fkie_cve-2018-5446
Vulnerability from fkie_nvd
Published
2018-05-04 18:29
Modified
2025-05-22 18:15
Summary
Medtronic 2090 CareLink Programmer uses a per-product username and password that is stored in a recoverable format.
Impacted products



{
  "configurations": [
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:medtronic:2090_carelink_programmer_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "884B08DC-B6F9-4B34-9679-61C256437DF9",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:medtronic:2090_carelink_programmer:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "215ED381-B433-4D05-97AD-4E7287E5820D",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    }
  ],
  "cveTags": [],
  "descriptions": [
    {
      "lang": "en",
      "value": "Medtronic 2090 CareLink Programmer \n\nuses a per-product username and password that is stored in a recoverable format."
    },
    {
      "lang": "es",
      "value": "Todas las versiones de Medtronic 2090 Carelink Programmer se han visto afectadas por un nombre de usuario y una contrase\u00f1a per-product que est\u00e1n almacenadas en un formato recuperable que podr\u00eda permitir que un atacante con acceso f\u00edsico a un 090 Programmer obtenga credenciales per-product para la red de despliegue del software."
    }
  ],
  "id": "CVE-2018-5446",
  "lastModified": "2025-05-22T18:15:22.840",
  "metrics": {
    "cvssMetricV2": [
      {
        "acInsufInfo": false,
        "baseSeverity": "LOW",
        "cvssData": {
          "accessComplexity": "LOW",
          "accessVector": "LOCAL",
          "authentication": "NONE",
          "availabilityImpact": "NONE",
          "baseScore": 2.1,
          "confidentialityImpact": "PARTIAL",
          "integrityImpact": "NONE",
          "vectorString": "AV:L/AC:L/Au:N/C:P/I:N/A:N",
          "version": "2.0"
        },
        "exploitabilityScore": 3.9,
        "impactScore": 2.9,
        "obtainAllPrivilege": false,
        "obtainOtherPrivilege": false,
        "obtainUserPrivilege": false,
        "source": "nvd@nist.gov",
        "type": "Primary",
        "userInteractionRequired": false
      }
    ],
    "cvssMetricV30": [
      {
        "cvssData": {
          "attackComplexity": "LOW",
          "attackVector": "PHYSICAL",
          "availabilityImpact": "NONE",
          "baseScore": 5.3,
          "baseSeverity": "MEDIUM",
          "confidentialityImpact": "HIGH",
          "integrityImpact": "NONE",
          "privilegesRequired": "NONE",
          "scope": "CHANGED",
          "userInteraction": "NONE",
          "vectorString": "CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:N/A:N",
          "version": "3.0"
        },
        "exploitabilityScore": 0.9,
        "impactScore": 4.0,
        "source": "nvd@nist.gov",
        "type": "Primary"
      }
    ],
    "cvssMetricV31": [
      {
        "cvssData": {
          "attackComplexity": "HIGH",
          "attackVector": "PHYSICAL",
          "availabilityImpact": "NONE",
          "baseScore": 4.9,
          "baseSeverity": "MEDIUM",
          "confidentialityImpact": "HIGH",
          "integrityImpact": "NONE",
          "privilegesRequired": "NONE",
          "scope": "CHANGED",
          "userInteraction": "NONE",
          "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:C/C:H/I:N/A:N",
          "version": "3.1"
        },
        "exploitabilityScore": 0.5,
        "impactScore": 4.0,
        "source": "ics-cert@hq.dhs.gov",
        "type": "Secondary"
      }
    ]
  },
  "published": "2018-05-04T18:29:00.523",
  "references": [
    {
      "source": "ics-cert@hq.dhs.gov",
      "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-2090-29901.html"
    },
    {
      "source": "ics-cert@hq.dhs.gov",
      "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-058-01"
    },
    {
      "source": "af854a3a-2127-422b-91ae-364da2661108",
      "tags": [
        "Third Party Advisory",
        "US Government Resource"
      ],
      "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01"
    }
  ],
  "sourceIdentifier": "ics-cert@hq.dhs.gov",
  "vulnStatus": "Modified",
  "weaknesses": [
    {
      "description": [
        {
          "lang": "en",
          "value": "CWE-257"
        }
      ],
      "source": "ics-cert@hq.dhs.gov",
      "type": "Secondary"
    },
    {
      "description": [
        {
          "lang": "en",
          "value": "CWE-522"
        }
      ],
      "source": "nvd@nist.gov",
      "type": "Primary"
    }
  ]
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
  • Confirmed: The vulnerability is confirmed from an analyst perspective.
  • Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
  • Patched: This vulnerability was successfully patched by the user reporting the sighting.
  • Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
  • Not confirmed: The user expresses doubt about the veracity of the vulnerability.
  • Not patched: This vulnerability was not successfully patched by the user reporting the sighting.


Loading…

Loading…