cveTags": [], "descriptions": [ { "lang": "en", "value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications use deterministic keys, which allows unauthenticated, physically proximate attackers to brute-force the keys via Bluetooth Low Energy." }, { "lang": "es", "value": "En SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i y AnyDana-A, el protocolo de comunicaci\u00f3n de la bomba de insulina y sus aplicaciones m\u00f3viles AnyDana-i y AnyDana-A, utilizan claves deterministas, lo que permite a atacantes no autenticados y f\u00edsicamente pr\u00f3ximos llevar a cabo ataques de fuerza bruta por medio de Bluetooth Low Energy" } ], "id": "CVE-2020-27264", "lastModified": "2024-11-21T05:20:57.980", "metrics": { "cvssMetricV2": [ { "acInsufInfo": false, "baseSeverity": "LOW", "cvssData": { "accessComplexity": "LOW", "accessVector": "ADJACENT_NETWORK", "authentication": "NONE", "availabilityImpact": "NONE", "baseScore": 3.3, "confidentialityImpact": "PARTIAL", "integrityImpact": "NONE", "vectorString": "AV:A/AC:L/Au:N/C:P/I:N/A:N", "version": "2.0" }, "exploitabilityScore": 6.5, "impactScore": 2.9, "obtainAllPrivilege": false, "obtainOtherPrivilege": false, "obtainUserPrivilege": false, "source": "nvd@nist.gov", "type": "Primary", "userInteractionRequired": false } ], "cvssMetricV31": [ { "cvssData": { "attackComplexity": "LOW", "attackVector": "ADJACENT_NETWORK", "availabilityImpact": "HIGH", "baseScore": 8.8, "baseSeverity": "HIGH", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H", "version": "3.1" }, "exploitabilityScore": 2.8, "impactScore": 5.9, "source": "nvd@nist.gov", "type": "Primary" } ] }, "published": "2021-01-19T21:15:14.110", "references": [ { "source": "ics-cert@hq.dhs.gov", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "sourceIdentifier": "ics-cert@hq.dhs.gov", "vulnStatus": "Modified", "weaknesses": [ { "description": [ { "lang": "en", "value": "CWE-330" } ], "source": "ics-cert@hq.dhs.gov", "type": "Secondary" }, { "description": [ { "lang": "en", "value": "CWE-330" } ], "source": "nvd@nist.gov", "type": "Primary" } ] }
Vulnerability from fkie_nvd
Published
2021-01-19 17:15
Modified
2024-11-21 05:20
Severity ?
Summary
SOOIL Developments CoLtd DiabecareRS, AnyDana-i, AnyDana-A, The communication protocol of the insulin pump and AnyDana-i,AnyDana-A mobile apps doesn't use adequate measures to authenticate the pump before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the keys and spoof the pump via BLE.
References
▶ | URL | Tags | |
---|---|---|---|
ics-cert@hq.dhs.gov | https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01 | Third Party Advisory, US Government Resource | |
af854a3a-2127-422b-91ae-364da2661108 | https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01 | Third Party Advisory, US Government Resource |
Impacted products
Vendor | Product | Version | |
---|---|---|---|
sooil | anydana-a_firmware | * | |
sooil | anydana-a | - | |
sooil | anydana-i_firmware | * | |
sooil | anydana-i | - | |
sooil | diabecare_rs_firmware | * | |
sooil | diabecare_rs | - |
{ "configurations": [ { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:anydana-a_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "808DBA89-C2C2-4C03-814D-DB0B4DFF8B22", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:anydana-a:-:*:*:*:*:*:*:*", "matchCriteriaId": "2624582E-762F-4CDB-B974-1BC971F1544D", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:anydana-i_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "98E25B7D-C1ED-4144-8411-171C7B6426E4", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:anydana-i:-:*:*:*:*:*:*:*", "matchCriteriaId": "E25CB863-07D9-4E90-8022-6C2B4C52EDFC", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:diabecare_rs_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "C5C154F1-7F29-4EC0-9971-76A02C6DA8BC", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:diabecare_rs:-:*:*:*:*:*:*:*", "matchCriteriaId": "8628A530-AF34-453E-968A-40DD5B8EE456", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" } ], "cveTags": [], "descriptions": [ { "lang": "en", "value": "SOOIL Developments CoLtd DiabecareRS, AnyDana-i, AnyDana-A, The communication protocol of the insulin pump and AnyDana-i,AnyDana-A mobile apps doesn\u0027t use adequate measures to authenticate the pump before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the keys and spoof the pump via BLE." }, { "lang": "es", "value": "SOOIL Developments CoLtd DiabecareRS, AnyDana-i, AnyDana-A, El protocolo de comunicaci\u00f3n de la bomba de insulina y las aplicaciones m\u00f3viles AnyDana-i, AnyDana-A no utilizan las medidas adecuadas para autenticar la bomba antes de intercambiar claves, lo que permite no autenticarse a atacantes f\u00edsicamente cerca para escuchar las claves y falsificar la bomba por medio de BLE" } ], "id": "CVE-2020-27272", "lastModified": "2024-11-21T05:20:58.667", "metrics": { "cvssMetricV2": [ { "acInsufInfo": false, "baseSeverity": "LOW", "cvssData": { "accessComplexity": "MEDIUM", "accessVector": "ADJACENT_NETWORK", "authentication": "NONE", "availabilityImpact": "NONE", "baseScore": 2.9, "confidentialityImpact": "PARTIAL", "integrityImpact": "NONE", "vectorString": "AV:A/AC:M/Au:N/C:P/I:N/A:N", "version": "2.0" }, "exploitabilityScore": 5.5, "impactScore": 2.9, "obtainAllPrivilege": false, "obtainOtherPrivilege": false, "obtainUserPrivilege": false, "source": "nvd@nist.gov", "type": "Primary", "userInteractionRequired": true } ], "cvssMetricV31": [ { "cvssData": { "attackComplexity": "LOW", "attackVector": "ADJACENT_NETWORK", "availabilityImpact": "NONE", "baseScore": 5.7, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "NONE", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "REQUIRED", "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:H/I:N/A:N", "version": "3.1" }, "exploitabilityScore": 2.1, "impactScore": 3.6, "source": "nvd@nist.gov", "type": "Primary" } ] }, "published": "2021-01-19T17:15:12.613", "references": [ { "source": "ics-cert@hq.dhs.gov", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "sourceIdentifier": "ics-cert@hq.dhs.gov", "vulnStatus": "Modified", "weaknesses": [ { "description": [ { "lang": "en", "value": "NVD-CWE-noinfo" } ], "source": "nvd@nist.gov", "type": "Primary" } ] }
Vulnerability from fkie_nvd
Published
2021-01-19 17:15
Modified
2024-11-21 05:20
Severity ?
Summary
SOOIL Developments CoLtd DiabecareRS, AnyDana-i ,AnyDana-A, communication protocol of the insulin pump & AnyDana-i,AnyDana-A mobile apps doesnt use adequate measures to protect encryption keys in transit which allows unauthenticated physically proximate attacker to sniff keys via (BLE).
References
▶ | URL | Tags | |
---|---|---|---|
ics-cert@hq.dhs.gov | https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01 | Third Party Advisory, US Government Resource | |
af854a3a-2127-422b-91ae-364da2661108 | https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01 | Third Party Advisory, US Government Resource |
Impacted products
Vendor | Product | Version | |
---|---|---|---|
sooil | anydana-a_firmware | * | |
sooil | anydana-a | - | |
sooil | anydana-i_firmware | * | |
sooil | anydana-i | - | |
sooil | diabecare_rs_firmware | * | |
sooil | diabecare_rs | - |
{ "configurations": [ { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:anydana-a_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "808DBA89-C2C2-4C03-814D-DB0B4DFF8B22", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:anydana-a:-:*:*:*:*:*:*:*", "matchCriteriaId": "2624582E-762F-4CDB-B974-1BC971F1544D", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:anydana-i_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "98E25B7D-C1ED-4144-8411-171C7B6426E4", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:anydana-i:-:*:*:*:*:*:*:*", "matchCriteriaId": "E25CB863-07D9-4E90-8022-6C2B4C52EDFC", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" }, { "nodes": [ { "cpeMatch": [ { "criteria": "cpe:2.3:o:sooil:diabecare_rs_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "C5C154F1-7F29-4EC0-9971-76A02C6DA8BC", "versionEndExcluding": "3.0", "vulnerable": true } ], "negate": false, "operator": "OR" }, { "cpeMatch": [ { "criteria": "cpe:2.3:h:sooil:diabecare_rs:-:*:*:*:*:*:*:*", "matchCriteriaId": "8628A530-AF34-453E-968A-40DD5B8EE456", "vulnerable": false } ], "negate": false, "operator": "OR" } ], "operator": "AND" } ], "cveTags": [], "descriptions": [ { "lang": "en", "value": "SOOIL Developments CoLtd DiabecareRS, AnyDana-i ,AnyDana-A, communication protocol of the insulin pump \u0026 AnyDana-i,AnyDana-A mobile apps doesnt use adequate measures to protect encryption keys in transit which allows unauthenticated physically proximate attacker to sniff keys via (BLE)." }, { "lang": "es", "value": "SOOIL Developments CoLtd DiabecareRS, AnyDana-i, AnyDana-A, el protocolo de comunicaci\u00f3n de la bomba de insulina y las aplicaciones m\u00f3viles AnyDana-i, AnyDana-A no utilizan las medidas adecuadas para proteger las claves de cifrado en tr\u00e1nsito, lo que permite a un atacante f\u00edsicamente pr\u00f3ximo no autenticado rastrear claves por medio de (BLE)" } ], "id": "CVE-2020-27270", "lastModified": "2024-11-21T05:20:58.577", "metrics": { "cvssMetricV2": [ { "acInsufInfo": false, "baseSeverity": "LOW", "cvssData": { "accessComplexity": "MEDIUM", "accessVector": "ADJACENT_NETWORK", "authentication": "NONE", "availabilityImpact": "NONE", "baseScore": 2.9, "confidentialityImpact": "PARTIAL", "integrityImpact": "NONE", "vectorString": "AV:A/AC:M/Au:N/C:P/I:N/A:N", "version": "2.0" }, "exploitabilityScore": 5.5, "impactScore": 2.9, "obtainAllPrivilege": false, "obtainOtherPrivilege": false, "obtainUserPrivilege": false, "source": "nvd@nist.gov", "type": "Primary", "userInteractionRequired": true } ], "cvssMetricV31": [ { "cvssData": { "attackComplexity": "LOW", "attackVector": "ADJACENT_NETWORK", "availabilityImpact": "NONE", "baseScore": 5.7, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "NONE", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "REQUIRED", "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:H/I:N/A:N", "version": "3.1" }, "exploitabilityScore": 2.1, "impactScore": 3.6, "source": "nvd@nist.gov", "type": "Primary" } ] }, "published": "2021-01-19T17:15:12.520", "references": [ { "source": "ics-cert@hq.dhs.gov", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" }, { "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Third Party Advisory", "US Government Resource" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "sourceIdentifier": "ics-cert@hq.dhs.gov", "vulnStatus": "Modified", "weaknesses": [ { "description": [ { "lang": "en", "value": "CWE-522" } ], "source": "nvd@nist.gov", "type": "Primary" } ] }
CVE-2020-27272 (GCVE-0-2020-27272)
Vulnerability from cvelistv5
Published
2021-01-19 16:18
Modified
2024-08-04 16:11
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- KEY EXCHANGE WITHOUT ENTITY AUTHENTICATION CWE322
Summary
SOOIL Developments CoLtd DiabecareRS, AnyDana-i, AnyDana-A, The communication protocol of the insulin pump and AnyDana-i,AnyDana-A mobile apps doesn't use adequate measures to authenticate the pump before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the keys and spoof the pump via BLE.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||
---|---|---|---|---|
n/a | SOOIL Developments CoLtd DiabecareRS,AnyDana-i,AnyDana-A |
Version: Dana Diabecare RS,AnyDana-i,AnyDana-A versions prior to 3.0 |
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-04T16:11:36.580Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "product": "SOOIL Developments CoLtd DiabecareRS,AnyDana-i,AnyDana-A", "vendor": "n/a", "versions": [ { "status": "affected", "version": "Dana Diabecare RS,AnyDana-i,AnyDana-A versions prior to 3.0" } ] } ], "descriptions": [ { "lang": "en", "value": "SOOIL Developments CoLtd DiabecareRS, AnyDana-i, AnyDana-A, The communication protocol of the insulin pump and AnyDana-i,AnyDana-A mobile apps doesn\u0027t use adequate measures to authenticate the pump before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the keys and spoof the pump via BLE." } ], "problemTypes": [ { "descriptions": [ { "description": "KEY EXCHANGE WITHOUT ENTITY AUTHENTICATION CWE322", "lang": "en", "type": "text" } ] } ], "providerMetadata": { "dateUpdated": "2021-01-19T16:18:13", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2020-27272", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "SOOIL Developments CoLtd DiabecareRS,AnyDana-i,AnyDana-A", "version": { "version_data": [ { "version_value": "Dana Diabecare RS,AnyDana-i,AnyDana-A versions prior to 3.0" } ] } } ] }, "vendor_name": "n/a" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "SOOIL Developments CoLtd DiabecareRS, AnyDana-i, AnyDana-A, The communication protocol of the insulin pump and AnyDana-i,AnyDana-A mobile apps doesn\u0027t use adequate measures to authenticate the pump before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the keys and spoof the pump via BLE." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "KEY EXCHANGE WITHOUT ENTITY AUTHENTICATION CWE322" } ] } ] }, "references": { "reference_data": [ { "name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01", "refsource": "MISC", "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2020-27272", "datePublished": "2021-01-19T16:18:13", "dateReserved": "2020-10-19T00:00:00", "dateUpdated": "2024-08-04T16:11:36.580Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2020-27266 (GCVE-0-2020-27266)
Vulnerability from cvelistv5
Published
2021-01-19 21:17
Modified
2024-08-04 16:11
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-603 - USE OF CLIENT-SIDE AUTHENTICATION
Summary
In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a client-side control vulnerability in the insulin pump and its AnyDana-i and AnyDana-A mobile applications allows physically proximate attackers to bypass user authentication checks via Bluetooth Low Energy.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||
---|---|---|---|---|
n/a | SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A |
Version: Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0 |
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-04T16:11:36.390Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "product": "SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A", "vendor": "n/a", "versions": [ { "status": "affected", "version": "Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0" } ] } ], "descriptions": [ { "lang": "en", "value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a client-side control vulnerability in the insulin pump and its AnyDana-i and AnyDana-A mobile applications allows physically proximate attackers to bypass user authentication checks via Bluetooth Low Energy." } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-603", "description": "USE OF CLIENT-SIDE AUTHENTICATION CWE-603", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2021-01-19T21:17:33", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2020-27266", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A", "version": { "version_data": [ { "version_value": "Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0" } ] } } ] }, "vendor_name": "n/a" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a client-side control vulnerability in the insulin pump and its AnyDana-i and AnyDana-A mobile applications allows physically proximate attackers to bypass user authentication checks via Bluetooth Low Energy." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "USE OF CLIENT-SIDE AUTHENTICATION CWE-603" } ] } ] }, "references": { "reference_data": [ { "name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01", "refsource": "MISC", "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2020-27266", "datePublished": "2021-01-19T21:17:33", "dateReserved": "2020-10-19T00:00:00", "dateUpdated": "2024-08-04T16:11:36.390Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2020-27270 (GCVE-0-2020-27270)
Vulnerability from cvelistv5
Published
2021-01-19 16:17
Modified
2024-08-04 16:11
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- UNPROTECTED TRANSPORT OF CREDENTIALS CWE523
Summary
SOOIL Developments CoLtd DiabecareRS, AnyDana-i ,AnyDana-A, communication protocol of the insulin pump & AnyDana-i,AnyDana-A mobile apps doesnt use adequate measures to protect encryption keys in transit which allows unauthenticated physically proximate attacker to sniff keys via (BLE).
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||
---|---|---|---|---|
n/a | SOOIL Developments CoLtd DiabecareRS,AnyDana-i,AnyDana-A |
Version: Dana DiabecareRS,AnyDana-i,AnyDana-A,versions prior to 3.0 |
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-04T16:11:36.599Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "product": "SOOIL Developments CoLtd DiabecareRS,AnyDana-i,AnyDana-A", "vendor": "n/a", "versions": [ { "status": "affected", "version": "Dana DiabecareRS,AnyDana-i,AnyDana-A,versions prior to 3.0" } ] } ], "descriptions": [ { "lang": "en", "value": "SOOIL Developments CoLtd DiabecareRS, AnyDana-i ,AnyDana-A, communication protocol of the insulin pump \u0026 AnyDana-i,AnyDana-A mobile apps doesnt use adequate measures to protect encryption keys in transit which allows unauthenticated physically proximate attacker to sniff keys via (BLE)." } ], "problemTypes": [ { "descriptions": [ { "description": "UNPROTECTED TRANSPORT OF CREDENTIALS CWE523", "lang": "en", "type": "text" } ] } ], "providerMetadata": { "dateUpdated": "2021-01-19T16:17:59", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2020-27270", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "SOOIL Developments CoLtd DiabecareRS,AnyDana-i,AnyDana-A", "version": { "version_data": [ { "version_value": "Dana DiabecareRS,AnyDana-i,AnyDana-A,versions prior to 3.0" } ] } } ] }, "vendor_name": "n/a" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "SOOIL Developments CoLtd DiabecareRS, AnyDana-i ,AnyDana-A, communication protocol of the insulin pump \u0026 AnyDana-i,AnyDana-A mobile apps doesnt use adequate measures to protect encryption keys in transit which allows unauthenticated physically proximate attacker to sniff keys via (BLE)." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "UNPROTECTED TRANSPORT OF CREDENTIALS CWE523" } ] } ] }, "references": { "reference_data": [ { "name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01", "refsource": "MISC", "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2020-27270", "datePublished": "2021-01-19T16:17:59", "dateReserved": "2020-10-19T00:00:00", "dateUpdated": "2024-08-04T16:11:36.599Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2020-27268 (GCVE-0-2020-27268)
Vulnerability from cvelistv5
Published
2021-01-19 21:17
Modified
2024-08-04 16:11
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-602 - CLIENT-SIDE ENFORCEMENT OF SERVER-SIDE SECURITY
Summary
In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a client-side control vulnerability in the insulin pump and its AnyDana-i and AnyDana-A mobile applications allows physically proximate attackers to bypass checks for default PINs via Bluetooth Low Energy.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||
---|---|---|---|---|
n/a | SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A |
Version: Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0 |
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-04T16:11:36.584Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "product": "SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A", "vendor": "n/a", "versions": [ { "status": "affected", "version": "Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0" } ] } ], "descriptions": [ { "lang": "en", "value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a client-side control vulnerability in the insulin pump and its AnyDana-i and AnyDana-A mobile applications allows physically proximate attackers to bypass checks for default PINs via Bluetooth Low Energy." } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-602", "description": "CLIENT-SIDE ENFORCEMENT OF SERVER-SIDE SECURITY CWE-602", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2021-01-19T21:17:53", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2020-27268", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A", "version": { "version_data": [ { "version_value": "Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0" } ] } } ] }, "vendor_name": "n/a" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a client-side control vulnerability in the insulin pump and its AnyDana-i and AnyDana-A mobile applications allows physically proximate attackers to bypass checks for default PINs via Bluetooth Low Energy." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "CLIENT-SIDE ENFORCEMENT OF SERVER-SIDE SECURITY CWE-602" } ] } ] }, "references": { "reference_data": [ { "name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01", "refsource": "MISC", "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2020-27268", "datePublished": "2021-01-19T21:17:53", "dateReserved": "2020-10-19T00:00:00", "dateUpdated": "2024-08-04T16:11:36.584Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2020-27256 (GCVE-0-2020-27256)
Vulnerability from cvelistv5
Published
2021-01-19 20:46
Modified
2024-08-04 16:11
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-798 - USE OF HARD CODED CREDENTIALS
Summary
In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a hard-coded physician PIN in the physician menu of the insulin pump allows attackers with physical access to change insulin therapy settings.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||
---|---|---|---|---|
n/a | SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A |
Version: Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0 |
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-04T16:11:36.321Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "product": "SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A", "vendor": "n/a", "versions": [ { "status": "affected", "version": "Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0" } ] } ], "descriptions": [ { "lang": "en", "value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a hard-coded physician PIN in the physician menu of the insulin pump allows attackers with physical access to change insulin therapy settings." } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-798", "description": "USE OF HARD CODED CREDENTIALS CWE-798", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2021-01-19T20:46:30", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2020-27256", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A", "version": { "version_data": [ { "version_value": "Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0" } ] } } ] }, "vendor_name": "n/a" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a hard-coded physician PIN in the physician menu of the insulin pump allows attackers with physical access to change insulin therapy settings." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "USE OF HARD CODED CREDENTIALS CWE-798" } ] } ] }, "references": { "reference_data": [ { "name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01", "refsource": "MISC", "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2020-27256", "datePublished": "2021-01-19T20:46:30", "dateReserved": "2020-10-19T00:00:00", "dateUpdated": "2024-08-04T16:11:36.321Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2020-27276 (GCVE-0-2020-27276)
Vulnerability from cvelistv5
Published
2021-01-19 16:18
Modified
2024-08-04 16:11
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-290 - AUTHENTICATION BYPASS BY SPOOFING
Summary
SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i & AnyDana-A mobile apps doesn't use adequate measures to authenticate the communicating entities before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the authentication sequence via Bluetooth Low Energy.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||
---|---|---|---|---|
n/a | SOOIL Developments CoLtd DiabecareRS,AnyDana-i,AnyDana-A |
Version: Dana DiabecareRS, AnyDana-i, AnyDana-A All versions prior to 3.0 |
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-04T16:11:36.643Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "product": "SOOIL Developments CoLtd DiabecareRS,AnyDana-i,AnyDana-A", "vendor": "n/a", "versions": [ { "status": "affected", "version": "Dana DiabecareRS, AnyDana-i, AnyDana-A All versions prior to 3.0" } ] } ], "descriptions": [ { "lang": "en", "value": "SOOIL Developments Co Ltd DiabecareRS,AnyDana-i \u0026 AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i \u0026 AnyDana-A mobile apps doesn\u0027t use adequate measures to authenticate the communicating entities before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the authentication sequence via Bluetooth Low Energy." } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-290", "description": "AUTHENTICATION BYPASS BY SPOOFING CWE-290", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2021-01-19T16:18:20", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2020-27276", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "SOOIL Developments CoLtd DiabecareRS,AnyDana-i,AnyDana-A", "version": { "version_data": [ { "version_value": "Dana DiabecareRS, AnyDana-i, AnyDana-A All versions prior to 3.0" } ] } } ] }, "vendor_name": "n/a" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "SOOIL Developments Co Ltd DiabecareRS,AnyDana-i \u0026 AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i \u0026 AnyDana-A mobile apps doesn\u0027t use adequate measures to authenticate the communicating entities before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the authentication sequence via Bluetooth Low Energy." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "AUTHENTICATION BYPASS BY SPOOFING CWE-290" } ] } ] }, "references": { "reference_data": [ { "name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01", "refsource": "MISC", "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2020-27276", "datePublished": "2021-01-19T16:18:20", "dateReserved": "2020-10-19T00:00:00", "dateUpdated": "2024-08-04T16:11:36.643Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2020-27264 (GCVE-0-2020-27264)
Vulnerability from cvelistv5
Published
2021-01-19 20:46
Modified
2024-08-04 16:11
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-330 - USE OF INSUFFICIENTLY RANDOM VALUES
Summary
In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications use deterministic keys, which allows unauthenticated, physically proximate attackers to brute-force the keys via Bluetooth Low Energy.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||
---|---|---|---|---|
n/a | SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A |
Version: Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0 |
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-04T16:11:36.425Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "product": "SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A", "vendor": "n/a", "versions": [ { "status": "affected", "version": "Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0" } ] } ], "descriptions": [ { "lang": "en", "value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications use deterministic keys, which allows unauthenticated, physically proximate attackers to brute-force the keys via Bluetooth Low Energy." } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-330", "description": "USE OF INSUFFICIENTLY RANDOM VALUES CWE-330", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2021-01-19T20:46:53", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2020-27264", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A", "version": { "version_data": [ { "version_value": "Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0" } ] } } ] }, "vendor_name": "n/a" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications use deterministic keys, which allows unauthenticated, physically proximate attackers to brute-force the keys via Bluetooth Low Energy." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "USE OF INSUFFICIENTLY RANDOM VALUES CWE-330" } ] } ] }, "references": { "reference_data": [ { "name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01", "refsource": "MISC", "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2020-27264", "datePublished": "2021-01-19T20:46:53", "dateReserved": "2020-10-19T00:00:00", "dateUpdated": "2024-08-04T16:11:36.425Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }
CVE-2020-27269 (GCVE-0-2020-27269)
Vulnerability from cvelistv5
Published
2021-01-19 21:17
Modified
2024-08-04 16:11
Severity ?
VLAI Severity ?
EPSS score ?
CWE
- CWE-294 - AUTHENTICATION BYPASS BY CAPTURE-REPLAY
Summary
In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications lacks replay protection measures, which allows unauthenticated, physically proximate attackers to replay communication sequences via Bluetooth Low Energy.
References
► | URL | Tags | |||
---|---|---|---|---|---|
|
Impacted products
Vendor | Product | Version | ||
---|---|---|---|---|
n/a | SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A |
Version: Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0 |
{ "containers": { "adp": [ { "providerMetadata": { "dateUpdated": "2024-08-04T16:11:36.578Z", "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE" }, "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "title": "CVE Program Container" } ], "cna": { "affected": [ { "product": "SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A", "vendor": "n/a", "versions": [ { "status": "affected", "version": "Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0" } ] } ], "descriptions": [ { "lang": "en", "value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications lacks replay protection measures, which allows unauthenticated, physically proximate attackers to replay communication sequences via Bluetooth Low Energy." } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-294", "description": "AUTHENTICATION BYPASS BY CAPTURE-REPLAY CWE-294", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2021-01-19T21:17:59", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ], "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2020-27269", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A", "version": { "version_data": [ { "version_value": "Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0" } ] } } ] }, "vendor_name": "n/a" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications lacks replay protection measures, which allows unauthenticated, physically proximate attackers to replay communication sequences via Bluetooth Low Energy." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "AUTHENTICATION BYPASS BY CAPTURE-REPLAY CWE-294" } ] } ] }, "references": { "reference_data": [ { "name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01", "refsource": "MISC", "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01" } ] } } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2020-27269", "datePublished": "2021-01-19T21:17:59", "dateReserved": "2020-10-19T00:00:00", "dateUpdated": "2024-08-04T16:11:36.578Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }